Spectroscopic Analysis for biological samples : towards in situ sample analysis of body fluids Gilwon Yoon September 27, 2006 Seoul National University of Technology
2
Spectra of water, Hb(RBC), albumin, glucose from visible to NIR (water compensated)
3
Absorption spectrum of water
4
Involved Key Technologies Spectroscopic detection target component Interfering substances inhomogeneous medium Visible/IR Light source Light interaction with tissue High S/N electronic detection Chemometrics Clinical test Statistical analysis Prediction of concentration
5
Spectroscopic Analysis – Statistical Methods Influence of measurement setup : Transmission or reflection measurement Influence of red blood cell (hemoglobin) in partial least squares regression (PLSR) analysis Independent Component Analysis (ICA) – a method without calibration process
6
I. Influence of measurement setup : Transmission or reflection measurement (a)(b) Light source Mono- chromator slit DetectorSample Detector
7
Comparison between reflectance and transmittance Jeon, Hwang, Hahn, Yoon (2006), 11:1:014022, Journal of Biomedical Optics wavelength region [nm] SEC [mg/dl] SEP [mg/dl] diffuse reflectance (10mm thick sample) 1100-250027.38275.44 1100-185015.91437.54 1850-250030.57192.00 diffuse transmittance (1mm thick sample) 1100-1800 2064-2338 3.2224.69 1100-18002.8826.77 2064-23384.5043.51 diffuse transmittance (2mm thick sample) 1100-1830 2050-2392 26.4639.07 1100-183040.4783.58
8
Diffuse reflectance between 1100 and 1850 nm (a) SECV with respect to the optimal number of factor, (b) Loading vector of calibration model, (c) Regression vector of calibration model, (d) Prediction of glucose illustrated with the intralipid concentrations of sample solutions.
9
Diffuse transmittance with 1 mm thick samples (a) SECV with respect to the optimal number of factor, (b) Loading vector of calibration model, (c) Regression vector of calibration model, and (d) Prediction of glucose concentrations.
10
II. Influence of red blood cell (hemoglobin) in partial least squares regression (PLSR) analysis Absorption becomes much stronger towards longer wavelengths Dominance of hemoglobin Interferences among the substances in blood or extracellular fluid Effect of preprocessing methods Biological Samples in the near infrared (1000 – 2500 nm)
11
a) Whole blood spectra of 98 samples and saline spectrum, b) Whole blood spectra are correlated with hemoglobin and glucose concentrations at each wavelength and computed correlations coefficients are shown.
12
What is a maximally achievable accuracy ? The standard error of glucose prediction was 25.5 mg/dl and the coefficient of variation in prediction was 11.2%. Kim and Yoon (2006), 11: 041128, Journal of Biomedical Optics Calibration set Prediction set mean value of glucose SEP a (r Pre b )VC Pre c [%] Hb cal Hb pre 22825.5 (0.9764)11.2 Hb high Hb mid 21323.1 (0.9817)10.8 Hb low 22148.7 (0.9279)22.0 Hb mid Hb high 20739.3 (0.9465)19.0 Hb low 22146.9 (0.9328)21.2 Hb low Hb high 20774.2 (0.8672)35.8 Hb mid 21333.8 (0.9603)15.9
13
III. Independent Component Analysis (ICA) – method without calibration process Identification of pure, or individual, absorption spectra of constituent components from the mixture spectra without a priori knowledge of the mixture. This method was tested with a two-component system consisting of aqueous solution of both glucose and sucrose, which exhibit distinct but closely overlapped spectra. ICA combined with principal component analysis was able to identify a spectrum for each component, the correct number of components, and the concentrations of the components in the mixture. This method does not need calibration process. Hahn and Yoon (2006), in print, 45:32, November, Applied Optics
14
Pure, or individual, water-subtracted absorption profiles of Glucose (G) and Sucrose (S)
15
25 measured mid-IR spectra for the mixtures of glucose and sucrose. Water absorption was subtracted to enhance the absorption profile of each component.
16
Extracted pure-component spectra from measured IR spectra of 25. ‘Pure’ and ‘ICA’ represent pure-component absorption spectrum and the ICA-method extracted absorption spectrum respectively.
17
Scatter plot for the reference concentrations and ICs from measured mid-IR spectra
18
Measurement geometry or setup – loading factor analysis can provide actual contribution of wavelength in prediction Dominant absorber such as RBC(hemoglobin) and water in near infrared effect substantially. A proper care is needed. A new method that does not require no concentration information and calibration process is introduced. Summary in Spectroscopic analysis