Low Voltage Technology Pulse Generator Technology Lead Technology
3
Pulse Generator
4
1958, 8 October...
5
Identity ADx XL DR
6
Feed Through Telemetry Coil Sensor X-ray marker Capacitors Defib. protection Crystal
7
All bonding is doubled for safety reasons Golden Thread Welded Bonds
8
Magnet test rate 100 85 Magnet Rate <99/min: (and normal energy consumption) 6 months F-U interval Magnet Rate ~93/min: (and normal energy consumption) Replacement within 3 months Magnet Rate <85/min: Immediate replacement Elective Replacement Indicator 98.6 BOL 86.3 ERI EOL EOJ
9
Stimulation rate 63.2 65.9 67.4 69.0 70.6 64.5 EOLBOL 10203040 Battery impedance kOhms Elective Replacement Indicator
10
5 10 15 20 25 Current drain µA 0,5ms1,0ms 1,0 Volt 0,5ms1,0ms 2,5 Volt 0,5ms1,0ms 7,5 Volt Single-chamber unit, rate 60min -1 and 500Ω lead imp. Intrinsic current consumption Stimulation current consumption Exemple of current drain
11
Lead Connectors
12
IS-1 Connector in the Header
13
Tightening the Set Screw
14
Silicone rubber sealing membrane How to find the set screw
15
VS 1 IS -1 Cordis 3.2 mm VS 1B IS -1 Different 3.2 mm Short Bore Connectors
16
IS -1 Cordis 3.2 mm VS 1B IS -1 VS 1A / IS -1 VS 1B / IS -1 Different 3.2 mm Long Bore Connectors
17
VS 1B / IS -1 IS -1 Sensolog* Dialog* Sensorithm* Regency* Paragon* Synchrony* Trilogy* Affinity* Entity Integrity Identity ADx Family * Integrity µ-size Identity µ-size ADx Family µ-size Microny Not µ-size cans * Not M/S version Connectors in our Products
18
Connectors M, 5 mm S, 6 mm K, 3.2 mm (IS-1) T, 3.2 mm (IS-1) Bifurcated 3.2 mm (VDD) C, 3.2 mm inline (VDD) (for old version AddVent, obsolete now)
19
Leads
20
Endocardial Myocardial Epicardial Lead position
21
Bi-polarUni-polar Polarity configuration
22
Advantages - Bi-polar Less risk of myopotential oversensing Less risk of musclestim. Less risk of Crosstalk and Far Field Sensing Advantages - Uni-polar Smaller diameter Larger spikes on ECG With a bipolar lead, you can program the device to e.g. stimulate unipolar but sense bipolar ( ) Uni-polar or bi-polar leads?
23
Far-field sensing Sensing intrinsic activity in the other chamber Crosstalk Sensing pacemaker- activity in the other chamber
25
Bi-polar Uni-polar IS-1 connector
26
Coaxial Coil Design
28
Passive Active Fixation methods
29
Passive and Active Fixation
30
Myocardial lead
31
J-shaped lead (1421T)
32
This fractal coating creates a microscopic area ~4-600 times larger than the macro- scopic area (~15 cm 2 vs 3,2 mm 2 (Isoflex)) Titanium nitride (TiN) coating
33
Lead position Inflammatory reaction Scar tissue Polarisation losses Insulation defects/ Conductor failure Factors affecting threshold
34
Insulation defect Conductor failure Lead defects
35
Reaction to sensing III 0=None T=Triggered I=Inhibited D=Dual (T+I) Rate Modulation IV 0=None R=Rate Modulation What is stimulated? I 0=None A=Atrium V=Ventricle D=Dual (A+V) S=Single (A or V) Multisite Pacing V 0=None A=Atrium V=Ventricle D=Dual (A+V) What is sensed? II 0=None A=Atrium V=Ventricle D=Dual (A+V) S=Single (A or V) The Revised NASPE/BPEG Generic Code for Antibradycardia, Adaptive-Rate, and Multisite Pacing, PACE, Volume 25, No. 2, February 2002 NASPE/BPEG Generic Pacemaker Code