Download presentation
Presentation is loading. Please wait.
Published byRylan Manor Modified over 9 years ago
1
A Bolometer Array Instrument for Millimeter Wavelength Philippe Camus, CNRS-CRTBT (Grenoble) Milano, July 6th, 2001
2
Bolometric Millimeter Camera Sunyaev-Zel’dovich effect High-Z objects and star formation regions imaging at 1mm => wavelength 1.2mm /2.1mm Which detector at 2.1mm? IRAM 30m Field 5X5 arcmin 256 pixel@2.1mm 1024 pixel@1.2mm
3
Today and future Instruments MAMBO (IRAM 30m) 37 bolometer array on Si3N4 membrane (MPfir)
4
Optical Layout Cryostat Dilution 100-300mK Adaptation to ‘any’ detector Arrays
5
NbSi Bolometer (CSNSM/ Service d’Aéronomie) Bolometer on thin Si 3 N 4 Memebrane 5X5mm e=100nm Thin film geometry optimisation (typ. 100X400 e=100nm) NEP = 5.10 -16 W/Hz 1/2 @300mK or 5.10 -17 W/Hz 1/2 @100mK
6
Small Array Development (6X6 detectors) NbSi layer co-evaporation & Lithography Objective 8X8 bolometers
7
Electronic Read-Out Multiplexing is required to reduce the number of output amplifiers (JFET) Low dissipation/low temperature operation (HEMTor MOSFET?) Objective of 8 detector/amplifier
Similar presentations