Readout Electronics Alexander Karakash & Natalia Kondrat’eva MEPHI for DESY
2
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 2 Outline 1.Measurement of main characteristics of ASIC’s PreAmp versus temperature. 1.Measurement of characteristics DAC
3
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 3 Experimental setup for measurement for ASIC linearity i
4
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 4 Calibration mode [DAC=OFF, G=0,1pF, T=min, R(10k)=OFF; C(inj)=68pF] Measurement of linearity
5
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 5 Calibration mode
6
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 6 Calibration mode Noise is independent on temperature G/ T = -0,27%/ o C With increasing of T the input and output range is increased too Peak position does not depend on T Pedestal depend on T ~ 2,5mV/ o C observed sharp shift of pedestal with sharp change of T When door of thermostat was closed and thermostat was turn off, temperature around ASIC was higher then outside of thermostat on +5 - +8 o C Reference component depends on temperature. It has small heat capacity
7
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 7 Physics mode [DAC=OFF, G=0,5pF, T=max, R(10k)=ON; C(inj)=68pF] Measurement of linearity
8
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 8 Physics mode
9
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 9 Physics mode Noise is independent on temperature G/ T = -0,11%/ o C With increasing of T the input and output range is decrease Peak position does not depend on T up to 50 C Pedestal depend on T ~ 1,5mV/ o C Using calibration input of ASIC does not give correct results for dependency gain from temperature.
10
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 10 Experimental setup for measurement of DAC Setting of ASIC: DAC=ON Calibration mode: G=0,1pF T=min R(10k)=OFF
11
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 11 Measurements of DAC With increasing of DAC’s setting noise sharp increases after the 200 code It was observed effect – jump setting voltage for some setting of DAC. Jump of voltage was approximately 40 mV (2 bits) Gain is independent on DAC’ setting practically T=25 o C
12
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 12 Dependencyof DAC voltage and noise from temperature Setting voltage depends on temperature V/ T~3mV/ o C Noise depends on temperature ~ +2% In thermostat noise was smaller then outside It is necessary: to protect HAB from pickup noise; to use correct schematic solutions for reject pickup noise
13
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 13 Regime auto generation of DAC DAC’s auto generation effect was observed. This effect appeared when DAC=ON only. This effect may be associated with evaluation board? But …
14
24/03/2005Alexander Karakash & Natalia Kondrat'eva MEPhI for DESY 14 Summary & Questions & Remarks ASIC’s PreAmp & ASIC’s DAC have good thermal stability in both mode. But it is still necessary to stabilize temperature around HAB DAC’s max working range is limited by its noise. It is necessary: – to protect HAB from pickup noise; – to use correct schematic solutions to reject pickup noise. The next version of ASIC must have auto generation suppression.