Ionization Chamber Array for External Beam Radiotherapy
Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014
2
Project goal A short summary Absolute dose measurement
MV X-Rays ~100÷230MeV protons Absolute dose measurement Machine quality assurance Patient quality assurance
3
Outline Experimental Activity Characterization at IBA DosLab
Characterization at Klinikum rechts der Isar & University of California (San Francisco)-Radiation Oncology dep. Characterization at Proton Therapy Center Czech s.r.o. Ongoing activities and future work ARDENT project: trainings, conferences, secondments
4
Experimental Activity
Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014
5
Detector evolution 80 ionization chambers, 3.5mm pitch First Prototype
Measurement setup 80 ionization chambers, 3.5mm pitch Second Prototype
6
Characterization at IBA DosLab
Prototype optimization Main debug issue: time stability of the chamber current. Signal (normalized) Time (s) Example: two 60Co irradiations (detector biased at 100V). Current ~ 10pA Signal (normalized) Time (s) Signal about 8pA Optimization of signal routing led to a better time stability of the sensor: Pretty good stability in time of the signal Prompt rise and fall time 60Co, 150V bias, Tint=0.8s Sensitivity 200pC/Gy Fluctuation 0.3% (st.dev.) -20 40 60
7
Characterization at IBA DosLab
Charge collection efficiency Charge collection efficiency higher than 99% already at 150V (99.5% ± 0.3% at 250V)
8
Characterization at IBA DosLab
Sensitivity dependence on dose per pulse ± 0.8% sensitivity change on dose per pulse in the range 0.09÷2.67mGy/pulse
9
Characterization at IBA DosLab
Tissue to phantom ratio measurement Very good agreement with PPC05 (better than 0.5%) at every depth of measurement
10
Characterization at IBA DosLab
Elekta Agility Output Factors Good agreement with reference for small fields, reduced sensitivity with large fields -3.7% -2.4% a:Si flat panel, a prototype optimized for dosimetry
11
Characterization at IBA DosLab
Beam profiles (photons)
12
Characterization at MRI
Machine QA: virtual wedges, MLC QA, dose in depth (see pics) MapCheck2 (SunNuclear) diode array Varian Clinac DHX Secondment, 22nd April - 3rd May, Klinikum rechts der Isar (Munich)
13
Characterization at MRI
Example of patient QA EBT3 Varian Trilogy, Millennium MLC, VMAT delivery technique
14
Characterization at UCSF
Clinical tests & clinical training Secondment, 28th July - 8th August, UCSF department of Radiation Oncology; Detector characterization through clinical test; Clinical training: Machine QA; Introductory training to Pinnacle TPS; Use of gafchromic films; Joined the daily activity of medical physicists.
15
Characterization at UCSF
Example of small field dosimetry and linac commissioning 1x1cm2 Varian TrueBeam Virtual wedge 45°
16
Characterization at UCSF
Example of linac commissioning measurement Water phantom as reference
17
Characterization at PTC czech s.r.o.
Dose linearity PBS mode only 1ms pulse duration, 10ms pulse period, 3.5mm spot at 226MeV 0.02-2MU/spot 33x17 spot map Output norm. to 1MU/spot -0.04% 1.7% -0.9% 0.8% -0.1% -0.05% 0.01% PPC05 signal (nC) IC array signal (pC) Ref detector: PPC05
18
Characterization at PTC czech s.r.o.
Beam profiles (protons) Lynx: a scintillator based detector for PBS mode machine QA
19
Ongoing activities/Future work
Long term stability measurements with 60Co Further test with PBS proton beams; Paper about the characterization of the 1D IC array prototype; Extend the technology to a 2D detector - components drawing already started, first prototype assembling foreseen for the end of the year.
20
Trainings, conferences, secondments…
Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014
21
Secondments Upcoming:
Klinikum rechts der Isar, Munich ( ) UCSF RadOnc dep., S.Francisco ( ) Upcoming: B&A secondment at STMicroelectronics (Catania, July 2015 tentative) UCSF (early 2015 tentative)
22
Conferences, trainings…
AAPM, Austin (TX), Poster presentation: “An Ionization Chamber Array with High Spatial Resolution for External Beam Radiotherapy”, M.Togno, D.Menichelli, J.J.Wilkens. Upcoming: AAPM 2015, … Trainings: Clinical trainings at MRI & UCSF (already mentioned); Radiation protection training (IBA Dosimetry); Training on SBRT radiotherapy technique (Klinikum rechts der Isar); ARDENT B&A training (CERN): Technology Transfer, Management, Entrepreneurship. PhD activities: already spent more than 2 weeks at the University to teach students at physics labs.
23
Thank you! Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014