1
Radiation safety in CT
2
RADIATION DOSES DURING CT EXAMS ARE CLASSIFIED AS:
LOW DOSE RADIATION AS COMPARED TO DOSES GENERATED FROM NUCLEAR EVENTS
3
BODY RESPONE TO RADIATION:
SOMATIC EFFECTS GENETIC EFFECTS
4
SOMATIC EFFECT MANIFESTED IN THE INDIVIDUAL EXPOSED
5
ACUTE SOMATIC RADIATION SYNDROME STAGES:
PRODROMAL- NAUSEA, VOMITING, DIARRHEA LATENT PERIOD – NO SYMPTOMS MANIFEST – ORGAN DAMAGE, DEATH OR RECOVERY HEMATOPOIETIC GI CNS
6
MOST COMMON SOMATIC EFFECTS
CATARACT FORMATION SKIN CANCER THYROID CANCER BREAST CANCER LEUKEMIA
7
EFFECT ON AN EMBRYO OR FETUS
MOST PRONOUNCED DURING THE FIRST TRIMESTER!!!!
8
ALTERNATIVE EXAMS SHOULD BE SCHEDULED IF PATIENT IS PREGNANT
CT OR MRI ?
9
FETAL MRI
10
GENETIC EFFECT MANIFESTED IN THE FUTURE GENERATIONS!
11
RADIATION DOSES FROM CT SCANNERS ARE THE HIGHEST IN DIAGNOSTIC RADIOLOGY!!!!
HOWEVER:
12
RADIATION EXPOSURE SHOULD OCCUR :
AFTER CONSIDERATION BY THE PHYSICIAN
13
ALARA
14
RADIOLOGIC UNITS (CUSTOMARY)
EXPOSURE, INTENSITY ROENTGEN (R) ABSORBED DOSE RADIATION ABSORBED DOSE (rad) DOSE EQUIVALENT RADIATION EQUIVALENT MAN (rem)
15
RADIOLOGIC UNITS (SI) EXPOSURE, INTENSITY ------ C/kg
ABSORBED DOSE GRAY (Gy) DOSE EQUIVALENT SIEVERT (Sv) 1 Gy = 100 rad 1 Sv = 100 rem
17
PENCIL IONIZATION CHAMBER METHOD OF MEASURING CT DOSE (CT DOSE INDEX---CTDI)
19
CT PHANTOM & IONIZATION CHAMBER
27
PRINCIPLE OF OPERATION OF IONIZATION CHAMBER
29
THE AMOUNT OF COLLECTED CHARGE IN I. C
THE AMOUNT OF COLLECTED CHARGE IN I.C. IS PROPORTIONAL TO THE AMOUNT OF IONIZATION WHICH IS PROPORTIONAL TO THE AMOUNT OF RADIATION PASSING THROUGH THE CHAMBER. THE TOTAL ELECTRIC CHARGE GENERATED BY AN X-RAY BEAM IS REPRESENTED BY Q AND IS MEASURED IN COULOMBS.
30
IN 1981 CENTER FOR DEVICES AND RADIOLOGICAL HEALTH SUGGESTED AN EASY AND ACCURATE METHOD TO MEASURE PATIENT DOSE IN CT. IT RELIED ON ON CTDI AND MSAD (MULTIPLE SCAN AVERAGE DOSE)
31
MSAD SERIES OF CT SCANS ARE PERFORMED ON A PATIENT . BETWEEN EACH SCAN PATIENT IS MOVED A BED INDEX (BI) DISTANCE.
37
MSAD vs CTDI
39
FACTORS IN PATIENT PROTECTION
TIME DISTANCE SHIELDING
40
SHIELDING IF PERSONNEL PRESENT IN THE ROOM DURING SCAN MUST BE SHIELDED AND STAND FAR FROM THE SOURCE ( INVERSE SQUARE LAW)
41
GONADAL SHIELDS 4-5 CM WITHIN THE X-RAY BEAM
42
CONTACT SHIELDS 360 ° AROUND PATIENT
47
CT DOSE IN SPIRAL SCANNING
CTDI CAN ALSO BE USED PITCH MSAD OR CTDI
48
MORE USEFUL MEASUREMENT IS
EFFECTIVE DOSE TAKES INTO ACCOUNT QUALITY OF RADIATION AND ORGAN SENSITIVITY EXPRESSED IN Sv
50
EFFECTIVE DOSE (mSv) HEAD------------ 2 CHEST ---------- 8
CONVENTIONAL HEAD CHEST ABDOMEN PELVIS SKULL CHEST PA ABDOMEN --- 1 PELVIS
51
DOSE REDUCTION IN CT BED INDEX SPIRAL PITCH COLLIMATION (PREPATIENT)
NOISE DOSE
52
DOSE REDUCTION IN CT SID OID kVp (15 % RULE) mAs DOSE