Download presentation
1
David E. Hintenlang, Ph.D., DABR
CT Scanner QC Presented by: David E. Hintenlang, Ph.D., DABR Medical Physics Program Director Department of Nuclear and Radiological Engineering Department of Biomedical Engineering American College of Medical Physics 2004 Annual Meeting Scottsdale Arizona
2
CT Quality Control - Motivations
Appropriate diagnostic benefit vs risk Image Quality Noise Contrast Spatial resolution Radiation Risk Dosimetry indices (ie. CTDI)
3
Presentation Overview
Introduction & Background General categories of QC tests Specific QC tests and procedures QC Program recommendations & comparisons Acceptance tests “Annual” QC tests “Daily” QC tests Conclusions & References
4
Background Expanded CT capabilities over last decade
Helical and Multislice CT Decreases scan times - Changes in patient dose Increased utilization of CT Complex systems, high utilization, large contribution to patient dose demands a continuous and robust QA program
5
QC Objectives Complexity of system offers opportunities for many individual QC tests Level of testing balanced against QC Objectives QC Categories: Acceptance Testing Annual Testing Daily QC ACR Accreditation
6
Categories of QC Tests Safety Mechanical accuracy Imaging Performance
Display System
7
General QC Tests : Safety
Electrical Mechanical Laser Radiation
8
General QC Tests: Mechanical Accuracy
Ensure that aligned patient anatomy is imaged Alignment light accuracy Alignment of table and gantry Table and gantry positioning Slice localization from scout images
9
General QC Tests: Image Quality
Noise and Field Uniformity CT Number Linearity Low Contrast Detectability Spatial Resolution Display and Hard Copy Image Quality
10
General QC Tests: Dosimetry
CTDI Head & Body Patient Dosimetry
11
Alignment Laser Accuracy
Internal & External scan plane lights Verify correct distance between internal & external scan plane lights Wrap film around centered phantom Make pin hole marks along illuminated lines
12
Scan 1 slice @ 1mm thickness
Repeat for external alignment laser Measure distance between line of pierced holes and center of exposed slice Misalignment should be less than 2 mm
13
Table - Gantry Alignment
To ensure that longitudinal axis of table is aligned with isocenter Measure to locate & mark central axis of table at: head end middle foot end Hang plumb bob from center of gantry - to table top Measure distances between plumb bob and central axis of table Repeat for various gantry angles Table should be aligned within +/- 5 mm
14
Table Incrementation Accuracy
Evaluate accuracy/ reproducibility of longitudinal table motion Tape ruler along table edge Tape pointer on frame around mid-point of ruler Load table Move table mm in one direction & return Repeat for a total of 3-times in each direction Average & Standard Deviation should be less than 3 mm
15
Table Incrementation Accuracy & Collimation (axial scan)
Evaluate accuracy & reproducibility of longitudinal table motion & width of x-ray beam Wrap film around centered head phantom Scan 5 slices for several combinations of slice thickness and gap. Measure slice thickness and gaps from film Mark table frame at one end of the phantom Return to original table position and note any difference from original position Interslice gap should have < 0.5 mm discrepancy Slice thickness should be accurate to within 1 mm
16
Table Incrementation, scanned volume, helical pitch accuracy
Evaluate accuracy & reproducibility of longitudinal table motion & helical pitch Wrap film around centered head phantom Mark table frame at one end of phantom Record indicated table position Obtain volume scans for various slice thicknesses and pitch Measure from film: Total length scanned Spiral slice thickness Spiral gaps (pitch)
17
kVp Accuracy Lock x-ray tube position
Center voltage divider using CT specific filer pack in center of x-ray beam Check kVp’s at : 80, 100, 140 kVp Tube potential should be within +/- 5%
18
Half-Value Layer Lock x-ray tube position
Set scanning parameters per manufacturer (80 kVp and mAs to obtain 200 mR) Mount CT chamber parallel to table long axis, centered on isocenter Vary Al thicknesses Should be at least 2.3 mm 80 kVp
19
Exposure Reproduceabiity & Linearity
Head phantom in head holder on table Insert CT ion chamber in center of phantom Perform 4 scans at 80 kVp, 100 mA, 1s Record exposure for each Repeat at: 200 mA, 300 mA Repeat at : 120 kVp, 140 kVp
20
Radiation Profile Width
Evaluates the accuracy of pre-patient collimator settings Tape film around centered head phantom positioned flat on table with 0o gantry angle Scan one slice at each slice thickness Lowest kVp and tube current available space about 2 cm apart Measure width of each slice Should be within 1 mm of selected thickness
21
Slice Sensitivity Profile
Evaluate the actual width of the imaged slice Set standard scanning parameters Scan phantom insert with “ramps” for each available slice thickness For each image determine average pixel values of: a) metal ramp and b) acrylic background Set window width < 10 HU of min width setting Set window level at average of ramp and background values Measure width of ramp and multiply by tanq
23
Image Quality Measures
Noise and Field Uniformity CT Number Linearity Low Contrast Detectability Spatial Resolution Reconstruction times Scout view accuracy
24
Phantoms & Test Tools Testing phantoms available from manufacturers & third parties Early CT phantoms may not provide sensitivity require to test modern CT systems
25
Noise & Field Uniformity Axial Scan
Assess noise level under simulated clinical conditions Set standard scanning parameters Position a centered homogenous phantom Scan and select ROI’s Obtain mean pixel value and standard deviation for each ROI Std. Dev. Central ROI < 10 HU Max difference between peripheral and central ROI’s < +/- 5 HU
26
Noise & Field Uniformity Volume Scan
Assess noise level under simulated clinical conditions Set standard scanning parameters Position a centered homogenous phantom For each voume scanned measure 3 reconstructed slices center & both ends of volume Scan and select ROI’s Obtain mean pixel value and standard deviation for each ROI Std. Dev. Central ROI < 10 HU Max difference between peripheral and central ROI’s < +/- 5 HU
27
CT Number Linearity CT numbers should be consistent from system to system Set standard scanning parameters Image phantom insert with 5 rods of different materials Determine mean pixel values of each object & background Perform for low, mid, & high kVp settings CT number should be within 50 HU for each material
28
Low Contrast Detectability
Evaluate scanner low contrast object discrimination Set standard scanning parameters Scan low-contrast detectability insert Determine mean value of each hole & surrounding material Contrast = difference in HU of objects & background Determine Number of objects visible Should meet manufacturer’s specifications
29
Spatial Resolution Evaluate high contrast spatial resolution
Set standard scanning parameters Scan Spatial Resolution Insert Determine highest frequency group visually resolved for each image Should meet manufacturer specifications
30
Spatial Resolution MTF
Derive the MTF as a measure of limiting spatial resolution Scan spatial resolution insert For each group of high contrast resolution patterns determine the mean value & std. deviation. Determine & plot MTF Repeat using bone reconstruction filters Should meet manufacturer’s specifications
31
Image Reconstruction Time
Evaluate reconstruction times Set parameters for typical head scan Scan anthropomorphic head phantom Measure time required to obtain 1 image slice Subtract out data collection time Repeat for anthropomorphic chest phantom and typical body slice Should meet manufacturer specifications
32
Scout Views Evaluate scout view production
Position anthropomorphic head phantom Perform scout view at 120 kVp, 160 mA, 1 mm thickness Scan length 0 500 mm Perform scout views for 0, 45 and 90 degree gantry angles Record time required to perform and print images Repeat with anthropomorphic chest phantoms
33
Display and Hard Copy Image Quality
Ensure that information in the video signal is accurately transferred to display media Reduce room illumination to normal viewing level Display SMPTE Pattern Adjust brightness and contrast controls 5% and 95% patches should be visible
34
Dosimetry Measurement: CTDI
Determine a measure of radiation dose to tissue under different scan conditions Set standard scanning parameters (i.e. head) Position CTDI head phantom Scan with CT ion chamber inserted sequentially in each of 5 positions Calculate CTDI using peripheral and central values Repeat with CTDI body phantom Repeat for various scanning parameters as desired
35
CT Equipment Quality Control Program
ACR Accreditation Program for CT QC program established & implemented under supervision of a qualified medical physicist Requirements Acceptance testing upon installation Annual medical physicist survey Continuous Quality Control
36
Acceptance Testing Series of measurements performed to verify that a CT system conforms to required specifications Specific tests/technical specifications should be included in bid request or purchase agreement Many more combinations of operating parameters exist than can be practically tested Must acquire sufficient data to adequately characterize system performance
37
Considerations for Acceptance Tests
Determine most important performance specs to test in view of: Clinical needs Scanner capabilities Devise a short list of 4-5 “standard” scanning conditions: Standard body Standard head Spine Pediatric Body Pediatric Head Pertinent performance tests should be performed for each standard condition to provide a performance baseline Select a few additional scanner conditions that test performance limits
38
AAPM Report #39 Recommendations
Essential Tests Localization light accuracy Table Incrementation Radiation Profile Sensitivity Profile Image Noise Field Uniformity High Contrast Resolution Display & Hard Copy Image Quality Radiation Dose Optional Tests Alignment of table to gantry Table gantry tilt mAs Linearity kVp Accuracy Scan Field Dependence (QCT) Alternative Tests Slice localization from Scout image High contrast resolution – MTF Low contrast detectability
39
Annual ACR Requirements
Localization light accuracy Alignment of table to gantry Table gantry tilt Slice localization from Scout image Table Incrementation Slice thickness -Sensitivity Profile Image Noise Field Uniformity High Contrast Resolution Low Contrast Detectability Artifact Evaluation CT Number Accuracy & Linearity Display & Hard Copy Image Quality Radiation Dose – CTDI Patient Radiation Dose for select exams Safety Evaluation Visual inspection Audible/visual signals Posting requirements Scattered radiation measurements Other regulatory tests
40
Monthly to Semi-Annual Tests per AAPM Report # 74
Localization light accuracy Slice localization from Scout image Table Incrementation Slice thickness -Sensitivity Profile Image Noise Field Uniformity High Contrast Resolution Low Contrast Detectability Artifact Evaluation CT Number Accuracy & Linearity Display & Hard Copy Image Quality Radiation Dose – CTDI Dosimetry from Scout Image Safety Evaluation Visual inspection Audible/visual signals Posting requirements Scattered radiation measurements Specified by ACR but not AAPM 74 Patient Radiation Dose for select exams Alignment of table to gantry Table gantry tilt
41
Continuous Quality Control
Required for all CT units Qualified medical physicist should determine tests, frequency, and responsible individual Facility Considerations CT Usage On-site radiological technologist should be identified with responsibility for conducting routine QC
42
Continuous Quality Control Tests
Localization light accuracy Slice thickness Image Noise Field Uniformity High Contrast Resolution Low Contrast Detectability Artifact Evaluation CT Number Accuracy Display & Hard Copy Image Quality
43
Daily Tests- AAPM Report #74
CT Number Accuracy Image Noise Field Uniformity Artifact Evaluation All can be obtained from same CT phantom scans Sensitive to a wide variety of scanner problems
44
Aids to Daily QC i.e. AutoQA LiteTM 1
A software program to evaluate and monitor performance of CT scanners Provides database tool for tracking a limited number of tests: Image Noise CT Number accuracy Field Uniformity Slice Width Spatial Resolution 1. The Institute for Radiological Image Sciences, Inc. , MedSoft Technologies, Inc.
45
AutoQA Lite Overview Provides automated analysis of CT phantom images
CatPhan series ACR Accreditation phantom Toshiba InnerVision Vendor specific (GE, Siemens, and Philips) Operates on PC with Dicom 3.0 CT image formats
46
L
47
Conclusions ACR provides an excellent outline for CT QC Programs
A wide variety of tests exist Medical Physicist has latitude to define requirements for “continuous” QC testing
48
CT QC References: American College of Radiology Website
(links to “accreditation”) AAPM Report No. 39, Specification and acceptance testing of computed tomography scanners, American Institute of Physics, 1993. AAPM Report No. 74, Quality control in diagnostic radiology, Medical Physics Publishing, 2002. ImPACT, CT Scanner Acceptance Testing, Version 1.02, 2001 The ACR CT Accreditation Program and the Medical Physicist, Maynard High, ACMP 2003 Annual Meeting
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.