Download presentation
Presentation is loading. Please wait.
Published byRobert Cooke Modified over 11 years ago
1
Dual Chamber Temporary Pacing Operations & Troubleshooting
2
Indications for Dual Chamber Temporary Pacing
Complete heart block Sinus bradycardia Bradycardia with congestive heart failure Atrial and/or ventricular ectopic arrhythmia Reentrant tachycardias During pulse generator replacement
3
Indications for Dual Chamber Temporary Pacing
Prior to the implant of a permanent pacemaker Temporary support of a patient after heart surgery Acute myocardial infarction complicated by heart block High rate burst stimuli for the treatment of some tachyarrhythmias
4
The NASPE/BPEG Generic (NBG) Code
Position I II III IV V Category Chamber(s) Paced Chamber(s) Sensed Response to Sensing Programmability Rate Modulation Antitachyarrhythmia Function(s) O = None A = Atrium V = Ventricle D = Dual (A+V) O = None A = Atrium V = Ventricle D = Dual (A+V) O = None T = Triggered I = Inhibited D = Dual (T+I) O = None P = Simple Programmable M = Multiprogrammable C = Communicating R = Rate Modulation O = None P = Pacing S = Shock D = Dual (P+S) Manufacturer’s Designation Only S = Single (A or V) S = Single (A or V) Note: Positions I through III are used exclusively for antibradyarrhythmia function
5
Asynchronous Modes AOO VOO DOO
6
DVI 1 2 3 4 5 6 Vs ° Vp * Ap = AV interval (140 ms)
Lower rate = 60 ppm V-A interval = 860 ms = PVARP (250 ms) = indicates cycle restarted by sensed or paced event
7
Action or Response to a Sensed Event
DDD Pacing Chamber Paced Chamber Sensed Action or Response to a Sensed Event D D D
8
DDD Pacemaker Provides: Results in: AV Synchrony Rate Variability
Changes in cardiac output Management of rhythm Improved quality of life
9
Parameter Adjustments of the Temporary DDD Pacemaker
Lower Rate A–V Interval Upper Rate Output Sensitivity Refractory Period (PVARP)
10
ECG Function DDD Mode
11
Timing Intervals in milliseconds (ms)
A–V 250 ms V–V 1000 ms DDD / 60 / 250 / 125 / 155
12
Calculation of Atrial Escape Interval
Atrial Escape Lower Rate – A–V Interval = Interval V–V – A–V = V–A (ms) (ms) (ms) (60 ppm) 1000 – 250 = 750 (60 ppm) 1000 – 150 = 850 (70 ppm) 850 – 200 = 650
13
Proper Atrial Sensing is the “Heart” of Physiologic Pacing
14
A Pace / V Pace A–V Sequential Pacing V–A 750 ms A–V 250 ms V–V
DDD / 60 / 250 / 150
15
Programmed Lower Rate 1000 ms Actual V–V 960 ms
A Pace / V Sense Atrial Pacing with Normal A–V Conduction Programmed Lower Rate 1000 ms Actual V–V 960 ms DDD / 60 / 200 / 150
16
A Sense / V Pace Atrial Synchronous Pacing
Spontaneous Atrial Rate = 55 ppm Spontaneous Atrial Rate = 110 ppm DDD / 50 / 150 / 150
17
A Sense / V Sense Spontaneous P-Wave with Normal A–V Conduction
Spontaneous Atrial Rate = 65 ppm Spontaneous P-R Interval = 160 ms DDD / 60 / 200 / 150
18
Timing Operations of DDD Pacing
Ventricular Rate Upper Rate Atrial Tracking Lower Rate Atrial Rate
19
Definitions PVARP – Post Ventricular Atrial Refractory Period
Time after Vs or Vp when atrial rates are ignored A–V – Time from As or Ap to Vp in ms TARP – Total Atrial Refractory Period in ms
20
Refractory Period T-wave sensing P-wave sensing Increase to prevent:
Decrease to permit tachy pacing
21
Wenckebach Operation
22
Wenckebach Operation W = Wenckebach Interval AV AV W AV W AV W AV
PVARP PVARP PVARP PVARP PVARP Upper Rate Upper Rate Upper Rate Upper Rate Upper Rate As As As As As As Vp Vp Vp Vp Vp W = Wenckebach Interval
23
2 : 1 Block
24
2 : 1 Block As As As As As As As As Vp Vp Vp Vp AV AV AV AV PVARP
Upper Rate Upper Rate Upper Rate Upper Rate As As As As As As As As Vp Vp Vp Vp
25
Physiology of Retrograde Conduction
1. Loss of A-V synchrony due to a PVC 2. Sensed retrograde activation 3. A-V interval initiated PVC 4. Prolongation of A-V interval 5. Ventricular pacing synchrononized to retrograde P-waves
26
Pacemaker Mediated Tachycardia (PMT)
Fast ventricular paced rhythm Synchronized to retrograde- conducted P-waves Not normal sinus tachycardia
27
Thresholds
28
Feature of the Temporary DDD Pacemaker
Output Control Atrial Ventricular Pulse Amplitude (ma) Pulse Width (ms)
29
Feature of the Temporary DDD Pacemaker
Sensitivity 2.5 1.25 0.5
30
Troubleshooting
31
Undersensing Failure of the pacemaker to sense intrinsic R-waves or intrinsic P-waves
32
Undersensing Battery depletion Decreased QRS voltage Fusion beat
Dislodged/fractured lead Inappropriate sensitivity setting
33
Oversensing Inhibition of the pacemaker by events the pacemaker should ignore, e.g. EMI, T-waves, and myopotential
34
Oversensing Myopotential inhibition EMI
T-waves outside of refractory period Dislodged/fractured lead Inappropriate sensitivity setting
35
No Output Pacemaker fails to emit stimuli at the programmed intervals
36
No Output Faulty cable connection Dislodged/fractured lead
1 2 3 4 5 AP AS AP AP AP VP VP VP VP VP No atrial output (on ECG) Battery depletion/pacemaker off Oversensing Faulty cable connection Dislodged/fractured lead
37
Loss of Capture Electrical stimuli delivered by the pacemaker does not initiate depolarization of the atria or ventricle
38
Loss of Capture Inappropriate output setting Dislodged/fractured lead
Atrial Noncapture Ventricular Noncapture Fusion Inappropriate output setting Increased resistance to conduction QRS complex not visible Dislodged/fractured lead Tissue is refractory Faulty cable connection
39
Ventricular Safety Pacing
Ap = Atrial Pace Vp = Ventricular Pace Ap Vp Programmed A-V Interval Ventricular Blanking at Atrial Rate 12 ms Ventricular Safety Pace Interval 110 ms Programmed A-V Interval Ventricular Sensing within this period triggers a ventricular output at an A-V interval of 110 ms instead of programmed A-V interval. Ventricular Sensing during the remainder of the programmed A-V interval inhibits the ventricular output. If there is no sensed event, a ventricular output occurs at the programmed A-V interval.
40
Ventricular Safety Pacing (VSP)
1 2 3 4 5 Vs Vp * Ap = AV interval (160 ms) Lower rate = 54 ppm Upper rate = 180 ppm V-A interval = 860 ms = PVARP (250 ms) = indicates cycle restarted by sensed or paced event
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.