Download presentation
Presentation is loading. Please wait.
Published byLilian Armstrong Modified over 5 years ago
1
2008 Progress Seminar 유방암 자가 진단기 Subtitle 1 Subtitle 2 (발표자이름) (날짜)
2
Abstract 유방암 자가진단 (not only at clinics) magnified pressure LCD
Probe x9 Conventional Pressure sensor (Force Sensor Resistor) LCD 4.5cm 4.5cm Flexible PCB (copper) Breast Capacitive Pressure Sensor Fabric Sensor (Silver Paste) 유방암 자가진단 (not only at clinics) magnified pressure
3
Conventional Sensor LCD 저항으로 이루어진 센서 문제점 Debugging 완료.
Probe x9 Conventional Pressure sensor (FSR) LCD 4.5cm 4.5cm Breast 저항으로 이루어진 센서 문제점 Debugging 완료. 2x8 array sensors Progress ADC, calibration 중 Pressure mapping needed after.
4
Fabric Sensor LCD Capacitive Pressure Sensor Progress
Probe LCD Capacitive Pressure Sensor Progress Circuit design 중. Consider Fabric Unstable Reproduce/ use FPCB 4.5cm 4.5cm Breast Fabric Sensor (silver paste) Line pitch
5
Fabric Sensor Reduce crosstalk and wiring error from 20% to 1% 6mm
Capacitance measured: 3.5pF (uncompressed) to 5.8pF(5N/cm^2)
6
Fabric Sensor Capacitance measured: 3.5pF (uncompressed) to 5.8pF(5N/cm^2) Electrode array – silver coated yarn Compressible Spacer – 6mm thickness, 50% compressibility at 1.8N/cm^2 <<Muller Textil>> Shielding – EMI – conductive grounding pads Electrode array – silver paste on the fibre Metalized nylon fibres Compressible Spacer – 3mm thickness, 50% compressibility at 1.8N/cm^2 <<Muller Textil>>
7
Fabric Sensor Circuit Bandpass Filter? AD7745
Capacitance to digital converter Sine(Triangle) wave oscillator
8
Flexible PCB LCD Capacitive Pressure Sensor
Probe Flexible PCB (copper) LCD 4.5cm 4.5cm Breast Capacitive Pressure Sensor 10x10 copper line – ADC MUX 연결 dielectric 제질: 폴리 이미드 사용 Progress 2pF difference Fabric circuit 완성되면 apply to the FPCB as well.
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.