Download presentation
2
Corneal topography orbscan
S.A.A MORTAZAVI MD Associate professor of ophthalmology Feiz hospital 1390
3
ORBSCAN SYSTEM Use the principle of projection
Forty scanning slit beams (20 from the left and 20 from the right with up to 240 data points per slit ) to scan the cornea and measure independently the X,Y & Z locations
4
Orbscan imaging Forty slit images are acquired in two 0.7 second periods Each of the 40 slit images triangulates one slice of ocular surface Distance between data slices average 250 microns
6
ORBSCAN Orbscan I only slit scan topography
Orbscan II the placidodisc added in orbscan I
8
ORBSCAN The images used to construct the anterior corneal surface,posterior corneal surface,anterior iris and anterior lens surfaces Data regarding the corneal pachymetry and anterior chamber depth
9
Elevation Orbscan measure elevation
Elevation is important the only complete scaler measure of surface shape Both slope & curvature can be mathematically derived from a single elevation map
10
BEST FIT SPHERE (BFS) The computer calculates a hypothetical sphere that matches as close as possible to the actual corneal shape being measured Compares the real surface to the hypothetical sphere showing areas above the surface of the sphere in warm colours and areas below the surface in cool colours
11
Topography quad map The upper left : anterior float
The upper right : posterior float The lower left : keratometric pattern The lower right : pachymetry map
15
NORMAL BAND SCALE Highlights the abnormal areas in the cornea in orange to red colors The normal areas are all shown in green Helpful in generalized screening in preoperative examination
17
AXIAL MAP Provides detailed keratometric information across the diameter of the cornea K readings are between certain values the cornea must be neither too steep nor too flat
18
AXIAL MAP To create a good quality corneal flap in LASIK if either extremes (too steep or too flat) is the case, this can lead to surgical flap complications K readings of more than 48 D are an indication of potential keratoconus
21
Middle box Keratometric readings White to white distance in mm
Angle kappa readings The thinnest point of cornea irregularity within the central 3 mm & 5 mm
23
PACHYMETRY MAP The orbscan measures thickness from the tear film layer to descemet’s membrane and is thicker than that obtained with ultrasound Adjustment factor (acoustic factor) ,the default setting is 92% Provides a reading showing the thinnest point of the cornea that may not necessarily be the central reading
25
PACHYMETRY MAP Thinnest point <470 micron
In pathological corneas, thinnest point is often displaced inferotemporal Difference of >100 microns from the thinnest point to the values at 7mm optical zone
26
ELEVATION MAP The green colour is referred as refrence sphere (at sea level ) The warmer colours are above this level and the cooler colours are below
29
ANTERIOR ELEVATION MAP
Looking at a proper scale in the cornea , can see height differences Compare the height of the actual cornea to a best fit sphere
30
Posterior map The highest elevation value as a keratoconus indicator or at least as a screen for patients may be at risk of developing keratectasia 55 D elevation as an absolute cut off
31
ELEVATION DATA The difference between the highest and lowest points is a potential keratoconus indicator if over 100 microns (Rousch criteria)
32
DIAGNOSTIC CRITERIA Power map changes Posterior elevation maps
Pachymetry Composite/integrated topography information
33
POWER MAP Mean corneal power >45D
In addition to steep corneal curvatures-the bowtie or broken bowtie appearance indicative of early keratoconus Central corneal asymetry a change within central 3mm optical zone of the cornea of more than 3D
34
Irregularity in central cornea
Greater than 1.5 D in 3 mm zone and greater than 2.0 D in 5 mm zone is considered abnormal and cause for concern
35
POSTERIOR ELEVATION MAP
Many surgeons think the first sign of keratoconus appears on the posterior surface of the cornea 3.13% of population screened for laser surgery had posterior ectasia criteria by orbscan , despite having axial topography classified as normal
37
POSTERIOR ELEVATION MAP
The most common reference surface for viewing elevation maps is the best fit sphere A best fit sphere (BFS) >55D on the posterior profile , indicative of posterior ectasia
38
Posterior float difference
Greater than 50 micron generally accepted as abnormal In corneas thinner than normal over 40 as abnormal
39
Posterior elevation map
The location of the steepest part of the posterior float should be relatively central , but is a more concern it be located away from the center and in an area of corneal thinning Posterior float difference;40 to 50 microns seems to be the maximum difference
41
Correlation of signs of the highest point
Highest point on the posterior elevation coincides with the highest point of anterior elevation , the thinnest on pachymetry and the point of steepest curvature on the power map
43
Although high posterior elevation and ratio between two elevation maps rarely used as exclusion criteria alone , but by considering these together , more conclusive information can be obtained
44
Risks of ectasia indices
Number of abnormal maps Posterior float difference >0.050 3mm & 5mm irregularity Peripheral thickness changes Astigmatism variance between eyes Steep k’s –mean power map
46
Three step rule One abnormal map ; perform with caution
Two abnormal map ; with concern Three abnormal map ;contraindicated
50
Composite/integrated information
Similarly between anterior & posterior profiles a forward bending of areas shown above the BFS and association with the thinnest point on the cornea Inferotemporal displacement of the highest point
53
Abnormal tear film Can significantly distort the readings
The significant change in surface quality and validity of the dry eye
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.