Pacemaker Malfunctions Even less amusing!. Pacemaker Codes (NASPE/BPEG) Position I IIIII Category Chamber(s) Chamber(s) Response to paced sensed sensing.

Similar presentations


Presentation on theme: "Pacemaker Malfunctions Even less amusing!. Pacemaker Codes (NASPE/BPEG) Position I IIIII Category Chamber(s) Chamber(s) Response to paced sensed sensing."— Presentation transcript:

1 Pacemaker Malfunctions Even less amusing!

2 Pacemaker Codes (NASPE/BPEG) Position I IIIII Category Chamber(s) Chamber(s) Response to paced sensed sensing O=None O=None O=None A=Atrium A=Atrium T=triggered V=Ventricle V=Ventricle I=Inhibited D=Dual (A+V) D=Dual(A+V) D=Dual (T+I)

3 Triggered l Means different things n If an intrinsic event is sensed, the pacemaker will trigger pacing to that chamber OR to another chamber –P wave will trigger an AV interval –R wave will trigger pacing to the ventricle

4 Triggered? Atrial paced spike triggers AV interval Triggered ventricular pacing spike is in refractory period of ventricular cycle

5 A O O Chamber Paced Chamber Sensed Action or Response to a Sensed Event AOO Pacing Asynchronous mode Not sensing so can’t react

6 V V I Chamber Paced Chamber Sensed Action or Response to a Sensed Event VVI Pacing

7 Even more! l Fourth letter- Programmable Functions n R = rate modulation n C = Communicating n M = Multiprogrammable n S = simple programmable n O = None l Fifth letter – antitachycardic functions n O = none n P = paced n S = shock

8 Rate Modulation l Atrium or Ventricle is pacing or tracking above the set upper limit (permanent pacemaker) n Used to help compensate for increased demands –Sensor can be: l Change in temperature l Change in movement or body position l Change in pH l Change in minute ventilation

9 D D D R Chamber Paced Chamber Sensed Response to a Sensed Event DDDR Pacing Rate modulated

10 Pacing Malfunctions l Often a two person job n One person supports the patient n One person troubleshoots the pacemaker

11 Failure to pace (not tryin’ to tickle) l No pacing spikes or not enough coupled with the intrinsic rhythm to ensure good CO l Pacing energy not being transmitted from generator to patient

12 Failure to pace ( not trying to tickle)

13 How do you fix this? l Pacing spikes not visible (often an equipment problem) n Pacer on? n Rate set correctly? n Battery fresh ? n Connections tight? n Get another cable n Get another generator n Often an equipment problem-no energy thrown

14 Failure to capture (target is not ticklish/not ticklin’ hard enough) l Pacing Arm is throwing out energy-target is not responding

15 Loss of Capture (target is not ticklish) Nothing happens in response to arm throwing out impulse

16 How do you fix this? l Myocardium is not responding to stimulation n Battery fresh? n Connections tight? –Make sure lead is in contact with myocardium n mA high enough? n Assess electrolytes, oxygenation, acid base balance n Ischemic tissue? Fibrin sleeve? n Reposition patient or patient’s arm on side of pacemaker if subclavian entry

17 Undersensing (ticklin’ too much) l Intrinsic rhythm not seen so pacing arm throws out energy

18 Undersensing (ticklin’ at the wrong time) Pacer eyeball is set too high! Can’t see target moving so the arm throws out a stimulus

19 The danger of “not seeing”... Pacer arm throws out stimulus during vulnerable portion of cardiac cycle

20 How do you fix this l Pacermaker not seeing correctly n Battery fresh n Connections tight –Leads in contact with myocardium –Lead fracture n Do a sensitivity threshold to get eyes at proper level

21 Oversensing ( not ticklin’ enough) l Pacemaker misinterprets noncardiac events as intrinsic activity

22 Oversensing (not ticklin’ enough) Pacer eyeball thinks it sees cardiac activity so the arm does not throw out an stimulus

23 How do you fix this l Pacermaker not seeing correctly n Battery fresh n Connections tight –Leads in contact with myocardium –Lead fracture n Do a sensitivity threshold to get eyes at proper level

24 Let’s practice

25 Other Complications l Infection l Pulmonary embolus l Venous Thrombus l Myocardial perforation l Endocarditis

26 Nursing Care l Assess heart rate and rhythm l Protect patient from injury l Troubleshoot malfunctions l Evaluate pacemaker function l Keep patient informed

27 Sample documentation l Temporary pacemaker (0800) n Epicardial wires ( 2 ventricular) n Pacing threshold 3; mA set at 6 n Sensitivity threshold 4 mV; mV set at 2 mV n Rate set at 60; patient’s intrinsic rate 30 n Patient pacing about 50% of time. 100% capture when paced. l Temporary pacemaker (1400) n Pacemaker off. Pacing wires grounded and taped to chest wall. Site care with betadine.

28 Flippin’ a switch l What’s the deal with the magnet?

29 No Magnet

30 With Magnet

31 Rapid Atrial Pacing

32 l Used only in the atrium l Pace at rates from 80-800 n Need to get higher than patient’s ATRIAL rate l Physician must be present when in use n Nurse may only connect l Always have defibrillator available

33 What are the possibilities? l Single chamber pacemaker l Dual chamber pacemaker l Free standing RAP machine n Think Frankenstein

34 Rapid Atrial Pacing (single chamber) l Flip open top of pacemaker n Press enable to begin l Set rate according to physician instruction or let MD set rate n Rate will be higher than patient’s atrial rate l Press hold to deliver until instructed to stop n Will be done in bursts n Pacing light will flash

35 Rapid Atrial Pacing (dual chamber) l Found on bottom half of pacer l Must hit menu screen and scroll to menu 3 l Set rate by turning round knob l Press select to deliver and HOLD until instructed to release Menu Select 3 DDD 80 440 800 RAP 320 SELECT Press toDeliver Rapid Atrial Pacing

36 Permanent pacemakers

37 Differences l You cannot access pacing generator l You cannot see leads l Helps interpretation of strip if you know how pacemaker is set up n Should have card n Old records n Request info from cardiologist

38 You are in trouble! What can you try if you have a malfunction with a permanent pacemaker that is compromising your patient?

39 Biventricular pacemakers? l Pacing leads in the RV and into the coronary sinus l Allow both ventricles to pace synchronously l Improves heart function for patients with CHF l EKG will not look different from single ventricular pacing

40 The END