1
Measures of solid hydrogen contamination problem in the liquid helium
6th International Workshop on Cryogenics Operations on November 2014 Measures of solid hydrogen contamination problem in the liquid helium Hiroshi Ikedaa,*, Yutaka. Kondo a Research Facility Center for Science and Technology Cryogenics Division, University of Tsukuba, Tsukuba, Ibaraki , Japan
2
CONTENT Introduction History of L280 in University of Tsukuba
The trouble with solid hydrogen Purity measurement and state understanding of internal purifier (New method !) Improving method the purity of the regeneration gas Gas chromatograph mass spectrometer impurity analysis of helium liquefaction machine Consideration of decreased purity (Changes in Linde helium Liqufier) Conclusions
3
Introduction ・In recent years, several universities that renewed their helium liquefier system have reported that solid hydrogen was mixed within the liquid helium produced by a helium liquefier. ・The University of Tsukuba has also installed a Linde L280 helium liquefier system in October Similar to other universities, cases where the helium flow of cryogenic equipment with flow impedance was blocked began to be reported around March ・In this work we focused on the internal purifier of the helium liquefier for the problem where solid hydrogen is mixed in the liquid helium produced by a helium liquefier.
4
History of L280 in University of Tsukuba
5
Decreased purity of helium recovery gas bag
About 1 year passed Decreased purity of helium recovery gas bag
6
helium liquefier flow ( L280)
7
Internal purifier Flow
Gasbag PV3415 CV3401 V3410 CV3430 CV3420 Flow Meter Cold Gas Line Impurer Helium Gas Atmospheric release PV3435 TI3475 TI3465 Heater PI3401 PI3445 HP Line LP Line Liqide Air CV3410
8
Decreased purity of helium recovery gas bag
⇒ 95% (Nomaly=99%) Differential pressure occurs when introduced into the internal purifier gas was decreased purity
9
Trend graph of the differential pressure occurs
ΔP=0.15MPa Cool down Mode Regeneration Mode Purify Mode Buffer tank
10
The trouble with solid hydrogen
11
The trouble with solid hydrogen
Flow impedance blocked by solid hydrogen Quantum Design MPMS PPMS (equipment to control temperature while limiting the flow of helium by a filter and capillary)
12
Trouble Instrument Worldwide! Japan Korea China India Spain France
Germany Worldwide!
13
Purity measurement and state understanding of internal purifier
New method!
14
Gas bag purity is reduced to 95% by the regeneration gas, introduction into the internal purifier, differential pressure occurs. ‣ For gas purity measurement discharged from the internal purifier was installed simple helium purity monitor immediately after the flow meter.
15
Simple purity monitor installation flow diagram
Gas Bag CV3415 CV3401 V3410 CV3430 CV3420 Flow Meter Cold Gas Flow Impurer Helium Gas CV3435 TI3475 TI3465 Heater PI3401 PI3445 HP Line LP Line Liqid Air 100% Purity Monter CV3410 Atmospheric release
16
Simple helium gas purity monitor
New Method! Ulvac WP-01 Pirany type SKD-20
17
A result of the purity measurement
The purity of the regeneration gas was reduced to 33% at maximum Initial setting valu ・Purification operation state Cold end temperature TI3475 = 32.5 K Internal pressure PI3445 = 2.5 MPa ・ Regeneration during operation Exit flow meter flow rate = 120 L/min Regeneration completion temperature TI3465 = 140 K Heater control temperature TI3475 = 95 K Warming heater output R3470 = 18 %
18
Improving method the purity
of the regeneration gas
19
Comparison with the liquefier
Internal purifier Cold end temperature Internal purifier Exit flow rate TCF50 L280 27.5 K 200 L/min 32.5 K 120 L/min High temperature! Low flow rate!
20
Setting changes 120L/min⇒190L/min CV3415 CV3401 V3410 CV3430 CV3420
Down fromD32.5K Gas Bag CV3415 CV3401 V3410 CV3430 CV3420 Flow Meter Cold Gas Flow Impurer Helium Gas CV3435 TI3475 TI3465 Heater PI3401 PI3445 HP Line LP Line Liqid Air 100% Purity Monter CV3410 Atmospheric release
21
Internal purifier regeneration gas purity measurement results
32.5K 120L/min
22
Internal purifier regeneration gas purity measurement results
29.0K 190L/min 26.0K 190L/min 32.5K 120L/min
23
Trend graph of the regeneration gas purity after improvement
26.0K 190L/min ΔP=0.039MPa
24
Befer After ΔP=0.15MPa ΔP=0.039MPa 32.5K、120L/min 26.0K、190L/min
25
Gas chromatograph mass spectrometer impurity analysis of helium liquefaction machine
26
Gas chromatograph mass spectrometer
SHIMADZU GC-2014 O
27
Gas chromatograph mass spectrometer impurity analysis of helium liquefaction machine
28
Improvement result The purity of the regeneration gas, 33% ⇒ 99%
Cold end temperature change, the regeneration gas flow rate increase The purity of the regeneration gas, 33% ⇒ 99% Pressure difference is reduced. Purification operation has stabilized
29
Consideration of decreased purity (Changes in Linde helium Liqufier)
30
1996 TCF50 Flow diagrams Adosorber 3 point ! 80K 20K LHe GHe LN2
Gas Bag InPur GHe Adosorber 3 point !
31
2010 L280 Flow diagrams Adsober Only One 20K LN2 LHe GHe Gas Bag
InPur GHe Gas Bag
32
2014 L140 New Flow diagrams Adosober Revival ! 20K LN2 LHe GHe Gas Bag
InPur GHe
33
Conclusions Regeneration gas purity was improved from 33% to 99% by the setting change of internal purifier. Purification operation has stabilized generation of the differential pressure is reduced. By monitoring established the purity helium simple monitor, it can insight into the operating state of the internal purifier at a low cost. As the cause of differential pressure and decreased purity of the regeneration gas occurred, in addition to the incorrect setting, it is considered as the cause did not have the adsorber inside purifier.