Download presentation
1
Gamma Camera Quality Control
Without adequate quality control, the functional reliability of a nuclear medicine facility is in doubt. Frank P. Dawry Physx.home.comcast.net
2
Quality Control: A Crucial Component In Ensuring Quality Patient Care
Routine conduction of equipment quality control is a necessity for ensuring quality patient care by establishing quality imaging. All instrumentation specific to the operations of a nuclear medicine facility must undergo quality control testing on a regular basis to assure reliability and safety. Quality patient care can only be achieved if optimal diagnostic accuracy is demonstrated through appropriate quality control procedures. According to published guidelines written by the American Society of Nuclear Cardiology (ASNC), the Joint Commission on the Accreditation of Healthcare Organizations (JCAHO), the Society of Nuclear Medicine (SNM), as well as radiation safety guidelines issued by the Nuclear Regulatory Commission (NRC), agreement states and other radiology societies, quality control of equipment must be performed routinely and documented. Frank P. Dawry Physx.home.comcast.net
3
Quality Control testing The routine assessment of an instrument in order to confirm acceptable system performance Part of an overall nuclear medicine quality assurance program Survey meter calibration Dose calibrator QC testing Scintillation (thyroid) probe testing Scintillation camera QC testing Computer software testing Monitor performance testing Staff performance ‘Outcomes’ testing Frank P. Dawry Physx.home.comcast.net
4
Standards NEMA – National Electrical Manufacturers Association
Establishes definitions, quantitative measurements of performance characteristics and reporting techniques for the specification of scintillation camera parameters Spatial linearity Energy resolution Flood field uniformity Spatial resolution (intrinsic and extrinsic) System alignment SPECT reconstructed spatial resolution Whole body spatial resolution Multiple window spatial registration Count rate performance System sensitivity Detector shielding Detector sensitivity System volume sensitivity Performed during acceptance testing prior to first use Verified during total system evaluation annually Frank P. Dawry Physx.home.comcast.net
5
Routine Camera Quality Control procedures
Daily Visual checking the energy spectrum – “peaking” Evaluating the system performance with flood sources – “daily floods” Weekly Checking the spatial resolution with a bar phantom Checking the system linearity with a bar phantom Correcting for system and collimator non-uniformities Performing camera uniformity analysis Monthly Checking the center of rotation for tomographic scans System sensitivity Frank P. Dawry Physx.home.comcast.net
6
Scintillation Camera Performance
Extrinsic, or System measurements – with the collimator attached Give the best indication of clinical performance Intrinsic measurements– with the collimator off Useful for comparing the performance of different cameras because they uncouple the camera performance from the collimator performance Frank P. Dawry Physx.home.comcast.net
7
Camera operation James R Halama, PhD Loyola UniversityMedical Center
Frank P. Dawry Physx.home.comcast.net
8
Camera operation James R Halama, PhD Loyola UniversityMedical Center
Frank P. Dawry Physx.home.comcast.net
9
Camera operation James R Halama, PhD Loyola UniversityMedical Center
Frank P. Dawry Physx.home.comcast.net
10
Camera operation James R Halama, PhD Loyola UniversityMedical Center
Frank P. Dawry Physx.home.comcast.net
11
Camera operation James R Halama, PhD Loyola UniversityMedical Center
Frank P. Dawry Physx.home.comcast.net
12
Frank P. Dawry Physx.home.comcast.net
13
Uniformity The measure of a camera’s response to a uniform irradiation of the camera surface. Intrinsic uniformity is measured using a radionuclide point source (~150 uCi Tc-99m) placed approximately 5 crystal diameters away from an uncollimated camera System or Extrinsic uniformity is measured using a sheet source placed directly on the surface of a collimated camera 5 to 10 millicuries (mCi) of Co-57 or 25 mCi Tc-99m in fillable-water tank. Frank P. Dawry Physx.home.comcast.net
14
Sheet Sources 57Co Sheet Source T1/2 = 270 days 122 keV
10-15 mCi at time of purchase. 99mTc Sheet Source (water filled) T1/2 = 6 hrs. 140 keV 10-15 mCi at time of filling. Frank P. Dawry Physx.home.comcast.net
15
Daily Flood 5 million counts 256 x 256 acquisition matrix
Resulting image should show a uniform distribution across the entire field of view Frank P. Dawry Physx.home.comcast.net
16
Uniformity parameters
UFOV – useful field of view CFOV – center field of view Integral uniformity The difference between the maximum and the minimum pixel value divided by the sum of these two values X 100 = +/- 100 X (Max – Min)/(Max + Min) Differential uniformity The largest difference between any two pixels within a set of 5 contiguous pixels in a row or column (rate of change) = +/- 100 X (Hi – Low)/(Hi + Low) Frank P. Dawry Physx.home.comcast.net
17
Frank P. Dawry Physx.home.comcast.net
18
Spatial Resolution / System Linearity “BARS”
Resolution - the ability of a camera to separately image two objects. If the two objects are seen as a single blurred spot in the image until they are 5mm apart, the resolution is said to be 5mm. Linearity - the ability of a camera to reproduce a straight line. 3 million counts, 256 x 256 matrix for extrinsic 30 million counts 512 x 512 matrix for intrinsic Frank P. Dawry Physx.home.comcast.net
19
Spatial Resolution James R Halama, PhD Loyola UniversityMedical Center
Frank P. Dawry Physx.home.comcast.net
20
Spatial Resolution James R Halama, PhD Loyola UniversityMedical Center
Frank P. Dawry Physx.home.comcast.net
21
Spatial Linearity Deviation from straight line of less than 1
Spatial Linearity Deviation from straight line of less than 1.0 mm for UFOV PLES Phantom Quadrant Bar Phantom James R Halama, PhD Loyola UniversityMedical Center Frank P. Dawry Physx.home.comcast.net
22
Intrinsic Uniformity James R Halama, PhD
Loyola UniversityMedical Center Frank P. Dawry Physx.home.comcast.net
23
Extrinsic Uniformity James R Halama, PhD
Loyola UniversityMedical Center Frank P. Dawry Physx.home.comcast.net
24
Resolution phantoms Moire effect Frank P. Dawry Physx.home.comcast.net
25
UB gamma camera test pattern
One image per detector head is all that's needed to equally and effectively test all quadrants of the gamma camera. Frank P. Dawry Physx.home.comcast.net
26
Uniformity corrections
Correcting for system and collimator non-uniformities due to small variations in sensitivity. Applied to tomographic (SPECT) reconstructed images Acquired for each collimator that is used for SPECT imaging 30 to 120 million count acquisition 256 or 512 matrix size Frank P. Dawry Physx.home.comcast.net
27
Uniformity corrections
James R Halama, PhD Loyola UniversityMedical Center Frank P. Dawry Physx.home.comcast.net
28
Camera uniformity analysis
Similar to a daily flood except greater number of counts acquired ≥30 million counts 256 x 256 matrix Useful for tracking of uniformity trends Frank P. Dawry Physx.home.comcast.net
29
Center of rotation - COR
Used to verify the integrity of the motion of the gantry during tomographic imaging COR value is the difference between the measured COR and the theoretical COR Tc-99m point source either centered in the field of view or off centered Acceptable limits < ±2 mm Frank P. Dawry Physx.home.comcast.net
30
Frank P. Dawry Physx.home.comcast.net
31
Center of rotation - COR
James R Halama, PhD Loyola UniversityMedical Center Frank P. Dawry Physx.home.comcast.net
32
SPECT Phantoms – system performance testing
Jaszczak Frank P. Dawry Physx.home.comcast.net
33
SPECT Phantoms – tomographic system performance testing
Frank P. Dawry Physx.home.comcast.net
34
SPECT Phantoms – tomographic system performance testing
James R Halama, PhD Loyola UniversityMedical Center Frank P. Dawry Physx.home.comcast.net
35
SPECT Phantoms – planar system performance testing
James R Halama, PhD Loyola UniversityMedical Center Frank P. Dawry Physx.home.comcast.net
36
Society of Nuclear Medicine ‘simulator’ phantom program
Lumbar Spine Cardiac Frank P. Dawry Physx.home.comcast.net
37
SPECT phantom Frank P. Dawry Physx.home.comcast.net
38
ICANL INTERSOCIETAL COMMISSION FOR THE ACCREDITATION OF NUCLEAR LABORATORIES
Frank P. Dawry Physx.home.comcast.net
39
PET/CT TEST (PET ONLY) Blank scan Normalization Absolute activity calibration Preventive maintenance FREQUENCY Daily After a hardware change or per manufacturer recommendations* Every 6 months, or per manufacturer recommendations* plus consideration for the CT portion and the merging of the two imaging modalities. Frank P. Dawry Physx.home.comcast.net
40
SPECT/CT in review Similar to SPECT with consideration for CT portion and the merging of the two imaging modalities. Frank P. Dawry Physx.home.comcast.net
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.