Presentation is loading. Please wait.

Presentation is loading. Please wait.

Physical & Psychophysical Evaluation of a New Digital Specimen Radiography System for Use in Mammography Elizabeth Krupinski, PhD Hans Roehrig, PhD University.

Similar presentations


Presentation on theme: "Physical & Psychophysical Evaluation of a New Digital Specimen Radiography System for Use in Mammography Elizabeth Krupinski, PhD Hans Roehrig, PhD University."— Presentation transcript:

1 Physical & Psychophysical Evaluation of a New Digital Specimen Radiography System for Use in Mammography Elizabeth Krupinski, PhD Hans Roehrig, PhD University of Arizona William Schempp, PhD MedOptics Corporation

2 Rationale w Specimen radiography of excised breast tissue is critical to ensure that an appropriate tissue sample has been obtained or that complete removal of the lesion has been successful.

3 Objectives w To physically characterize the new high- resolution digital specimen radiography system from MedOptics Corporation. w To evaluate human observer detection performance using images acquired with the new system.

4 The Digital System w Camera 1024 x 1024, 1” x 1” large area high-res CCD 2:1 fiber optic taper bonded to CCD (2”x2” FOV) standard Lanex Regular screen coupled to large end of taper image data digitized at 500 Kpix/sec, 12-bits w Faxitron MICRO 50 x-ray source 15-50 kVp range 1-90 sec time selection SID = 50 cm

5 Faxitron HVL & Eeff

6 Camera X-radiation Response 1.0010.00100.001000.00 Exposure [mR] 1.00 10.00 100.00 1000.00 10000.00 C o r r e c t e d M e a n [ A D U ] o r R M S [ A D U ] Camera Response (Mean and Stdv (rms) in Digital Units [ADU]) to X-radiation (mR) 30 kVp 20 kVp 30 kVp 20 kVp 25 kVp Corrected Mean Curve Fit for 30 kVp log(Y) = 1.04655*log(X) + 1.7888 Corrected RMS Curve Fit for 30 kVp log(Y) = 0.50577*log(X) - 0.291

7 Camera Dark Signal & Noise 0.010.101.0010.00100.00 Time [sec] 1.00 10.00 100.00 1000.00 M e a n o r S t d e v [ G L ] Dark Signal Dark Noise Uncorrected Dark-Signal and Dark-Noise of Uncooled CCD as Function of Time

8 Noise Power Spectrum 25 kVp 0.002.004.006.008.0010.00 Spatial Frequency (lp/mm) 0.01 0.10 1.00 10.00 100.00 N o i s e P o w e r ( m m ^ 2 ) Noise Power Spectra for Exposures at 25 kVp Exposure Time 1 sec Exposure Time 4 sec

9 Noise Power Spectrum 30 kVp 0.002.004.006.008.0010.00 Spatial Frequency (lp/mm) 0.10 1.00 10.00 N o i s e P o w e r ( m m ^ 2 ) Noise Power Spectra for Exposures at 30 kVp Exposure Time 1 sec Exposure Time 3 sec

10 Comparison NPS & Square MTF 0.002.004.006.008.0010.00 Spatial Frequency (lp/mm) 0.00 0.01 0.10 1.00 10.00 100.00 N o i s e P o w e r ( m m ^ 2 ) o r M T F ^ 2 Scaled MTF^2 from Curve Fit Noise Power Spectrum 25 kVp, 4 sec Noise Power Spectrum and MTF for Specimen Radiography System

11 MTF 0.002.004.006.008.0010.00 Spatial Frequency (lp/mm) 0.00 0.20 0.40 0.60 0.80 1.00 M T F ( r e l. U n i t s ) MTF of the CCD-based Camera of the MedOptics Specimen Radiography System; derived from the Fourier Transform of profiles of x-ray images of a 10 micron wide x-ray slit ("Line-Spread-Function") Dashed Line 25 kVp Solid Line 30 kVp Polynomial Coefficients Curve Fit: Polynomial of degree 2 Degree 0: 0.997419 Degree 1: -0.220671 Degree 2: 0.0125387

12 MTF from Square Wave Response 0.002.004.006.008.0010.00 Spatial Frequency (lp/mm) 0.00 0.20 0.40 0.60 0.80 1.00 M T F ( r e l. U n i t s ) MTF of the CCD-based Camera of the MedOptics Specimen Radiography System; derived from the Fourier Transform of x-ray images of a lead-bar pattern ("Squarewave Response") Dashed Line 25 kVp Solid Line 30 kVp

13 Observer Performance Study w 2 CDMAM contrast-detail phantoms Commercially available large version Small experimental version w Imaging conditions Traditional plain film Fuji Computed Radiography (CR) MedOptics Digital System w 25 kV & 30 kV @ High (3600 GL), medium (2400 GL) & low (1200 GL) exposures

14 Film Image Optical Densities

15 Viewing Conditions w Traditional film & Fuji CR film standard viewbox @ 1100 ft-L extraneous light masked w Fuji CR digital & MedOptics Digital DataRay DR110 1728 x 2304 140 ft-L Perceptually linearized with Barten curve

16 Observers & Task w 10 undergrad students in each condition w On outline of phantom indicate location of all disks visible in the phantom(s) w Window/Level & zoom available on monitor w Ambient lights turned off in all conditions

17 Large Phantom Results F = 68.44, p < 0.0001

18 Small Phantom Results F = 52.62, p < 0.0001

19 Performance Summary w Increases monotonically low to high exposures w 25 kV > 30 kV w MedOptics Digital System Best w Traditional plain film worst w Fuji CR monitor & film in middle w Minor differences between phantoms

20 Conclusions w MedOptics Digital Biopsy System yields high detection performance w Currently mammographers image core biopsy samples with either traditional systems or stereotactic units Traditional takes too long Stereotactic requires patient to get out of unit to image sample w Digital avoids these problems & is fast


Download ppt "Physical & Psychophysical Evaluation of a New Digital Specimen Radiography System for Use in Mammography Elizabeth Krupinski, PhD Hans Roehrig, PhD University."

Similar presentations


Ads by Google