Presentation is loading. Please wait.

Presentation is loading. Please wait.

Shortwave and Microwave Diathermy part(2). Induction Electrodes (Drum Electrode) One Or More Monopolar Coils Rigidly Fixed In A Housing Unit May Use More.

Similar presentations


Presentation on theme: "Shortwave and Microwave Diathermy part(2). Induction Electrodes (Drum Electrode) One Or More Monopolar Coils Rigidly Fixed In A Housing Unit May Use More."— Presentation transcript:

1 Shortwave and Microwave Diathermy part(2)

2 Induction Electrodes (Drum Electrode) One Or More Monopolar Coils Rigidly Fixed In A Housing Unit May Use More Than One Drum Depending On Area Treated Penetration – Deeper Soft Tissues Toweling Important 2

3 Heating With Continuous SWD Patient Sensation Provides Basis For Recommendations Of Continuous SWD Dose I (Lowest) (<38 W) - No Sensation of Heat Dose II (Low) (~80 W)- Mild Heating Sensation Dose III (Medium) (80-300 W) - Moderate or Pleasant Heating Sensation Dose IV (Heavy) (>300 W) -Vigorous Heating Within Pain Threshold 3

4 Pulsed SWD Referred To By Different Names – Pulsed Electromagnetic Energy (PEME) – Pulsed Electromagnetic Field (PEMF) – Pulsed Electromagnetic Energy Treatment (PEMET) 4

5 5 Pulsed SWD

6 PSWD Mean Power Calculations Pulse Period (Pulse on + off time) = – Peak Pulse Power (W) / Pulse Rep Freq (Hz) Percentage On Time = – Pulse Duration (msec) / Pulse Period (msec) Mean Power = – Peak Pulse Power (W) / Percentage on Time 6

7 Example Peak Pulse Power = 800 W Pulse Duration =.4 ms Pulse Frequency = 200 Hz Pulse Period = 800 W / 200 Hz = 4 ms % on time =.4 / 4 =.10 or 10% Mean Power = 10% of 800 = 80 W – End Result = thermal effect – < 38 W = no heating 7

8 Pulsed SWD Interrupted Output Delivered In Series Of High-Frequency Bursts (40-400 usec) Pulse Rate Selected With Pulse Frequency Control (1–1000 Hz) Off-Time Longer Than On-Time Low Mean Power Output Uses Drum Electrode 8

9 Shortwave Diathermy vs.Ultrasound n Pulsed SWD Produces The Same Magnitude And Depth Of Muscle Heating as 1MHz Ultrasound (Draper, JAT 1997) 9

10 Treatment Time Most Typically SWD Treatments Last For 20-30 Minutes Remember As Skin Temperature Rises Resistance Falls 10

11 11 When Should Diathermy Be Used? l If The Skin Or Some Underlying Soft Tissue Is Tender And Will Not Tolerate Pressure l In Areas Where Subcutaneous Fat Is Thick And Deep Heating Is Required l Induction method l When The Treatment Goal Is To Increase Tissue Temperatures Over A Large Area l If The Skin Or Some Underlying Soft Tissue Is Tender And Will Not Tolerate Pressure l In Areas Where Subcutaneous Fat Is Thick And Deep Heating Is Required l Induction method l When The Treatment Goal Is To Increase Tissue Temperatures Over A Large Area

12 SWD Guidelines Indications – Continuous and Pulsed Precautions Contraindications 12

13 مولد الموجات القصيرة بالشرارة الكهربية Spark gap radio frequency generator RFC1 RFC2 SG1 C1 C2 C3

14 Vacuum tube sine wave circuits

15 Cutting and coagulation waves

16 مولد تيار الكي بالثيراترون Thyratron coagulation current generator Pulse generator RF OUTPUT V1V1 L3L3 C 1 50 pF C 2 0.001  F D1D1 R 1 10 K L1L1 L2L2 C 3 0.001  F R 2 10 K C 4 500 pF C 5 0.001  F From cut oscillator

17 Diathermy Semiconductor Circuits

18 Radiofrequency power amplifier semiconductor circuit

19 Electrosurgical Units (ESU) and Cautery Machines Function Electrosurgery is the application of a high-frequency electric current to biological tissue as a means to cut, coagulate, desiccate, or fulgurate tissue. Its benefits include the ability to make precise cuts with limited blood loss in hospital operating rooms or in outpatient procedures. Cautery, or electrocautery, is the application of heat to tissue to achieve coagulation. Although both methods are sometimes referred to as surgical diathermy. 19

20 How it works In electrosurgical procedures, the tissue is heated by an alternating electric current being passed through it from a probe. Electrocautery uses heat conduction from an electrically heated probe, much like a soldering iron. Electrosurgery is performed using an electrosurgical generator (also referred to as power supply or waveform generator) and a hand piece including one or several electrodes, sometimes referred to as an RF Knife, or informally by surgeons as a "Bovie knife" after the inventor. Bipolar electrosurgery has the outward and return current passing through the handpiece, whereas monopolar electrosurgery returns the current through a plate normally under the patient. Electrosurgery is commonly used in dermatological, gynecological, cardiac, plastic, ocular, spine, ENT, orthopedic, urological, neuro- and general surgical procedures as well as certain dental procedures. 20

21 21

22 22

23 23

24 24

25 25

26

27 Microwave Diathermy Two FCC Assigned Frequencies-2456 MHz and 915 MHz MWD Has Higher Frequency and Shorter Wavelength Than SWD Generates Strong Electrical Field and Relatively Little Magnetic Field Advantage: better focus wave on body, thereby more local heating affects Disadvantage: Depth Of Penetration Is Minimal In Areas With Subcutaneous Fat > 1 cm 27

28 Microwave Diathermy Unit A= Power Switch B=Timer C=Output Meter (indicates relative output in watts D= Power Output Level Knob E= Amber Light-Warming up / Red Light- Ready 28

29 MWD Applicators (Electrodes) Circular Shaped Applicators – 4” or 6” – Maximum Temperature At Periphery Rectangular Shaped Applicators – 4.5 x 5” or 5 x 21” – Maximum Temperature At Center 29

30 Microwave Applicator Set-Up Microwave Applicator Beams Energy To Patient Must Pay Attention To Cosine Law In 915 MHz Units Applicators Placed 1 cm From Skin 2456 MHz Units Have Manufacturer Recommended Distances and Power Outputs (Uses Antenna) 30

31 Best Treated areas for Microwave Tendons of foot, hand and wrist AC and SC joints Patellar tendon Distal tendons of hamstrings Achilles tendon Other areas of low subcutaneous fat 31


Download ppt "Shortwave and Microwave Diathermy part(2). Induction Electrodes (Drum Electrode) One Or More Monopolar Coils Rigidly Fixed In A Housing Unit May Use More."

Similar presentations


Ads by Google