Presentation is loading. Please wait.

Presentation is loading. Please wait.

Advances in Bioscience Education Summer Workshop Fluorescence and Electron Microscopy June 26 - 29, 2007 Biological Electron Microscope Facility Pacific.

Similar presentations


Presentation on theme: "Advances in Bioscience Education Summer Workshop Fluorescence and Electron Microscopy June 26 - 29, 2007 Biological Electron Microscope Facility Pacific."— Presentation transcript:

1 Advances in Bioscience Education Summer Workshop Fluorescence and Electron Microscopy June 26 - 29, 2007 Biological Electron Microscope Facility Pacific Biosciences Research Center University of Hawai’i at Manoa

2 What is a Microscope?  A tool that magnifies and improves resolution of the components of a structure  Has three components:  sources of illumination,  a magnifying system,  detectors.

3 Sources of Illumination  Light microscopes use a beam of light for illumination and include fluorescence and confocal microscopes  Electron microscopes use electrons as a source of illumination and include transmission and scanning electron microscopes.

4 Light and Electron Microscopes  Lenses are used to control a beam of illumination, magnify, and direct an image to a detector

5 Images and pictures are your data!

6 Epifluorescence Microscopy

7 Common Fluorescence Applications  Localize/identify specific organelles  Detect live cells vs. dead cells, necrotic vs. apoptotic cells  Determine cell membrane permeability  Localize antigen-specific molecules  Multiple labeling

8 Laser Scanning Confocal Microscope  Better resolution  Serial optical sections can be collected from thick specimens  Live or fixed cell and tissue imaging

9 Laser Scanning Confocal Microscopy Photos courtesy of Gregg Meada & Dr. Gert DeCouet, UHM And Dr. Chris Yuen and Dr. David Christopher Drosophila eye Plant Protoplast

10 Epifluorescence vs. Confocal Sample courtesy Gregg Meada & Dr. Gert DeCouet, UHM

11 Scanning Electron Microscopy (SEM)  View outer surface  Coat specimen with gold  No sectioning  High Mag (40x to 300,000x)  High resolution (better than 2 nm)

12 SEM Images

13 Transmission Electron Microscopy (TEM) View inside cell via sections magnification 120,000 X 50,000X

14 Conventional TEM Micrographs Skin Bacteria in cell Apoptosis Chloroplast CollagenVirus in cell

15 Ultra-microtomy  Ultrathin (60-90 nm) sectioning of resin- embedded specimens  Several brands/models available

16 Cryotechniques  Ultrarapid cryofixation  Metal mirror impact  Liquid propane plunge  Freeze fracture with Balzers 400T  Cryosubstitution  Cryoultramicrotomy – Ultrathin frozen sections (primarily for antibody labeling)

17 Immunolocalization  LM  Fluor/confocal  TEM  SEM with backscatter detector

18 Approaches to Immunolabeling  Direct Method: Primary antibody contains label  Indirect Method: Primary antibody followed by labeled secondary antibody  Amplified Method: Methods to add more reporter to labeled site

19 Two-step Indirect Method for Immunolabeling  Fluorescent- conjugated secondary antibody attaches to primary antibody that is bound to antigen

20 Immunolabeling for Transmission Electron Microscopy  Normally do Two-Step Method  Primary antibody applied followed by colloidal gold-labeled secondary antibody  May also be enhanced with silver

21 Colloidal Gold Immunolabeling for TEM  Colloidal gold of defined sizes, e.g., 5 nm, 10 nm, 20 nm, easily conjugated to antibodies  Results in small, round, electron-dense label easily detected with EM  Can be enhanced after labeling to enlarge size for LM or EM

22 Double-labeling Method  Use primary antibodies derived from different animals (e.g., one mouse antibody and one rabbit antibody)  Then use two different secondary antibodies conjugated with different sized gold particles

23 Preparation of Biological Specimens for Immunolabeling  Preserve tissue as closely as possible to its natural state while at the same time maintaining the ability of the antigen to react with the antibody  Chemical fixation OR  Cryofixation

24 Chemical Fixation  Antigenic sites are easily denatured or masked during chemical fixation  Glutaraldehyde gives good fixation but may mask antigens, plus it is fluorescent  Paraformaldehyde often better choice, but results in poor morphology, especially for electron microscopy  May use e.g., 4% paraformaldehyde with 0.5% glutaraldehyde as a good compromise

25 Embedding  Dehydrated tissue is embedded in a plastic resin to make it easier to cut thin sections

26 Steps in Labeling of Sections  Chemical fixation  Dehydration, infiltration, embedding and sectioning  Blocking  Incubation with primary antibody  Washing  Incubation with secondary antibody congugated with reporter (fluorescent probe, colloidal gold)  Washing, optional counterstaining  Mount and view

27 Controls! Controls! Controls!  Omit primary antibody  Irrelevant primary antibody  Pre-immune serum  Perform positive control  Check for autofluorescence  Check for non-specific labeling  Dilution series

28 Light Microscopes

29 Light Path in Fluorescence  Light delivered through excitation filter and then objective lens to specimen where it is absorbed;  emitted light goes back through objective lens through barrier filter and emission filter and then to detector.

30 Fluorescence  Light beam excites the fluorochrome, raising it to a higher energy state,  As it falls back to it’s original state, it releases energy in the form of a light of lower E and longer wavelength than original beam of light

31 Primary Ab = PDI secondary Ab = Alexafluor Blue light = exciting beam green and red light emitted Know Your Artifacts Autofluorescence  And use them to your advantage!  Green is label; orange-red is autofluorescence  Acts as counterstain

32 Fluorescence  Fluorochromes are excited by specific wavelengths of light and emit specific wavelengths of a lower energy (longer wavelength)

33 Filter Cubes for Fluorescence  Filter cubes generally have an excitation filter, a dichroic element, and an emission filter  The elements of a cube are selected for the excitation and fluorescence detection desired

34 Choose Fluorochrome/Filter Combos

35 Laser Scanning Confocal Microscopy  Fluorescence technique  Uses laser light for excitation  Improves image resolution over conventional fluorescence techniques  Optically removes out-of-focus light and detects only signal from focal plane  Can construct an in-focus image of considerable depth from a stack of images taken from different focal planes of a thick specimen  Can then make a 3-D image that can be tilted, rotated, and sliced

36 Principal Light Pathway in Confocal Microscopy  Laser light is scanned pixel by pixel across the sample through the objective lens  Fluorescent light is reflected back through the objective and filters (dichroic mirrors)  Adjustable pinhole apertures for PMTs eliminate out-of- focus flare  Image is detected by photomultiplier(s) and digitized on computer

37 TEM  Transmission Electron Microscope  Illumination source is beam of electrons from tungsten wire  Electromagnetic lenses perform same function as glass lenses in LM  Higher resolution and higher magnification of thin specimens

38 Specimen Preparation for TEM  Chemical fixation with buffered glutaraldehyde  Or 4% paraformaldehyde with >1% glutaraldehyde  Postfixation with osmium tetroxide  Or not, or with subsequent removal from sections  Dehydration and infiltration with liquid epoxy or acrylic resin  Polymerization of hard blocks by heat or UV  Ultramicrotomy – 60-80nm sections  Labeling and/or staining  View with TEM

39 High pressure freezing: Plant tissue is flash frozen in a pressure bomb -197 C Water in the tissue is replaced with acetone over 5 day period Acetone saturated tissue is embedded in resin Resin is cut in thin sections, 80 nm thick Add antibodies - immunolabeling Look under Electron microscope

40 Very Wrinkled

41 Chloroplast Carnage Pretty bad fixation

42 2nd time: stainings were done poorly, but there is hope… Back to the drawing board to start over. But what to correct? What to do different? Will it improve?

43 Despite mistakes, keep moving forward and ignore doubt and negativism that comes with pressure.

44 3 rd time A charm

45 Excellent preservation And Immunolabeling the 3 rd TIME

46 HIGH MAG

47 RE-search Not search Research time is spent: 70% trouble-shooting 15% success 15% communicating success. Must be repeated

48 ROOT

49

50

51 HOOK-o-PLASM

52 PDI in Vacuole

53

54 200 nm g CNGC in Golgi Apparatus

55 c 200 nm G PDI in Golgi Apparatus

56 Dividing mitochondria

57 Channel located to the plasma membrane

58 Channel located to the plasma membrane -plasmolysis

59 We learn more from mistakes than successes…


Download ppt "Advances in Bioscience Education Summer Workshop Fluorescence and Electron Microscopy June 26 - 29, 2007 Biological Electron Microscope Facility Pacific."

Similar presentations


Ads by Google