Design and construction of a mid-IR SPIDER apparatus 09/10/2012 Malte Christian Brahms Imperial College London 09/10/20121
2
Contents Purpose of SPIDER Working principle Design of mid-IR SPIDER User’s manual Outlook 09/10/20122
3
Purpose Pulse characterised completely by –Field as fct. of time or –Spectral amplitude and phase Oscillations in on the order of fs –Temporal resolution required too high Instead: Measure spectral amplitude and phase: S pectral P hase I nterferometry for D irect E lectric F ield R econstruction (SPIDER) 09/10/20123
4
Purpose SPIDERs available commercially Why build one? –Unusual wavelength in mid-IR –Specific variable wavelength SPIDER needed for TOPAS 09/10/20124
5
Contents Purpose of SPIDER Working principle Design of mid-IR SPIDER User’s manual Outlook 09/10/20125
6
Working principle Based on two replicas of the test pulse Displaced relative to each other –In time: Delay τ –In frequency: Shear Ω In frequency domain: 09/10/20126 Frequency Shear Time delay
7
Working principle Interferogram: 09/10/20127
8
Working principle 09/10/20128
9
Working principle 09/10/20129
10
Contents Purpose of SPIDER Working principle Design of mid-IR SPIDER User’s manual Outlook 09/10/201210
11
Design Two problems to solve: –Time delay –Frequency shear Solutions: –Delay: Split mirror Allows variable delay –Shear: SFG in BBO with chirped pulse 09/10/201211
12
Design – Time delay 09/10/201212 Split mirrors
13
Design – Frequency shearing Use SFG with chirped ancillary pulse replica –Chirp: Carrier frequency depends linearly on time: –SFG with same original frequency at two different times different output frequencies: Obvious solution: Piece of glass –But: Most glass almost dispersion-free at ca. 1500nm 09/10/201213 Stretched ancilla Test pulse replicas
14
Design – Frequency shearing Chirping the ancillary pulse Grating Compressor: 09/10/201214 Image: 2010 J. Phys. B: At. Mol. Opt. Phys. 43 103001, p.5 Walmsley et al.: The role of Dispersion in ultrafast optics, Rev. Sci. Instrum., Vol. 27, No.1, Jan 2001, p. 7
15
Design – Recombination 09/10/201215 Sum-frequency generation Focus delayed pulse-pair and chirped ancillary pulse into BBO crystal Image: 2010 J. Phys. B: At. Mol. Opt. Phys. 43 103001 p. 25 SHG + Original SFG
16
Design – Constraints On delay τ: Spectrometer resolution and shear –Fringe spacing 2π/τ must be sufficiently large τ cannot be too large –τ must be sufficiently large resolve peaks in FT On shear Ω: –Small enough to satisfy the sampling theorem: –Satisfied in practice, usual: Ω=10% of bandwidth 09/10/201216 T: Reconstruction window
17
Design – Constraints On chirp : –Only SFG with monochromatic field conserves shape of spectrum: In time domain: In frequency domain (if E anc is monochromatic): –Need slow-varying, quasi-monochromatic frequency lower limit on 09/10/201217
18
Design - Constraints Resolution: Shear size: Lower limit: 09/10/201218
19
Design – Constraints With numbers: At 1300-2000nm for a ca. 40fs pulse Both the constraints and the value are wavelength-dependent! Chirp determined by compressor angle and separation –Can vary angle easily, use mostly –If necessary, enough space to move one grating 09/10/201219
20
Design – Setup 09/10/201220 Spectromete r A1 A2 A3 f=100mm D-mirror BS (400- 1100nm)
21
Contents Purpose of SPIDER Working principle Design of mid-IR SPIDER User’s manual Outlook 09/10/201221
22
User’s Manual – Practical issues Calibration by simply adding ignores frequency-dependent behaviour (noise) in apparatus –Need to calibrate for this –Use signal without shear instead: Depending on wavelength and spectrometer: –Measure calibration either in SHG (blue) or original (red) –SPIDER in SFG For this SPIDER at 1300-2000nm: –Measure calibration and SPIDER signal in blue, first order 09/10/201222
23
User’s Manual – Practical issues Spectral amplitude: –Determined from calibration signal Take magnitude instead of phase angle: Determining shear Ω: –Measure compressor parameters or block one of sheared pair (spatially) –In practice: Use compressor parameters τ determined by software 09/10/201223
24
User’s manual – Procedure Good signal: –High fringe contrast –Well separated fringes –Near saturation Needed for this: –Phase-matching and crystal position –Spatial overlap (focus on crystal) –Temporal overlap (compressor translation) –Right amount of delay –SHG beams blocked (use aperture A3) 09/10/201224
25
User’s manual – Procedure 1.Take reading without ancillary pulse (calibration) –Block after beam splitter 2.Take SPIDER signal –Unblock ancillary, take trace 3.Analyse OR: Use Tobi’s TwinSPIDER (live operation) –Ask Tobi for more advice 09/10/201225
26
User’s Manual – Software 09/10/201226
27
Outlook This week or next: Use on TOPAS Later: Test on few-cycle pulses Intended for long-term use on TOPAS 09/10/201227