1 Therapeutic Ultrasound
Clinical Application © 2005 – FA Davis

2 FDA Labeling Requirements
Output frequency Effective Radiating Area (ERA) Beam Nonuniformity Ratio (BNR) Beam profile Date of last service/calibration should also be posted on the unit © 2005 – FA Davis

3 Treatment Area US heats a limited area
About the size of a catsup packet Treatment area should be 2 to 3 times the size of the ERA For larger areas, divide the treatment area into smaller zones © 2005 – FA Davis

4 Direct Coupling Immersion Method Pad/Bladder Method
Coupling Methods Direct Coupling Immersion Method Pad/Bladder Method © 2005 – FA Davis

5 Coupling Methods Ultrasonic energy cannot pass through the air
A coupling medium is required Medium should be water-based Coupling method should confirm to the body area The body area should be clean and relatively hair-free © 2005 – FA Davis

6 Direct Coupling Gel or Creams Only use approved coupling agents
Apply liberally to area Remove air bubbles by passing sound head over area (before power is increased) © 2005 – FA Davis

7 Direct Coupling Move the sound head s-l-o-w-l-y
4 cm/sec Moving the head faster decreases heating If the patient describes discomfort, decrease the output intensity © 2005 – FA Davis

8 Coupling Ability of Various Media
Substance Transmission Saran Wrap 98 Lidex ge, fluocinonide (.05%) 97 Thera-Gesic 97 Mineral oil 97 US Transmission gel 96 US Transmission lotion 90 Chempad-L 68 Hydrocortisone powder (1%) 29 Hydrocortisone powder (10%) 7 Eucerin cream 0 Myoflex 0 White petrolatum gel 0 © 2005 – FA Davis

9 Immersion Technique Used to treat irregularly shaped areas
The limb is immersed in a tub of degassed water If tap water is used, increase the output intensity by 0.5 w/cm2 Transducer is held appx. 1” from the body part Avoid the formation of air bubbles © 2005 – FA Davis

10 Pad (Bladder) Method A mass of conductive gel
Commercial pads Self-made bladders Conforms to the treatment area Commercial pads help limit the size of the treatment area © 2005 – FA Davis

11 Output Frequency Duration Duty Cycle Output Intensity
Output Parameters Output Frequency Duration Duty Cycle Output Intensity © 2005 – FA Davis

12 Output Frequency Determines the treatment depth 1 MHz Output
Deep (5 to 7 cm) tissues Rotator cuff, vastus intermedius, gastroc 3 MHz Output Superficial (up to 3cm deep) tissues Patellar tendon, MCL, brachialis Remember that adipose tissue is transparent to ultrasound © 2005 – FA Davis

13 Treatment Duration Depends on: Vigorous heating
Size of the treatment area Output intensity Therapeutic goals Vigorous heating 1 MHz output 10 to 12 minuts 3 MHz output 3 to 4 minutes Also see “Output Intensity” © 2005 – FA Davis

14 Duty Cycle Determine the proportion of thermal and nonthermal effects
High duty cycle: Predominantly thermal effects Low duty cycle: Predominantly nonthermal effects Thermal effect used in subacute and chronic conditions Nonthermal effects may be beneficial in acute stages © 2005 – FA Davis

15 Output Intensity – 1 MHz At a 100% duty cycle 1.75 W/cm2 output
Approximately 4 min treatment is required to increase tissue temp 4°C © 2005 – FA Davis

16 Output Intensity – 3 MHz At a 100% duty cycle 1.5 W/cm2 output
Approximately 10 min treatment is required to increase tissue temp 4°C © 2005 – FA Davis

17 Ultrasound and Electrical Stimulation
“Combo Treatment” © 2005 – FA Davis

18 US/Electrical Stimulation
Theoretically combines the benefit of the two treatments Used to treat: Trigger points Muscle spasm Output parameters: Thermal ultrasound Motor level electrical stimulation © 2005 – FA Davis

19 Set-up and Application
© 2005 – FA Davis

20 Contraindications Acute injuries (100% duty cycle) Ischemic areas
Areas of impaired circulation including arterial disease Over areas of deep vein thrombosis Anesthetic areas Over cancerous tumors Over sites of active infection or sepsis Over the spinal cord or large nerve plexus in high doses Exposed metal that penetrates the skin (e.g., external fixation devices) Areas around the eyes, heart, skull, or genitals Over the thorax in the presence of an implanted pacemaker Pregnancy when used over the pelvic or lumbar areas Over a fracture site before healing is complete Stress fracture sites or sites of osteoporosis Over the pelvic or lumbar area in menstruating female patients © 2005 – FA Davis

21 Precautions Symptoms may increase after the initial treatments.
Use caution when applying ultrasound around the spinal cord, especially after laminectomy. The use of ultrasound over metal implants is not contraindicated Keep the sound head moving Use caution when applying ultrasound over epiphyseal plates of growing bone © 2005 – FA Davis

22 Instrumentation Duty cycle: Adjusts between continuous and pulsed ultrasound application Frequency: Selects the depth of penetration Gel warmer: Used to preheat the transmission gel, primarily for patient comfort Intensity: Adjusts the output intensity. The WATT METER displays the output in either total watts or watts per square centimeter. Maximum head temperature: Prevents overheating in case the head is not properly coupled. Pause: Interrupts the treatment but retains the remaining amount of treatment time Power: Turns the unit ON or OFF Start-Stop: Initiates or terminates the production of ultrasound from the transducer. Timer: Sets the duration of the treatment. Watt meter: Displays the output of ultrasound in total watts or watts per square centimeter. © 2005 – FA Davis

23 Patient Preparation Establish that no contraindications are present
Determine the method and mode of ultrasound application to be used Clean the area to be treated Determine the type of coupling method to be used Identify a treatment area that is 2 to 3 times the ERA For direct coupling: spread the gel over the area to be treated Explain the sensations to be expected during the treatment: Mild to moderate warmth (but not pain or burning) for thermal treatments No subcutaneous sensations during pulsed ultrasound For thermal treatments pre-heating with a moist heat pack will decrease the treatment time required to reach vigorous heating levels Advise the patient to report any adverse, unusual, or painful sensations during the treatment. © 2005 – FA Davis

24 Initiation of the Treatment
Reduce the INTENSITY to zero before turning on the POWER. Select the appropriate mode for the output (CONTINUOUS or PULSED) Set the WATT METER to displays the appropriate output for the type of treatment Set the TIMER to the appropriate treatment duration Begin slowly moving the sound head over the medium Press the START button to begin the treatment session Units having low BNR may be moved at a slower rate than those with a higher BNR Slowly increase the INTENSITY Keep the sound head moving Move the head at a moderate pace (4 cm per second or slower) Use firm, yet not strong, overlapping strokes If pain is experienced: Move the sound head at a faster rate Use a lower duty cycle Use a lower the intensity If the gel begins to wear away or if the sound head begins sticking on the skin, depress the PAUSE button and apply more gel © 2005 – FA Davis

25 Phonophoresis Application
Preheating the area to increase the absorption of the medication Use only approved ultrasound transmission media. The direct coupling method is recommended Efficacy of phonophoresis using the bladder method has not been established Ensure that the skin is well moistened; avoid areas of dry skin Position the extremity to encourage circulation. Use a continuous output to maximize the effect of phonophoresis Unless the thermal effects of ultrasound are contraindicated Follow the procedures described in “Initiation of the Treatment” After treatment, cover the remaining medication with an occlusive dressing. © 2005 – FA Davis

26 Termination of the Treatment
If the treatment is being terminated prematurely: Reduce the intensity before removing the transducer Clean the remaining gel from the patient’s skin To ensure continuity of treatment sessions: Record the treatment parameters Output frequency Intensity Duration Duty cycle Running count of ultrasound treatments given for this condition Immediately initiate any post-treatment stretching © 2005 – FA Davis

27 Maintenance Daily Maintenance
Clean ultrasound head and transducer face as recommended by the manufacturer Monthly Maintenance Check all electrical cords for frays or kinks Check the sound head cable for frays or kinks Clean the transmitter face as recommended by the manufacturer. Federal regulations require that therapeutic ultrasound units be recalibrated annually by an authorized service technician © 2005 – FA Davis