1 Ionization Chamber Array for External Beam Radiotherapy
Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014

2 Project goal A short summary Absolute dose measurement
MV X-Rays ~100÷230MeV protons Absolute dose measurement Machine quality assurance Patient quality assurance

3 Outline Experimental Activity Characterization at IBA DosLab
Characterization at Klinikum rechts der Isar & University of California (San Francisco)-Radiation Oncology dep. Characterization at Proton Therapy Center Czech s.r.o. Ongoing activities and future work ARDENT project: trainings, conferences, secondments

4 Experimental Activity
Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014

5 Detector evolution 80 ionization chambers, 3.5mm pitch First Prototype
Measurement setup 80 ionization chambers, 3.5mm pitch Second Prototype

6 Characterization at IBA DosLab
Prototype optimization Main debug issue: time stability of the chamber current. Signal (normalized) Time (s) Example: two 60Co irradiations (detector biased at 100V). Current ~ 10pA Signal (normalized) Time (s) Signal about 8pA Optimization of signal routing led to a better time stability of the sensor: Pretty good stability in time of the signal Prompt rise and fall time 60Co, 150V bias, Tint=0.8s Sensitivity  200pC/Gy Fluctuation  0.3% (st.dev.) -20 40 60

7 Characterization at IBA DosLab
Charge collection efficiency Charge collection efficiency higher than 99% already at 150V (99.5% ± 0.3% at 250V)

8 Characterization at IBA DosLab
Sensitivity dependence on dose per pulse ± 0.8% sensitivity change on dose per pulse in the range 0.09÷2.67mGy/pulse

9 Characterization at IBA DosLab
Tissue to phantom ratio measurement Very good agreement with PPC05 (better than 0.5%) at every depth of measurement

10 Characterization at IBA DosLab
Elekta Agility Output Factors Good agreement with reference for small fields, reduced sensitivity with large fields -3.7% -2.4% a:Si flat panel, a prototype optimized for dosimetry

11 Characterization at IBA DosLab
Beam profiles (photons)

12 Characterization at MRI
Machine QA: virtual wedges, MLC QA, dose in depth (see pics) MapCheck2 (SunNuclear) diode array Varian Clinac DHX Secondment, 22nd April - 3rd May, Klinikum rechts der Isar (Munich)

13 Characterization at MRI
Example of patient QA EBT3 Varian Trilogy, Millennium MLC, VMAT delivery technique

14 Characterization at UCSF
Clinical tests & clinical training Secondment, 28th July - 8th August, UCSF department of Radiation Oncology; Detector characterization through clinical test; Clinical training: Machine QA; Introductory training to Pinnacle TPS; Use of gafchromic films; Joined the daily activity of medical physicists.

15 Characterization at UCSF
Example of small field dosimetry and linac commissioning 1x1cm2 Varian TrueBeam Virtual wedge 45°

16 Characterization at UCSF
Example of linac commissioning measurement Water phantom as reference

17 Characterization at PTC czech s.r.o.
Dose linearity PBS mode only 1ms pulse duration, 10ms pulse period, 3.5mm spot  at 226MeV 0.02-2MU/spot 33x17 spot map Output norm. to 1MU/spot -0.04% 1.7% -0.9% 0.8% -0.1% -0.05% 0.01% PPC05 signal (nC) IC array signal (pC) Ref detector: PPC05

18 Characterization at PTC czech s.r.o.
Beam profiles (protons) Lynx: a scintillator based detector for PBS mode machine QA

19 Ongoing activities/Future work
Long term stability measurements with 60Co Further test with PBS proton beams; Paper about the characterization of the 1D IC array prototype; Extend the technology to a 2D detector - components drawing already started, first prototype assembling foreseen for the end of the year.

20 Trainings, conferences, secondments…
Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014

21 Secondments Upcoming:
Klinikum rechts der Isar, Munich ( ) UCSF RadOnc dep., S.Francisco ( ) Upcoming: B&A secondment at STMicroelectronics (Catania, July 2015 tentative) UCSF (early 2015 tentative)

22 Conferences, trainings…
AAPM, Austin (TX), Poster presentation: “An Ionization Chamber Array with High Spatial Resolution for External Beam Radiotherapy”, M.Togno, D.Menichelli, J.J.Wilkens. Upcoming: AAPM 2015, … Trainings: Clinical trainings at MRI & UCSF (already mentioned); Radiation protection training (IBA Dosimetry); Training on SBRT radiotherapy technique (Klinikum rechts der Isar); ARDENT B&A training (CERN): Technology Transfer, Management, Entrepreneurship. PhD activities: already spent more than 2 weeks at the University to teach students at physics labs.

23 Thank you! Michele Togno - III ARDENT Annual Meeting, Schwarzenbruck - September, 30th 2014