Image Processing for Interventional MRI Derek Hill Professor of Medical Imaging Sciences King’s College London
2
Image Processing for Interventional MRI Derek Hill Professor of Medical Imaging Sciences University College London
3
The team Kawal Rhode Marc Miquel Redha Berboutkah David Atkinson Maxime Sermesant Rado Andriantsimiavona Kate McLeish Sebastian Kozerke Reza Razavi Vivek Muthurangu Sanjeet Hegde Jas Gill Pier Lambraise Cliff Bucknall Eric Rosenthal Shaqueel Qureshi
4
Context Interventional MRI provides particular opportunities and challenges for image analysis. –Hostile environment for computers –“real time” requirements –Link between acquisition and analysis
5
Overview Background to XMR guided interventions Integrating x-ray and MRI Automatic cathether tracking Integration of image analysis in acquisition
6
XMR X-ray + cylindrical bore MRI in the same room Becoming main platform for MR guided interventions –Resection control in neurosurgery –Endovascular procedures Not ideal for percutaneous procedures
7
Patient Staff XMR suite at Guy’s (funded of JREI, Philips Medical Systems and Charitable Foundation of Guy’s & St Thomas’)
8
XMR System at Guy’s Hospital XMR = hybrid X-ray/MR imaging Common sliding patient table Provides path to MR-guided intervention
9
XMR suite at Guy’s
10
Catheter manipulation
11
Visualizing catheters Fast imaging (70 msec per frame) –TE = 1.3, TR = 2.6 –SSFP sequence (balanced TFE) –Acquisition: 78 x 96, 80% FOV, 80% acq, SENSE factor 2 (ie: only 25 phase encodes!) Carbon dioxide filled balloon as contrast agent
12
Catheter Manipulation Images acquired with standard Philips real time or interactive sequences
13
Catheter Manipulation Miquel et al. Visualization and tracking of an inflatable balloon catheter using SSFP in a flow phantom and in the heart and great vessels of patients. Magn Reson. Med. 51(5):988-95 2004
15
Integrating x-ray and MRI XMR provide rapid transfer between modalities No capability to integrate the images X-ray and MRI provide complementary information Combined x-ray and MR has value in complex interventions eg: electrophysiology
17
T M1M1 Registration Matrix Calculation Overall registration transform is composed of a series of stages Calibration + tracking during intervention M2M2 M3M3 3D Image Space X-ray Table Space X-ray C-arm Space 2D Image Space Scanner Space R*P
18
XMR Registration: Software Overview
19
XMR Registration: Calibration Acrylic calibration object with 14 markers Interchangeable caps for MR and X-ray imaging Determine geometric relationship between MR and X-ray system Determine X-ray projection geometry MR X-ray
20
Calibration (1) Fixing flange for sliding table. (2) Saline-filled acrylic half cylinder with 20 divot cap markers in a helical arrangement. (3) Slot in acrylic base plate to allow sliding of half cylinder. (4) & (5) End stops. (6) Fixing to allow MR surface coil attachment
21
XMR Registration: MR Overlay on X-Ray
22
XMR Registration: 3D Reconstruction
23
XMR Registration: Phantom Validation T1-weighted MR volume + 5 pairs of tracked x-ray images using calibration object as a phantom 2D RMS Error = 4.2mm (n=35), Range = 1.4 to 8.0 mm 3D RMS Error = 4.6mm (n=17), Range = 1.7 to 9.0 mm “Registration and Tracking to Integrate X- ray and MR Images in an XMR Facility “, Rhode et al, TMI, Nov 2003.
25
Clinical Example Patient undergoing electrophysiology study prior to RF ablation of heart rhythm abnormality
26
MR Imaging - Anatomy SSFP three- dimensional multiphase sequence 5 phases 256x256 matrix 152 slices resolution=1.33 x 1.33 x 1.4 mm TR=3.0 ms TE=1.4 ms flip angle=45
27
MR Imaging - Motion SPAMM tagged imaging sequence 59 phases SA & 50 phases LA 256x256 matrix 11 slices SA & 4 slices LA resolution=1.33 x 1.33 x 8.0 mm TR=11.0 ms TE=3.5 ms flip angle=13 tag spacing=8 mm
28
X-ray Imaging + Electrical Mapping LAO ViewAP View Contact electrical mapping system Constellation catheter (Boston Scientific)
29
MR Anatomy Overlay
30
Catheter Reconstruction
31
Refining the Registration Errors due to limitations of registration technique and patient motion Basket point cloud meshed Rigid surface-to-image registration used to realign the basket mesh
32
Visualising the Electrical Data Cycle 1 - normal Cycle 2 - ectopic
33
Instantiation of model
34
Simulation results LV volume
35
Catheters re-visited Essential properties of catheters –Clearly visible –Safe mechanically electrically Magnetically Desirable properties –Automatic localization –Tip and length visible CO 2 filled balloon catheters are safe Tip location ambiguous Length not visible Cannot be localized automatically
36
Is there an image analysis solution? Find catheter automatically in modulus image? Is it easier to find in a phase image?
37
Better solution: change nucleus Fluorine is not present in body High NMR sensitivity Safe blood subsitutes available (eg: PFOB)
38
Catheter tracking(a)(b)(c) SSFP proton image plus fluorine projections Phantom setup
39
Catheter tracking(a)(b)(c) Automatic superposition Of catheter tip on proton image Phantom setup
40
Lumen visible
41
Dynamic scan
42
Catheter Tracking and Visualization Using 19 F Nuclear Magnetic Resonance Sebastian Kozerke 1,2, Sanjeet Hegde 3, Tobias Schaeffter 4, Rolf Lamerichs 5, Reza Razavi 3, Derek L. Hill 2 Magn. Reson. Med. 2004 (in press)
43
Image analysis combined with acquisition Real time MRI can provide high temporal resolution, but low quality Can we subsequently combine real time images to generate high image quality?
44
Real time MRI with slice tracking Real time undersampled radial acquisitions Navigator Slice tracking
45
Registration to compensate for motion Rigid body then non-rigid registration to correct motion During scanning
46
Demonstration on gated volunteer heart images (n=4) Undersampled images
47
Demonstration on gated volunteer heart images (n=4) Combined with no registration
48
Demonstration on gated volunteer heart images (n=4) Combined with rigid registration
49
Demonstration on gated volunteer heart images (n=4) Combined with rigid then non-rigid registration
51
Conclusions Interventional MRI is fertile area for image analysis Real time requirements New applications (eg: RF ablation) Improving guidance Novel acquisition and reconstruction incorporating image analysis