Download presentation
Presentation is loading. Please wait.
1
QUALITY CONTROL For the Technologist
2
TECHNOLOGIST RESPONSIBILITIES
DAILY Darkroom Cleanliness Sensitometry WEEKLY Screen Cleanliness Viewbox cleanliness Phantom Imaging MONTHLY Visual Checklist QUARTERLY Repeat/Reject Analysis Fixer Retention SEMIANNUALLY Compression Film/Screen Contact Darkroom Fog
3
DARKROOM CLEANLINESS NEVER: ALWAYS: Carpet a darkroom
Store boxes or cartons and never open ALWAYS: Have shelving with doors Replace filters and service air conditioning Wipe countertops with damp cloth Wipe film trays with anti-static solution Use a black light
4
SENSITOMETRY Must use same box of film
Must be performed before clinical films Check Developer temperature Flash sensitometric strip on emulsion side Speed and Contrast Index within +/-.15 B+F within +/-.03 Developer within +/-.5 degrees An increase in developer temp will affect other values: increase in B+F and decrease in density difference. If developer normal, those changes may indicate a need to change chemicals.
5
SENSITOMETRY When data falls outside of the limits Trends
Documentation
6
SCREEN CLEANLINESS Clean when 3-4 specks shown Use Antistatic solution
4x4 rayon or lintless 100% cotton pads Wipe in circular motion from center out Let dry Use one of many tools available to get rid of dust Check with black light Discard old film (costly) Load with new film
7
VIEWBOXES Clean weekly to remove dust, smudges, etc
Change bulbs every 12-18months Test luminance with light meter (not required by MQSA) When changing bulbs, clean inside of viewboxes and change all bulbs
8
PHANTOM IMAGING Monitors whole system
Simulates a 4.5 cm compressed breast Consists of 6 nylon fibers, 5 oxide specks, 5 tumor masses Should be performed: After any service After changes in chemistry or film Whenever changes in image quality suspected
9
PHANTOM PROCEDURE AEC placed same as a 4.5 cm breast
4mm acrylic disc placed away from objects Background density: center, 1.2 OD +/-.2 Density difference: .4 +/-.05 for films at 28kVp Chart mAs to evaluate reproducibility (should not vary more than +/-.15)
10
PHANTOM ANALYSIS Must see 4 fibers, 3 specks, 3 masses (10)
Can vary +/-1 but no lower than 10 objects. Viewed: By same person Same time of day On same viewbox Under same viewing conditions Using same type of magnifying lens
13
VISUAL CHECKLIST Takes inventory of room to ensure safety, comfort and convenience
14
FIXER RETENTION Performed right after daily strip on emulsion side in clear area Place drop of fixer and let sit 2 minutes (do not expose to bright light) Blot dry and place on white cardboard Compare to chart. No more than 5ug/cm squared (3 on the chart)
15
REASONS FOR ELEVATED FIXER
Inadequate wash water flow rate Partially depleted fixer (enough activity for clearing but not enough for hardening) Insufficiently hardened emulsion retains too much fixer for wash to remove. RESULT: Degrades stability of image
16
FIXER CORRECTION Increase fixer replenishment Increase water flow
17
REPEAT/REJECT ANALYSIS
To evaluate reasons films are repeated Points out insufficiencies, waste Can improve efficiency
18
REPEAT ANALYSIS PROCEDURE
Quarterly and at least 250 patients examined Films are categorized ACR states number of repeats should be no greater than 5%.
19
WAKE UP! ITS OVER!
20
COMPRESSION Image quality and safety test
Test in both power drive and manual MQSA states 25 – 47 lb range
21
How does compression improve quality?
Decrease scatter Decrease dose Decrease patient motion Increase contrast Increase uniformity of breast tissue Increase image sharpness by decreasing thickness thus decrease focal spot degradation
22
COMPRESSION PROCEDURE
Place towel Place scale Place friend’s (soon to be enemy’s) breast on top of scale Compress in power mode until it stops (never mind the screaming of your new enemy) Compress in manual until it stops (same as above)
23
COMPRESSION PROCEDURE
Place towel Place scale Place several towels on top of scale Compress in power until it stops Compress in manual until it stops
24
FILM/SCREEN CONTACT Performed on all new cassettes and/or semi-annual
More critical than general radiography Highlights areas of decreased sharpness Fine copper wire mesh screen used More critical because of its high resolution which is attained through single emulsion with small crystals and screens with small phosphors Damage to screens not noticeable with clinical films so this test highlights areas of decreased sharpness.
25
FILM/SCREEN CONTACT PROCEDURE
Clean cassettes Let sit for 15 minutes Employ 25-28kVp, no grid, manual mas with .5 seconds Paddle all the way up with acrylic Wire mesh on top of screen
26
FILM/SCREEN ANALYSIS .7-.8OD near chest wall Stand 3-6ft back
If densities over 1 cm appear, clean and retest. Remove if large density still appears. Several small densities acceptable.
27
DARKROOM FOG Inspect with all lights off
Be aware of luminescent or phosphorescent items Use radiation exposed film MQSA states fog should be no >.05 OD when exposed for 2 min. Problems usually lie in safelight position or filters
28
DARKROOM PROCEDURE Load cassette in total darkness
Expose phantom to 1.0 OD Inspect safelights Turn off lights, inspect Remove film, insert into template emulsion up Expose sections to safelights Run film then measure densities on each side of lines
30
Safelights Filters should correspond with green light sensitive film
Lamps ~4 ft from work area Wratten 1 or 2, Kodak GBX-2 good examples of filters No >15watt bulb in overhead fixture No >7.5 watt in closer fixtures Printing on filter should face out Change filters every 1-2 years Heat from higher watt bulbs could ruin filters Film should be stored away from radiation, heat, chemical fumes.
31
AIR QUALITY Temperature: 70 degrees F Humidity: 30-50%
Ventilation: for image quality and technologists’ health When film is exposed to excessive heat, its emulsion softens and is more susceptible to scratching. A cooler temp causes emulsion to crack and peel. If air dry <30%, static marks appear. If humidity above 50%, water droplets from air may cling to film and cause emulsion to clump. Image will appear as if misted with ink. Ventilation of processor: if no air going in or coming out (remove hose and put hand over hole) then streaking and mottling of emulsion can occur. Can give techs headaches and nausea due to chemical fume buildup.
33
PROCESSOR MAINTENANCE
Developer temperature Increase can cause fog Decrease can result in: Decrease in film speed and contrast Higher dose to pt Fixer temperature Decrease can affect archival of film Wash-water temperature Can cause biological growth in tank Can affect developer and fixer temps, improper wash
34
PROCESSING CONTINUED…
Chemical replenishment Rates inconsistent when processor left on and not used Check solution levels If solution levels too low, sediment could get in to the processor and cause damage. Would need to empty tanks and clean.
35
PROCESSING: STANDARD Typically used for general radiography
Faster cycle time (usually 90sec) Developer immersion ~23 sec Can accommodate single and double
36
PROCESSING: EXTENDED Difference from standard is developer immersion time (~45sec) “Push” more contrast and speed from the film Less exposure needed with increase in film speed Increase in tube life Downfall is an increase in noise
37
TYPES OF PROCESSORS Small desktop DISADVANTAGES ADVANTAGES
Straight-through require careful adjustments to replenishment Developer not sprayed evenly ADVANTAGES Cost, small, fewer rollers, two developer stations, Less chance of oxidation
38
TYPES OF PROCESSORS Large stand-alone DISADVANTAGES ADVANTAGES
Artifacts from rollers Space ADVANTAGES Agitates chemicals which increases supply of fresh chemicals
39
ARTIFACTS Processing Environmental Handling Positioning Equipment
White artifacts indicate pressure on emulsion before exposure
40
SENSITOMETRIC STRIP Establishing the baseline
Expose and process 5 strips Average readings for temporary base Process one strip per day for 5 days If strip is out of tolerance, correct and repeat Watch for trend due to seasoning effect Average these 5 strips for new base Baseline maintained unless film, chemistry or processing altered
41
CROSSOVER Performed when changing box of film
10-12 sheets remaining in old box Do not perform after processor cleaning
42
CROSSOVER PROCEDURE Perform in total darkness Expose 5 old and 5 new
Alternate old and new Distinguish old from new Measure steps used in daily QC Average 5 readings from each box Subtract old from new Add/Subtract results to/from original aim values
43
CROSSOVER RESULTS Results must be within +/- .10
If greater than .15 difference, contact film manufacturer representative
44
(Dudes Prefer) (Sex, Violence and Partying) over (Visiting) (Relatives For) (Christmas Day Feasts.)
46
QUALITY ASSURANCE Defined as all those planned and systematic actions necessary to provide high image quality.
47
QUALITY MANAGEMENT Part of the management function which determines and controls quality policy. Making appointments Call back protocol Obtaining previous mammograms Patient education Patient intake Problem solving
48
QUALITY CONTROL The set of operations necessary to maintain or improve quality.
49
QUALITY SYSTEM Assignment of responsibility Establishment of standards
Staff training Proper documentation of procedures QC of equipment at installation and throughout its life
50
MQSA of 1992 As of Oct 1, 1994, all mammography facilities in the U.S. (except those of Dept of Veterans Affairs) were required to be certified by the FDA as meeting mammography quality in order to lawfully continue to perform Mammography. Goal: to assure that Mammography is safe and reliable to allow detection of breast CA in its earliest most treatable stages.
51
MQSA REGULATIONS To operate lawfully, a mammography facility must be certified by the FDA. In order to be certified, a facility must first be accredited by a federally-approved private nonprofit or state accreditation body. To be accredited, the facility must undergo periodic review of clinical images; surveyed annually by a medical physicist; meet standards for personnel qualifications, equipment QA programs, and record keeping and reporting. Facility must undergo annual inspections.
52
ACCREDITATION A survey form completed by facility
Phantom images submitted to evaluate image quality Dosimeter visible on phantom and dose measured Clinical images submitted for evaluation Once initial paperwork approved, facility submits 30 days of processor QC strips CLINICAL IMAGES One set of fatty breasts and 1 set of dense breast images evaluated. Must demonstrate adequate positioning, compression, exposure level, contrast, exam identification, etc.
53
CERTIFICATION Awarded by the FDA as providing quality mammography services. This is awarded upon accreditation by a federally-approved accreditation body: ACR; States of Arkansas, California, and Iowa
55
RESPONSIBLE INDIVIDUALS
TECHNOLOGIST RADIOLOGIST PHYSICIST
56
Interpreting Physicians Initial Qualifications
State license Diagnostic radiology certification OR 3 months of formal training 60 hours of category I CME in mammography 240+ mammograms interpreted under direct supervision In the 6 months immediately prior to independent interpretation OR If board certified at first allowable time, within the last 2 years of residency
57
Interpreting Physicians Continuing Requirements
Continuing experience Interpret at least 960 examinations / 24 months Continuing education At least 15 category I CME credits / 36 months At least 6 of the 15 CME units must be in each mammographic modality used 8 hours of training in each mammographic modality before independent use
58
Radiologic Technologists Initial Qualifications
State license OR certified by FDA-approved body 40 hours of documented mammography training Performing at least 25 examinations under direct supervision 8 hours of training in each mammographic modality to be used
59
Radiologic Technologists Continuing Requirements
Continuing experience At least 200 examinations / 24 months Continuing education At least 15 CEUs / 36 months At least 6 of the 15 CEUs must be in each mammographic modality used 8 hours of training in each mammographic modality used before independent use
60
Radiologic Technologists Reestablishing Qualifications
Continuing experience Complete 25 examinations under direct supervision Continuing education Bring total continuing education credits up to 15 CEUs / 36 months
61
Medical Physicists Initial Requirements
Licensed or approved by a State OR Certified by FDA-approved body Master’s or higher degree in a physical science 20 semester hours of physics 20 hours of documented mammography survey training One facility and 10 units surveyed 8 hours of training in each modality surveyed before surveying independently
62
Medical Physicists Alternative Initial Requirements*
Licensed or approved by a State OR Certified by FDA-approved body Maintained active status For physicists who were qualified under the interim regulations
63
Medical Physicists Alternative Initial Requirements* (cont.)
Prior April 28, 1999, have: Bachelor’s degree in physical science with at least 10 semester hours of physics 40 hours of documented mammography survey training after earning the bachelor’s degree One facility and 20 units surveyed after earning the bachelor’s degree
64
Medical Physicists Continuing Requirements
Continuing experience Survey at least 2 facilities and 6 units / 24 months Continuing education At least 15 CEUs / 36 months 8 hours of training in any modality before surveying independently
66
MANUAL QC MANUAL QA MANUAL
67
QA MANUAL Details of acceptance
Personnel qualifications and documentations. Department policy Safety standards Biohazards
68
QC MANUAL The QC manual is pertinent to ease the inspection process. It lists: Test procedures Test frequencies Test standards Unacceptable results as listed as well as the corrections made for them.
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.