1 Basic fill color R 100 G 160 B 200 Accent 1 ”Logo Blue” R 0 G 39 B 118 Yellow (highlight) R 252 G 217 B 0 Accent 2 R 118 G 118 B 118 OP300 Technical training presentation Installation

2 OP 300 is delivered in 2 boxes OP 300 delivery content BOX 1 contains - Column and carriage - Accessories - Wall bracket - Manuals BOX 2 contains - Main Support - Rotating unit

3 Pre-installation requirements PC minimum requirements Windows XP Professional with SP2 (32-bit) 2.4 GHz Core 2 Duo dual core processor or equivalent 2 GB RAM 500 GB HDD or larger Gigabit Ethernet (CAT6) One available PCI-Express x16 bus slot for a GPU card 2 available USB ports (hardware key locks) 19" LCD display, 1280 x 1024 Time and persons For complete installation 1 day and 2 persons are needed Installation room Check minimum room size from pre-sales check-list. Physical environment specs in the pre-sales check-list. Unit mounts to the floor and wall with mounting hardware (not supplied). Radiation shielding Local requirements must be met

4 Product dimensions (78.9”) Range: 54.5” to 56.1”

5 Installation overview Stand up the column Attach unit to the wall and / or to the floor Lift the carriage and hang it on the column. Check leveling of the unit Connect all cables and attach exposure button and touch panel Carry out calibration procedure for the unit Install and configure Cliniview For detailed instructions please refer to Installation manual

6 Unpacking the unit; column box

7 Unpacking the unit; overhead box

8 Stand up the column

9 Three mounting variations: With one wall mounting bracket and bolting unit to the floor With two wall mounting brackets -- no floor drilling is needed Note: Base plate and second wall mounting bracket has to be ordered separately With one wall mounting bracket and base plate

10 For the Wall Mounting Bracket: Prior to mounting column to the mounting bracket, ensure that the bracket has room to adjust forward and back, otherwise there will be no room to level the unit front-to- back. Leveling is done after the overhead is mounted onto the column, using levels built into the overhead.

11 Remove the side covers

12 Cut the foam, leave the foam banded onto the overhead

13 Prepare for the overhead

14 Slide the overhead onto the column

15 Pins align the overhead prior to fixing with the bolts

16 Making the connections from overhead to column

17 Connections, continued

18

19 Mains power, handswitch and Ethernet connections

20 Remote handswitch terminals

21 Remove the packing foam

22 Remove the plywood

23 Remove the brackets

24 Configure the rotation stop Remove rotation stop bracket Rotate the tubearm 90 degrees CCW, as seen from above Reinstall rotation stop bracket

25 Level the unit Levels located in overhead Use column bracket to set level

26 Configuring the mains power jumpers Default 220V configuration 100V, 115V jumpers stored on side panel

27 Install and Wire the GUI

28 Set the column height microswitch

29 Install the side covers

30 Ceph arm mounting hardware

31 Ceph arm mounting brackets

32 Ceph head packaging

33 Mounting the ceph head

34 Level the ceph head

35 Ceph head connections

36 Ceph head connections, continued

37 Host computer video card Reconstruction video card packed in column box Open host computer, remove old video card from PCI slot (A) Locate “Y” power adapter (B) Install “Y” power adapter (C) Plug in reconstruction video card Ensure that two power plugs are connected to video card (D) Locate reconstruction dongle, plug into available USB port on the host computer A B C D

38 Powering on the unit

39 “Push any button” to initialize the unit

40 Configure the host computer’s NIC

41 Launch Cliniview Software

42 Configure Driver Left-ctrl/right click on driver icon to open the driver setup script Enter the IP address of the unit to be controlled

43 Launch S2 Terminal

44 Calibration of the unit How to start: Patient record must be active - select in the control panel - select “Service” in the control panel -check to see if any items have red exclamation points on them – those items must be recalibrated

45 Alignment and QA Tools Pan alignment cone assembly QA PhantomCT Geometry tool Leveling base

46 Acceptance limits Passed Calibration program is successfully done. Move on to next calibration. Not passed Adjustment is still needed. Follow the instructions the image (if any) and take another exposure. Some calibration programs are iterative and demands a few repetitions. Failed System could not decide what adjustment should be done in order for the calibration to succeed. This calibration status is always the result of some error condition. Taking another exposure will not help. The image may give a hint on what the problem is (e.g. no radiation, collimator severely tilted, image data corrupted…).

47 Pan geometry calibration Purpose: Panoramic geometry position calibration Acceptance limits: Passed / Not passed / Failed GUI informs if calibration has passed

48 CT Geometry calibration Purpose: Calibrates ct geometry, needed only with 3D units Acceptance limits: Passed / Not passed / Failed GUI informs if calibration has passed

49 Pan pixel calibration Purpose: Panoramic sensor pixel calibration Acceptance limits: Passed / Not passed / Failed GUI informs if calibration has passed Pan pixel calibration

50 CT pixel calibration Purpose: Performing pixel gain and dead pixel correction Acceptance limits: Passed / Not passed / Failed GUI informs when calibration is passed

51 Other calibration items Confidential, for distributors only51 Perform full calibration if red exclamation points are found on calibration screen Full calibration if ceph arm is added Use S2 service mode command “indicator service clear” to reset calibration period

52 Install the rear cover