MR Driving license course Day 2 – Buttonology
Stefan Skare MR-Centrum, KS
2
Recommended literature
These handouts To be ed to you Basics of MRI ladybird:/usr/local/info/education/basics_of_mri.tar EPI ladybird:/usr/local/info/education/EPI_APPS.PDF
3
Contents The scanner Overview of the interface
Stepping through a few relevant pulse sequences Sending the exam away TPS reset, reboot
4
MR scanner specs MRC1 MRC2 Strength Speed (Slew rate) 8 RF channels
33 mT/m over 45 FOV Speed (Slew rate) 120 T/m/s 8 RF channels MRC2 Zoom mode 40 mT/m over 35 FOV 150 T/m/s Whole body mode 23 mT/m over 45 FOV 75 T/m/s 8 RF channels
5
Dock & undock the bed
8
Move table all the way out
Align on Table in/out Fast or slow Set landmark (define origin = [0,0,0]) Advance to scan (move table to the center of tunnel) Stop table Back to Landmark (hold 2 secs) Move table all the way out (hold 2 secs)
9
Emergency buttons
12
RF coils for head imaging
Standard head coil 8 channel head coil “8HRBRAIN”
13
RF coils for head imaging
Standard head coil 8 channel head coil “8HRBRAIN”
15
Where to find the RF coils
16
fMRI equipment
17
Contents The scanner Overview of the interface
Stepping through a few relevant pulse sequences Sending the exam away TPS reset, reboot
18
Choose the correct coil !
3-plane Axial Sagittal Coronal Oblique Choose the correct coil ! Select 2D, unless 3D-SPGR MRC2 ONLY ! Choose gradient mode (Whole or Zoom) Always Zoom for EPI Choose pulse sequence Select imaging options
19
Contents The scanner Overview of the interface
Stepping through a few relevant pulse sequences Sending the exam away TPS reset, reboot
20
Pulse sequences Localizer 3D SPGR GE-EPI (fMRI) DW-EPI
21
(not necessary for first series)
Localizer Localizer Select FOV of cm Slice thickness e.g. 5 mm 0.0 1 Press Prepare to Scan (not necessary for first series) Press Save Series Select iso-center 1 slice per axis Press Scan
22
3D-SPGR q3dgrass2
23
First prep for scan if manual sync with paradigm
GE-EPI First prep for scan if manual sync with paradigm
24
Image allocation 1024 images / scan
# slices x # volumes < 1024 That’s why we use Brainwave (20,000 images) For diffusion tensor imaging there is also need for > 1024 images E.g. 50 slices x 55 diffusion directions 3,000 images We’ve now have a solution to run DTI via Brainwave (not product)
25
# diffusion directions #RR = #slices/2 or maybe “/3”
DW-EPI > 1024 images # diffusion directions Choose #RR = #slices/2 or maybe “/3” (Depending on heart rate)
26
Contents The scanner Overview of the interface
Stepping through a few relevant pulse sequences Sending the exam away TPS reset, reboot
28
Contents The scanner Overview of the interface
Stepping through a few relevant pulse sequences Sending the exam away TPS reset, reboot
29
Symptoms & Solutions MR system hangs Step1 - Hope that TPS reset works
Takes about 2 minutes You’ll need to make a “New Patient” afterwards. Type “mr” instead of patient ID to pick up the last exam. No new landmark is required. New Localizer is however necessary.
30
Symptoms & Solutions MR system hangs Step 2 - Reboot
Open a terminal window andtype “reboot” Step 3 – Hardware shutdown Instead of “reboot”, type “poweroff” When system is down, go in to the system cabinet in computer room and switch off and on OFF: 12, 11, 10…1 ON: 1,2…12 If you forget to “poweroff” before doing this, I’ll personally kill you with a spoon
31
1 2
33
This position is “ON”