W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Flexibility and Precision in Peripheral Arteries 0.035” OTW Self- Expanding Stent System
2
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only The Reality of Femoral Occlusions
3
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only In Summary Range of diameter Range of length Usable Catheter length Guide wire compatibility Introducer compatibility Construction & design Entry profile: On-stent Tantalum markers Ø 6, 7, 8, 9, 10, 12 mm 20, 30, 40, 60, 80, 120 mm 80 cm / 120 cm 0.035”6F multisegmented design with reinforced link-elements 0.043” 1 @ each end
4
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only The Stent
5
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Basics on NiTiNOL NiTiNOL was invented 1962 by the US-Naval Ordnance Laboratory, a department of the US-Navy. Nickel Titanium Naval Ordnance Laboratory They were looking for a very hard, anti-magnetic and corrosion resistant material They melted 45% Titanium and 55% Nickel, which developed the desired properties and even more. NiTiNOL is a brand name, there are other NiTi-alloys with other names like Sentinol, Sentalloy,...
6
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only What is Superelasticity?
7
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Properties of NiTiNOL Superelasticity In reality it is pseudo-elastic if it is cooled down below 20°C, it becomes inelastic like pure Tin (Zn) so that it can be deformed in all facilities without breaking (Martensitic low temperature phase). If it is rewarmed above 20°C, it moves back in its previous shape. Over 20°C it can be bended and twisted as desired and immediately it moves back to the old shape (Austenitic high temperature phase). Depending on the Alloy ingredients, the conversion temperature can vary from -15°C to +80°C.
8
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Shape NiTiNOL The material needs to be fixtured and constrained in the desired shape and heat-treated. Typically for superelastic material, a heat treatment in the 500°C range is adequate; the length of heat treatment varies with the equipment used for the heat treatment and the thermal mass of the shaping fixture. In a molten salt bath for example, the heat treatment time is generally between 2 and 5 minutes.
9
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Stent Design Highly flexible and conformable stent platform Laser cut, ultra-polished surfaces Multisegmented design well proven zigzag-Nitinol design oversized connecting links for improved fracture resistance Surface Quality Ultra high polished surface Increased corrosion resistance
10
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Stent Marker Two Tantalum markers (one on each end of the stent) for enhanced visibility High safety, because of Laser-welded markers Unique inside oriented marker, for improved mechanical properties Good radiopacity of the stent Strut thickness: 0.18 mm Strut width: 0.13 mm
11
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Mechanical Stent Behaviour Flexure Thin struts for high flexibilityThin struts for high flexibility Expansion Large strut width for sufficient radial strengthLarge strut width for sufficient radial strength Torsion Oversized connecting links for improved fracture resistanceOversized connecting links for improved fracture resistance Best Adaptability to the Vessel Dynamic Requirements Elongation Adequate strut dimensions, number of links and design for dynamic adaptabilityAdequate strut dimensions, number of links and design for dynamic adaptability
12
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Diameter dependent number of links Number of link elements: 3 link-elements Ø 6, 7 and 8 mm 5 link-elements Ø 9, 10 and 12 mm Optimized radial strength and response to different mechanical stress by the diameter specific design Oversized connecting links for improved fracture resistance. 1 2 31 2 3
13
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Shortening – a technical answer Shortening is dependent from the nominal length & diameter of the stent! When the stent is loaded in the delivery system, it is longer than the nominal length! Since the proximal marker on the SDS is movable, the stent edges are clearly marked to enable a proper positioning in the lesion. Even if the stent follows the sheath retraction for some mm, the distal markers stays always behind it. We have measured a shortening of 3,9% in the worst case
14
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only The Wallstent Story This mechanical behaviour is HISTORY!
15
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Metal-to Artery Ratio small large
16
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Model Mix Stent Ø 6.0 mm length 20, 30, 40, 60, 80, 100, 120 mm Stent Ø 7.0 mm length 20, 30, 40, 60, 80, 100, 120 mm Stent Ø 8.0 mm length 30, 40, 60, 80, 100, 120 mm Stent Ø 9.0 mm length 30, 40, 60, 80, 100 mm Stent Ø 10.0 mm length 30, 40, 60, 80, 100 mm Stent Ø 12.0 mm length 30, 40, 60, 80 mm Usable Shaft Length: 80 cm and 120 cm Stent Ø := unconstrained stent diameter length := nominal length @ unconstrained diameter
17
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only The Stent Delivery System
18
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only SDS Construction OTW 0.035” guide wire compatible Standard guidewire size, 0.018” is more exotic 6F shaft General accepted size for peripheral interventions 80 and 120 cm usable shaft length Suitable for antegrade or contralateral approaches Delivery by holding the stainless steel tube in place and sliding back the Y-piece with Tuohy-Borst Valve Straight forward technique Tuohy-Borst Valve, better known as variable hemostatic valve
19
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Radiopaque Markers on SDS Proximal marker (1) Proximal marker acts as the pusher to maintain the stent in place when the outer tube is retracted and the stent is deployed. Distal marker is integrated in the outer tube (2) Tip material is radiopaque (mixed with Barium Sulfate) (3) (1)(2)(3)
20
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only SDS Sheath Construction Full length braided sheath design This reinforcement combines pushability and flexibility to enable easy negotiation in tortuous anatomies without kinking ideal balance for successful contra-lateral approaches PTFE coating inside the sheath Reduction of friction between sheath and stent for a gradual, controlled and safe deployment of the stent in the lesion
21
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Tip Construction / Entry Profile New Tip design entry profile of 0,043 “ (1.10 mm), Tip composed of a corrugated tube and a radiopaque overmoulded tube NEW: edges were rounded to avoid entrapment during retrieval NEW: security seal between white overmoulded tip and blue inner tube.
22
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only The Pull-Back-Handle We have chosen the KISS approach! Keep It Simple & Safe Delivery by sliding back the Y-piece with Tuohy-Borst Valve while holding the stainless steel tube in place After flushing the inner lumen of the sheath through the side port… With one hand keep the steel tube in position, with the other hand slide back the Y-piece It is not a “Black-Box” Handle You see - what you get/do!
23
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Stent Positioning Controlled, precise positioning minimal shortening to obtain an adequate lesion coverage PTFE inner shaft coating for low-friction stent deployment low compliant core-tube for direct transition from pull-back handle to sheath retraction
24
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Product Features Excellent Flexibility of the Stent Platform and the Delivery System Inside shaft Teflon-liner for low friction deployment Optimal balance of flexibility, radial force and strut fracture resistance Diameter dependent design Ø 6, 7, 8 mm - 3 links Ø 9, 10, 12 mm - 5 links Atraumatic, kink-resistant tip design Braided outer catheter shaft with Pull-Back Handle
25
W e m a k e i d e a s c o m e a l i v e 001IMP120402 For internal use only Clinical Data StudyMARISABSOLTERESILENTFAST StentMarisDynalink+AbsoluteLifeStentLuminexx Mean stented length (mm) 119,8 ± 83,31132±7199.2±49.9 45 (lesion length) 12 month primary patency 81.4% (X-ray)63% (DUS)80% (DUS)68% (DUS)