Components of Radiographic Image Quality

Similar presentations


Presentation on theme: "Components of Radiographic Image Quality"— Presentation transcript:

1 Components of Radiographic Image Quality
Radiologic Technology 244 created: Fall 2005 Rev

2 Review handouts

3 Main Factors Affecting Recorded Detail
kVp & mAs Technique Selection (Time) Motion Object Unsharpness Focal Spot Size SID (Source to Image Distance) OID (Object to Image Distance) Material Unsharpness/ Film Screen Combo

4 Factors that affect Recorded Detail
Geometric unsharpness OID SID SIZE SHAPE Motion unsharpness (blurring) Intensifying Screens Film Speed / Composition Film – Screen contact Kvp & Mas (density / visibility)

5 GEOMETRIC QUALITIES DETAIL DISTORTION MAGNIFICATION

6 The degree of sharpness in an object’s borders and structural details.
How “clear” the object looks on the radiograph

7 Recorded Detail Other names: -sharpness of detail -definition
-resolution -degree of noise

8 What are these What does they measure?

9

10

11 Factors Affecting DENSITY
PATIENT THICKNESS,PATHOLOGY MAS & KVP SID

12

13

14 POOR DETAIL GOOD DETAIL

15 Motion Can be voluntary or involuntary
Best controlled by short exposure times Use of careful instructions to the pt. Suspension of pt. respiration Immobilization devices

16

17 Decrease Motion Unsharpness
Instruct patient not to move or breath Use Immobilization devices Use Short exposure times Lock equipment in place

18 NAME 4 CAUSES

19 Blurring of image due to patient movement during exposure.

20 Focal Spot Size Smaller x-ray beam width will produce a sharper image.
Fine detail = small focal spot (i.e. small bones) General radiography uses large focal spot Beam from penlight size flashlight vs. flood light beam

21

22

23

24 FOCAL SPOT ANGLE

25 Object Unsharpness Main problem is trying to image a 3-D object on a 2-D film. Human body is not straight edges and sharp angles. We must compensate for object unsharpness with factors we can control: focal spot size, SID & OID

26

27 SID Source to Image Distance
The greater the distance between the source of the x-ray (tube) and the image receptor (cassette), the greater the image sharpness. Standard distance = 40 in. most exams Exception = Chest radiography 72 in.

28 SID Shine a flashlight on a 3-D object, shadow borders will appear “fuzzy” On a radiograph it’s called ______________ A true border – _____ Farther the flashlight from object = sharper borders. Same with radiography.

29 OID Object to Image Distance
The closer the object to the film, the sharper the detail. OID , penumbra , sharpness  OID , penumbra , sharpness  Structures located deep in the body, radiographer must know how to position to get the object closest to the film.

30

31 Distortion Misrepresentation of the true size or shape of an object
MAGNIFICATION (size distortion) TRUE DISTORTION (shape distortion)

32 Shape Distortion Misrepresentation of the shape of an object
Controlled by alignment of the beam, part (object), & image receptor Influences: Central ray angulation & body part rotation

33

34

35

36 Elongation Foreshortened Normal

37

38 Distortion (x-ray beam not centered over object & film)
Distortion (object & film not parallel)

39 Central Ray Radiation beam diverges from the tube in a pyramid shape.
Photons in the center travel along a straight line – central ray Photons along the beam’s periphery travel at an angle When central ray in angled, image shape is distorted.

40 Distortion of multiple objects in same image (right) due to x-ray beam not being centered over objects.

41 Central Ray Angulation
Body parts are not always 90 degrees from one another Central ray angulation is used to demonstrate certain details that can be hidden by superimposed body parts. Body part rotation or obliquing the body can also help visualize superimposed anatomy. NAME 3 EXAMPLES

42 MAGNIFICATION caused by:
TUBE CLOSE TO THE PART (↓SID) PART FAR FROM THE CASSETTE (↑ OID) Compensate for MAG : ↑ OID by ↑ SID = “increase SID 7” for every 1” OID”

43 Size Distortion & SID Major influences: SID & OID
As SID , magnification  Standardized SID’s allow radiologist to assume certain amt. of magnification factors are present Must note deviations from standard SID

44

45 In terms of recorded detail and magnification, the best image is produced with a small OID and a large SID.

46 What can be done to improve the detail with a large OID?

47 Use a smaller FS

48 40” SID VS 72” SID

49

50 Size Distortion & OID If source is kept constant, OID will affect magnification As OID , magnification  The farther the object is from the film, the more magnification

51 How can it be measured?

52 Measuring % of Magnification
SID SOD

53 Measuring % of Magnification
What is the % of mag when you have a 72” SID and 4” OID? DO the math…………

54 Material Unsharpness Equipment used can contribute to image unsharpness Fast film/screen combinations = decrease in image sharpness Slower film/screen combinations = increase in image sharpness

55 Intensifying screens Lower patient dose Changes resolution of image
slow screens less LIGHT = better detail Faster – less detail (more blurring on edges)

56

57 Intensifying Screens: Review
Located inside the cassette (film holder) Calcium Tungstate Blue to purple light Rare Earth Green & Ultraviolet light

58 POOR SCREEN CONTACT FOAM BACKING HELPS TO PLACE INTENSIFYING SCREENS IN DIRECT CONTACT WITH THE FILM – NO GAPS IF GAPS – MORE LIGHT CAN BE EMITTED IN SPACE, CAUSING THE IMAGE TO BE OF POOR DETAIL

59

60 WIRE MESH SCREEN CONTACT TEST

61

62 Screen Speed Efficiency of a screen in converting x-rays to light is Screen Speed.

63 Spectral Matching (F/S)
What does it mean? Name the two types of screen phosphors What light spectrum do they emit?

64 Spectral Sensitivity Film is designed to be sensitive to the color of light emitted by the intensifying screens. Blue LIGHT– Conventional Calcium Tungstate screen Green, Yellow-Green LIGHT – Rare Earth screen

65 Spectral Matching (F/S systems)

66 Spectral Matching (F/S systems)
Red safe light

67

68 Safe lights What wattage bulb? Distance from counter top?

69 Review of Film Characteristics
Size of silver halide crystals & emulsion thickness determine speed of film and degree of resolution Speed – the response to photons Resolution – the detail seen

70 What are these What are they made of

71 Film Speed / Crystal size
Larger crystals or Thicker crystal layer Faster response= less detail, and less exposure (chest x-ray) Finer crystals / thinner crystal layer =Slower response, greater detail, more exposure (extremity)

72 IMAGE ON FILM SINGLE EMULSION = BETTER DETAIL
DOUBLE EMULISON = LESS DETAIL PARALLAX With double emulsion – an image is created on both emulsions – then superimposed – slight blurring of edges

73 Extremity vs Regular cassettes

74 QUANTUM MOTTLE Film grain, or graininess, refers to the tiny black spots that make up the visible image, one grain from each silver halide crystal exposed MORE COMMON IN CR SYSTEMS NOW NOT ENOUGH PHOTONS TO CREATE IMAGE

75 Factors Affecting mAs LIST 6 factors

76 Factors Affecting mAs Patient factors: size of pt., density of tissue, pathology kVp Distance - how Grids Film/Screen Combinations Processing

77 Technique /Denisty CHANGES

78 Log denisty H & D curve

79 a densitometer, measures film blackness. Film blackness is the relationship of the intensity of the light that hits the film from the view box (incident intensity) to the intensity of the light transmitted through the film (transmitted intensity). These measurements plotted on a graph produce a characteristic curve. The limitations of the human eye determine the useful density range in diagnostic radiography. The diagnostically useful range of densities is 0.25 to 2.5. The later module on exposure calculation considers this in more detail.

80 Film latitude ? What does it mean how does it plot on the curve?

81

82

83 Main Factors Affecting Recorded Detail
kVp & mAs Motion Object Unsharpness Focal Spot Size SID (Source to Image Distance) OID (Object to Image Distance) Material Unsharpness/ Film Screen Combo