Biotest Medizintechnik GmbH
Hemoglobin determination Biotest Hemoglobin System Accuracy and precision Biotest Medizintechnik GmbH
2
Products Cuvette Measuring device Biotest Hemoglobin
Biotest Hemoglobin Tester
3
Function Cuvette Biotest Hemoglobin Tester
Blood capillary, reagent and measuring cuvette in one. Biotest Hemoglobin Tester Photometer with integrated control and digital display
4
Function - Cuvette Chemical reaction Haemolysis by sodium desoxycholat
Sodium nitrite Sodium azid Methemo-globin Free Haemoglobin Azid- Methaemo-globin Erythrocytes
5
Function - Cuvette Absorbance curve
6
Measurements
7
Measurements Linearity and precision, n = 20 measurements by each blood sample. y = 1,0064x - 0,0773 R² = 0,9992 5,00 10,00 15,00 20,00 5,0 g/dl 10,0 g/dl 15,0 g/dl 20,0 g/dl Biotest Hemoglobin System, g/dl Mean value HiCN, i g/dl
8
Measurements
9
Function - Tester Photometer; two wave lengths
Evaluation and calculation Reading Evaluation
10
Self calibration 100% green light, 100% red light and dark current
Compared to moment of calibration If necessary automatic adjustment of LED current Evaluation
11
Resolution Max deviation = +/- 0,75 % or +/- 30 steps Absorption
256 g/L 0 = 100 % light 470 – 880 nm 12 bits = 4000 ”steps” Resolution
12
Calibration Absorption 256 g/L
Two points of calibration for linearity; 80 and 220 g/L Three control levels for linearity 120, 150, 180 g/L 0 = 100 % light 470 – 880 nm 12 bits = 4000 ”steps”
13
Function - Tester Calculation of Hb Extinction Turbidity Turbidity
Wave length [nm]
14
Capillary blood Hb Time 0 min 1 min 2 min 3 min
15
Results - Introduction
Accuracy How close to true value is the result Precision How close to each other are repeated measurements with the same sample
16
Results - Introduction
Accuracy and precision High accuracy Low precision Low accuracy High precision High accuracy High precision
17
Results - Introduction
Accuracy: Deviation (mean) in g/L Preferred; deviation in % Precision: Standard deviation SD Preferred; Coefficient of variation, CV %
18
Results Precision with control blood (n=50) Mean value 12,39 g/dl
30 Mean value 12,39 g/dl SD 0,084 g/dl CV 0,68 % 25 20 Numbers 15 10 5 12,1 12,2 12,3 12,4 12,5 12,6 12,7 Result [g/dl]
19
Features Cuvette Aimed for Hb-measurement with Biotest-Hemoglobin-Tester and compatible instruments High accuracy and precision Can be used for venous, capillary and arterial blood Developed to be price worth due to latest production and control technique
20
Features Tester Automatic calibration: no calibration or test cuvette needed Cuvette holder minimizing the risk of contamination Automatic monitoring of power supply with fast loading of batteries (1 hour loading > 40 hours of measuring) Automatic shut off with latest value saved Measures only cuvettes with properly haemolysed blood
21
Summary High accuracy and precision Easy and reliable system
Long time use on battery power
22