Ultrasound machine knobology
Copyright © 2010 Academy of Veterinary Imaging
2
Ultrasound machine knobology… In other words, what the heck do all of those knobs and buttons do, and how can adjusting them affect image quality?!?!
3
GE ultrasound Image optimization Features
Presets, depth, frequency, auto-optimize, focal points, gain/TGC Features Archiving, transducers, SRI, compounding, harmonics, 3-D, LOGIQView, Doppler
4
GE ultrasound GE features: Go to: Image optimization: Presets Depth
Frequency Auto-optimize Focal points Gain/TGC Archiving Transducers SRI CrossXBeam Harmonics 3-D LOGIQView % affected LOGIQ 7 LOGIQ 5 LOGIQ P5 LOGIQ e LOGIQ Book In this presentation, click on the Home button as pictured here to return to this page
5
Image optimization Image quality is comprised of many things
During an ultrasound exam, the ultrasound machine settings should be changed as needed to optimize image quality Understanding just a few of the knobs can accomplish this for most scans L7 L5 P5 Le LB
6
Presets Settings can be saved as presets
Viewed in Application or Preset screen Different presets for different transducers/structures/patient species/sizes The more ‘complicated’ machine settings are saved in presets, so these do not need to be adjusted during every scan L7 L5 P5 Le LB
7
Depth Adjust as needed throughout scan L7 L5 P5 Le LB Optimal
8
Frequency Better resolution Less depth Lower frequency
High frequency Better resolution Less depth Lower frequency Lower resolution Better for depth L7 L5 P5 Le LB
9
Auto optimize ‘Personal preference’ button.
Adds more contrast…Whites are whiter, blacks are blacker Try on or off. Chose the setting that looks better to you. Auto optimize must be deselected then reselected for every new structure imaged. L7 L5 P5 Le LB
10
Focal points Place focal points at the region of interest.
Improved resolution is seen where the focal points are. Place focal points at the region of interest. L7 L5 P5 Le LB
11
Gain Overall gain Amplification of received signal TGC
Time Gain Compensation Compensates for beam attenuation (loss of energy as it passes through tissue) L7 L5 P5 Le LB
12
LOGIQ 7 Frequency Focal points TGC Depth Gain Auto optimize L5 P5 Le
(touch screen) Le Frequency LB Focal points TGC Depth Gain Auto optimize
13
LOGIQ 5 Frequency Focal points Depth Gain Auto optimize TGC L7 P5 Le
LB Focal points Depth Gain Auto optimize TGC
14
LOGIQ P5 Focal Frequency points Depth TGC Gain Auto optimize L7 L5 LB
Le Focal points Frequency Depth TGC Gain Auto optimize
15
LOGIQ e Frequency Focal points TGC Depth Gain Auto optimize L7 L5 P5
LB TGC Depth Gain Auto optimize
16
LOGIQ Book Frequency Focal points TGC Depth Auto optimize Gain
(toggle) Le TGC Depth Auto optimize Gain
17
Archiving ‘P’ buttons can be programmed Default settings:
P1: Saves as raw dicom clip or still image (image appearance, M-mode, 3-D can be done on a raw dicom clip, but it is a very large file, thus too large to send) P2: Print on the thermal printer P3: Save as dicom
18
Archiving Images can be transferred to a CD or USB flash drive.
Depending on GE model, media must be formatted first Press Utilities button Select Connectivity, then removable media To eject media after saving images, press F3 Select the media to eject, then finalize
19
Some GE transducers 8C 12L Rectal i739 4C 3S
20
SRI Speckle Reduction Imaging: Improved contrast resolution
Improved overall clarity Smoothes regions where no feature or edges appear Maintains or enhances edges and borders Does not eliminate any information Increased Signal-to-Noise ratio
21
SRI Always is on in the LOGIQ e
Can adjust to have different levels in the LOGIQ 7 and LOGIQ P5 Level is set within the preset (Application) settings
22
Spatial compounding CrossXBeam Interrogates tissue at multiple angles
= GE’s spatial compounding Available on the LOGIQ e and LOGIQ P5 Interrogates tissue at multiple angles Introduces new tissue info Averages out speckle SRI and CrossXBeam together can have complimentary and additive effects Borders are similar for each steering angle, noise and speckle position change for each frame; successive frames average out noise and smooth borders
23
CrossXBeam Perpendicular angle of insonation generates higher signal
Images generated with compounding have more angles that are perpendicular to structures CrossXBeam OFF CrossXBeam ON
24
CrossXBeam Improved border definition Improved contrast resolution
Reduced angle dependence Reduced speckle and noise Increased visualization of biopsy needles Can reduce helpful artifact Can reduce frame rate LOGIQ P5 LOGIQ e
25
CrossXBeam: P5 Press X Beam to turn on
Press L and R buttons simultaneously to see side-by-side display of CrossXBeam on and CrossXBeam off
26
CrossXBeam: Le F9 or ‘Comp’ to turn on
Press L and R buttons simultaneously to see side-by-side display of CrossXBeam on and CrossXBeam off
27
Harmonics (THI) Display is from echos of a frequency different from the transmitted frequency (i.e. 2x transmitted frequency) Reduced artifact Improved axial and lateral resolution Better margin definition and increased contrast
28
Press THI or Harmonics button to turn on
Harmonics (THI) Press THI or Harmonics button to turn on Press L and R buttons simultaneously to see side-by-side display of harmonics on and harmonics off
29
3-D Volume imaging Appears more ‘life-like’ Infinite viewing planes
Volume vs. area affected can be assessed Appears more ‘life-like’ Infinite viewing planes May approximate CT or MRI imaging Scan parallel* or sweep (fan) Angle of imaging less operator dependent LOGIQ P5 LOGIQ e
30
3-D: P5 Press 3D to turn on Click arrow to go to images
Press Left button to start. Scan anatomy. Press Right button to stop.
31
3-D: Le Press F7 to turn on Click arrow to go to images
Press Left button to start. Scan anatomy. Press Right button to stop.
32
Use set button to alternate hand and rotate tools
Drag image to corner
33
Use set button to alternate hand and rotate tools
Rotate image
34
Use set button to alternate hand and rotate tools
Slice through
35
Use set button to alternate hand and rotate tools
Rotate again
36
3-D Use set button to alternate hand and rotate tools
Slice through another plane
37
LOGIQView Extended field of view
Better demonstration of anatomical relationships Ability to measure large structures LOGIQ P5 LOGIQ e
38
LOGIQView: P5 Press LOGIQ View to turn on
Press Left button to start, scan anatomy. Press Right button to stop.
39
Press Left button to start, scan anatomy. Press Right button to stop.
LOGIQView: Le Press F8 View to turn on Press Left button to start, scan anatomy. Press Right button to stop.
40
Scan in a straight plane or image will ‘bend’
LOGIQView Scan in a straight plane or image will ‘bend’
41
% affected Measure % of tendon affected by the lesion
Press Measure button Select % stenosis folder Select % sten (area) Trace around entire tendon Press set Trace around lesion % affected = %sten