Technology: An Introduction to GA 310 Instrument and troubleshooting
3
Scanner Plates with acrylamide gel Laser CCD Camera Instruments overview Detection on 377
4
Capillary/ies CCD Camera Buffer Laser Instruments overview Detection on 310
5
ABI Prism Technology From filter wheel to CCD camera PMT CCD Camera Laser 370 373 377 310 3100 3 730
6
“gel” flow DNA migration Capillary electrophoresis Electroendosmotic flow
7
silica GS entangled polymer Capillary electrophoresis Dynamic coating of capillary
8
Instruments 377 / Acrylamide gel
9
Instruments 310 / Capillary electrophoresis
11
Includes the positioning mechanism and the carrier which accommodates 48 or 96 tube trays Instruments 310 / Autosampler tray
12
>syringe >syringe drive >the pump block Instruments 310 / Pump block
13
Capillary and electrode are placed into the sample Voltage is applied “-” charged DNA enters the capillary as it migrates toward the “+” electrode at the other end of the capillary Capillary electrophoresis Electrokinetic injection
14
Instruments overview CCD camera detector 512 pixels VIRTUAL FILTER Dump Read 64
15
100% 75% 25% 0% Sequencing s/w overview Spectral overlap of dyes
16
The matrix is used to filter raw data in order to “extract” the rigt value for each color. Raw data still has signals also in the “wrong” colors, the multicomponent analysis subtracts the values for each peak giving a clear result. Sequencing s/w overview What is a matrix / spectral calib.
17
Direct dilutionSpin columnGel purification Exo I / SAP treatmentAmmonium Acetate ppt PCR PRODUCT SEQUENCING REACTIONS Reaction overview Cleaning PCR product
18
It is very important to roughly estimate the amount of template DNA Agarose gel (0.7 to 1.2 %) stained with EtBr + ladder or, even better, a scalar amount of a well quantified template (i.e. linearized pGEM) Spectrometer: OD 260 / OD 280 Fluorometer Sequencing overview Template amount evaluation 257550100 X ladder Y
19
Seq. troubleshooting Too much DNA TOP HEAVY NOT USABLE signal too strong NOT USABLE signal too weak
20
TOP HEAVY NOT USABLE signal too strong NOT USABLE signal too weak OK Seq. troubleshooting Too much DNA
21
Seq. troubleshooting Too strong signal
23
raw data Seq. troubleshooting Too little DNA
24
Actual primer Tm 43.5°C, estimated > 52°C Seq. troubleshooting Noise due to weak signal 42°C 45°C
25
<100 ng plasmid Seq. troubleshooting Noise due to weak signal 300 ng plasmid
26
Seq. troubleshooting Noise due to insertion / deletion Rev. Forw.
27
free dye blobs Seq. troubleshooting Dye terminators contamination
28
Centrisep Columns (PE Biosystems) Multiscreen plate (Millipore) DyeEx or DyeEx96 (Qiagen) EtOH + 3M NaOAc precipitation SAP (BAP) digestion (usb Corporation) Phenol:Chl extraction EtOH (+ MgCl 2 ) precipitation Seq. troubleshooting Free dye terminators removal o b s o l e t e
29
Seq. troubleshooting Degradation of stored seq. product
30
Seq. troubleshooting Matrix (spectral cal.) problem
31
100%T 45%C Seq. troubleshooting Matrix (spectral cal.) problem
32
Instr. related problems Spike due to CCD current (310) raw data anal. data
33
Polymer related problems Spikes due to old POP or dust raw data anal. data
34
In this case clean properly the capillary window with 70% EtOH and check the water quality (should be ddH 2 O). Unfiltered buffer solution in the buffer vial can generate a lot of spikes in the electropherograms. Instr. related problems High baseline + spikes on 310
35
Capillary related problems Loss of resolution change capillary!!!
36
Instr. related problems Syringe problem (waterfall) Wash new syringes carefully with 60°C HPLC water from Merck, and then good with cold water. Sometimes its necessary and helps to wash new syringes with 2 N NaOH (prepared with HPLC water from Merck and afterwards with cold HPLC water).
37
EtOH 70% cleaning Instr. related problems Capillary window cleaning (310)