Download presentation
1
Gas Detection Instruments
Direct-Reading Gas Detection Instruments Presented by Michael D. Shaw
2
“DIRECT-READING” Means that instrument provides display
in engineering units (usually ppm or percent) Implication (not always true) that unit is reading in real time
3
GASES OF INTEREST Combustible Oxygen (deficiency and enrichment) Toxic
4
MOST COMMON TECHNOLOGIES EMPLOYED IN INDUSTRIAL HYGIENE APPLICATIONS
5
CATALYTIC SENSORS (also known as Pellistors)
Theory - Target gas is oxidized on catalytic element The change in temperature causes a change in resistance that is measured by the meter Application – combustible gases
6
CATALYTIC SENSORS Advantage Disadvantages Long life
Different responses for each combustible gas Can be poisoned Most require at least 10% oxygen to work properly Limited to percent-level detection
7
METAL OXIDE SEMICONDUCTOR
(Also known as solid state, Figaro, or Taguchi sensors) Theory – Target gas reacts with MOS (SnO2 ) and changes its resistance – as measured by the meter Application - Nearly all oxidizable gases
8
METAL OXIDE SEMICONDUCTOR
Advantage Inexpensive Disadvantages Not selective, and this is often misrepresented Affected by humidity Not truly analytical—better as “go/no-go” detectors
9
NON-DISPERSIVE INFRARED (NDIR)
Theory – Target gas absorbs infrared light at a particular wavelength Via Beer’s Law calculations —concentration can be determined Application – Hundreds of compounds, especially organics
10
BEER’S LAW ( FOR EXTRA CREDIT )
"The deeper the glass, The darker the brew, The less of the incident Light that gets through" I = I0ekP I = intensity of light striking the detector I0 = measured signal at zero concentration of target gas k = system dependent constant P = concentration of target gas e = base for natural logarithms
11
NON-DISPERSIVE INFRARED (NDIR)
Advantage Extremely versatile Disadvantages Expensive Can be fragile and complicated to service
12
PHOTOIONIZATION DETECTOR (PID)
Theory - Ultraviolet light ionizes the target gas Current produced is proportional to concentration Application – Volatile Organic Compounds (VOC’s)
13
PHOTOIONIZATION DETECTOR (PID)
Advantage Detects a wide range of volatile organic compounds Quite effective when used in conjunction with chromatographic column → Portable GC’s Disadvantages Nonselective among organic vapors below ionization potential of lamp Affected by high humidity UV lamps are expensive
14
ELECTROCHEMICAL SENSORS
Theory – Target gas is adsorbed on an electrocatalytic sensing electrode, after passing through a diffusion medium, and is electrochemically reacted. The current produced by this reaction is directly proportional to the gas concentration. Applications Br2, CO, Cl2, ClO2, C2H4, ethylene oxide HCHO, H2, hydrazine, HBr, HCl, HCN, H2S, NO, NO2, O3, propylene oxide, SO2, and oxygen-- among others
15
ELECTROCHEMICAL SENSORS
Advantages Inexpensive Linear output Can be miniaturized Disadvantages Prone to interferences in unskilled hands Affected by temperature— readily correctable
16
WHAT IS THE MOST IMPORTANT THING IN GAS DETECTION WHEN USING
DIRECT-READING INSTRUMENTS? PROPER CALIBRATION!! Without a clean zero gas and an accurate verified calibration standard—there is no point in doing any gas detection
17
MONITORING PRODUCTS OFFERED BY
18
PORTABLE ANALYZER
19
SINGLE POINT APPLICATION
CONTINUOUS MONITORING SYSTEM SINGLE POINT APPLICATION
20
CONTINUOUS MONITORING SYSTEM
TWO POINT APPLICATION
21
MULTI-POINT (THREE OR MORE) APPLICATION
CONTINUOUS MONITORING SYSTEM MULTI-POINT (THREE OR MORE) APPLICATION
22
DATA ACQUISITION, ARCHIVING,
ARC-MAX® DATA ACQUISITION, ARCHIVING, AND REPORTING
23
ARC-MAX® MAIN SCREEN
24
ARC-MAX® SHIFT REPORT
25
ARC-MAX® TRENDING
26
ARC-MAX® ALARM LOG
27
For Survey Applications, We Recommend Our Nomad Data Logger
12-bit resolution Models available for 0-100mV, 0-2.5V and 4-20mA inputs Self-contained units also available for temperature, humidity and pH Sampling rate adjustable—1 second to 10 hours in 1 second intervals Storage capacity—16336 readings Excellent software included
28
Visualize your measurement session with an auto-scaling time history graph
29
Time history graph detail screen
Zoomed in on particular portion of the curve
30
Ability To View All Individual Data Points
32
CONCLUDING REMARKS We've looked at the most widely used operating
principles in direct-reading gas detection instruments We've touched upon the importance of calibration We've examined two approaches to data acquisition
33
SPECIAL THANKS TO PROFESSOR GEORGE BYRNS