Presentation is loading. Please wait.

Presentation is loading. Please wait.

Compton polarimetry for EIC Jefferson Lab Compton Polarimeters.

Similar presentations


Presentation on theme: "Compton polarimetry for EIC Jefferson Lab Compton Polarimeters."— Presentation transcript:

1 Compton polarimetry for EIC Jefferson Lab Compton Polarimeters

2 ERHIC 6.6 GeV to 21.2 GeV 9.4 MHz Repetion rate Up to 21 recirculations 50 mA with “gatling gun” design 80 % min polarization Similar to CEBAF Vadim Ptitsyn eRHIC Accelerator Design EIC2014

3 MEIC Storage ring – Ring ring 748.5 MHz = 1.33 ns bunch structure 3 A at 3 GeV and 180 mA at 11 GeV 2 macrobunch with one polarization 2.3 us Measure polarization average of the two macrobunch Every electron bunch crosses every ion bunch Warm large booster (up to 25 GeV/c) Warm 3-12 GeV electron collider ring Medium-energy IPs with horizontal beam crossing Injector 12 GeV CEBAF Pre-booster SRF linac Ion source Cold 25-100 GeV/c proton collider ring Three Figure-8 rings stacked vertically Electron cooling

4 Compton asymmetry e +  e’ +  ’ (( ) (( )

5 Hall A Compton chicane

6 Cavity power Green laser using IR seed laser and PPLN frequency doubling Around 5 kW power 10 kW reachable Lazer polarization flip Abdurahim Rakhman (2011) Phd Thesis Syracuse

7 Hall A Photon detector FADC readout SIS3320 250 MHz FADC Digital integration with 240 Hz helicity flip Record all the signal for a given helicity Compute integrated asymmetry for a pair

8 Happex III results Friend Nucl.Instrum.Meth. A676 (2012) 96-105 Friend Phd Thesis CMU 2012 Pe =89.41%

9 Hall C Compton Electron Detector Diamond microstrips used to detect scattered electrons  Radiation hard  Four 21mm x 21mm planes each with 96 horizontal 200 μm wide micro-strips.  Rough-tracking based/coincidence trigger suppresses backgrounds

10 Compton Electron Detector Measurements Polarization analysis:  Yield for each electron helicity state measured in each strip  Background yields measured by “turning off” (unlocking) the laser  Asymmetry constructed in each strip Strip number corresponds to scattered electron energy  Endpoint and zero-crossing of asymmetry provide kinematic scale  2-parameter fit to beam polarization and Compton endpoint

11 Polarization Measurements Q-Weak Run 2 – November 2011 to May 2012 P Moller +/- stat (inner) +/- point-to-point systematic (0.54%) P Compton +/- stat +/- preliminary systematic (0.6%) Photocathode re-activation 0.64% normalization unc. not shown Preliminary

12 Preliminary Systematic Uncertainties Systematic UncertaintyUncertainty ΔP/P (%) Laser Polarization 0.1%0.1 Dipole field strength (0.0011 T)0.02 Beam energy 1 MeV0.09 Detector Longitudinal Position 1 mm0.03 Detector Rotation (pitch) 1 degree0.04 Asymmetry time averaging 0.15% Asymmetry fit 0.3% DAQ – dead time, eff. Under study?? Systematic uncertainties still under investigation, but final precision expected to be better than 1%  DA- related systematics likely the most significant remaining issue to study

13 Simulation background Bremstrahlung Halo 1 kW green laser 1 A 3 GeV electron beam Halo contribution modeled on PEP II Photon detector signal Electron detector signal

14 Compton polarimeter in low-Q 2 chicane Same polarization as at the IP due to zero net bend Non-invasive continuous polarization monitoring Polarization measurement accuracy of ~1% expected No interference with quasi real photon tagging detectors c Laser + Fabry Perot cavity e - beam Quasi-real high-energy photon tagger Quasi-real low-energy photon tagger Electron tracking detector Photon calorimeter Possible implementation in low Q 2

15 Hall A Compton chicane Vertical motion of electron detector to move detector close to the beam ( up to 5 mm ) Photon detector on movable table

16 Conclusion Compton polarimetry at 1% level achieved at Jefferson Laboratory and aiming at 0.5 % for 12 GeV parity program Jefferson Lab ideal ground for Compton testing for EIC since Compton is non invasive – Photon detector testing straight forward – Electron detector testing doable with planning because of vacuum. Looking into Roman pot option for ease of detector swapping


Download ppt "Compton polarimetry for EIC Jefferson Lab Compton Polarimeters."

Similar presentations


Ads by Google