Download presentation
Published byGeorgiana Lindsey Modified over 9 years ago
1
Modes of Pacing Seoul National University Hospital
Department of Thoracic & Cardiovascular Surgery
2
Electrophysiology of Pacing
Stimulation & depolarization of myocardial tissue Sensing of intramyocardial electrical activity Current pacemaker technology Pulse generator
3
Stimulation & Depolarization of Myocardial Tissue
The myocardium must be excitable The stimulus current density (current per unit cross-sectional area) must be sufficiently high & of sufficient duration to depolarize a group of cells that is large enough to initiate impulse propagation in the myocardium The pacemaker-generated impulse then relies on the intrinsic properties of cardiac specialized conduction & myocardial tissue for depolarization of the entire heart
4
Factors affecting Sensing
Sensing is the detection of real or spontaneous cardiac depolarization Electrode size Configuration of electrode (unipolar, bipolar) Position of the lead tip within the heart
5
Pacing Lead Technology
Attachment of electrode Active fixation Passive fixation 1. Chronic ventricular pacing thresholds tend to be lower with passive lead, in part because of tissue injury with active fixation 2. Sensing characteristics are similar between active & passive leads 3. Epicardial electrodes historically have poorer performance over long term than endocardial lead
6
Pulse Generator Power source Time circuitry Sensing circuitry
Output circuitry for channels connected to the electrodes Transceiver for telemetric communication with programming device
7
Single Chamber Pacemaker
AOO AAI AAT VOO VVI VVT
8
Single Chamber Pacemaker
AAIR VVIR
9
Single Chamber Pacemaker
VVI I *
10
Intrinsic P and R waves
11
Atrial Sensing / Inhibition
12
Ventricular Sensing / Inhibition
13
Pacing Interval / Pacing Rate
The rate at which the pacemaker will pace if the patient does not have their own rhythm Expressed in either PPM or ms
14
Rate Conversion Conversion Pacing rate in PPM divided into 60,000 = ms
60,000 / 60 PPM = 1000 ms Interval in ms divided into 60,000 = PPM 60,000 / 1000 ms = 60 PPM
15
Ventricular Sensing / Inhibition
pacing interval
16
Capture Definition The depolarization and resultant contraction of the atria or ventricles in response to a pacemaker stimulus.
17
Threshold Stimulation Threshold Definition :
The minimum amount of electrical energy that consistently produces a cardiac depolarization
18
Atrial Capture Pacemaker Stimulus Atrial Capture
19
Ventricular Capture Ventricular Capture (paced ventricular
beats look like PVC’s or LBBB)
20
Refractory Period Refractory Period
21
Alert Period Definition The portion of timing cycle where the device senses electrical activity (e.g. cardiac) & responds in a preset or programmed manner
22
Alert Period Refractory Period Alert Period
23
Dual Chamber Pacemaker
VAT VDD DOO DVI DDI DDD
24
Dual Chamber Pacemaker
DDD T / I * I
25
Atrial & Ventricular Pacing
26
Sense in Atrium & Pace in Ventricle
27
Pace & Sense in Atrium & Pace in Ventricle
28
Pace Atrium & Sense Ventricle
29
Pace and Sense in Atrium Sense in Ventricle
30
AV / PV Delay AV Delay PV Delay
31
Post Ventricular Atrial Refractory Period (PVARP)
Definition Timeframe the atrial channel is refractory, initiated when there is either a ventricular pacer spike or sensed R-wave. PVARP
32
Atrial Alert Period Definition Timeframe after PVARP when the sense amplifier is open and can see P-waves
33
Atrial Alert Period Alert Period PVARP AV Delay PV Delay
34
Ventricular Refractory Period
35
Ventricular Alert Period
Definition Timeframe after the ventricular refractory period that the sense amplifier is open looking for R-waves
36
Ventricular Alert Period
Ventricular Refractory Period Ventricular Alert Period
37
Dual Chamber Pacing Minimum Rate or Base Rate
Lowest rate that the pacemaker will pace in the Atrium Maximum length of time the Pacemaker will wait for intrinsic activity
38
Dual Chamber Pacing Minimum Rate
39
Dual Chamber Pacing Maximum Rate / Maximum Tracking Rate
The fastest rate that the Ventricular channel can track intrinsic P-waves
40
Maximum Tracking Rate
41
Pacemaker Follow-up Aim Interval
Verify appropriate pacemaker operation Optimize pacemaker functions Interval Immediate and POD # 1 day Postoperative 3, 6, 12 months
42
Pediatric Aspects of Cardiac Pacing
Ventricular epicardial leads Thoracotomy approach Transdiaphragmatic approach Subcostal or subxiphoid incision Atrial epicardial leads Base of left atrial appendage Transvenous leads All the leads have the problem with child growth and results in high threshold, exit block, lead fracture. Even when a loop of lead was left in the pericardium, adhesions frequently formed that prevented unwinding of the loop.
43
Interrogation
44
Assess Current Pacemaker Function
Compare the surface ECG to markers to determine appropriate pacing (capture) & sensing
45
Measured Data Provides information on:
-Magnet Information, -Lead Status, -Battery Status
46
Diagnostics
47
Pacemaker Testing Sensing and Capture testing
should be performed to insure adequate safety margins are programmed
48
Ventricular Threshold Test
Sensing and Capture testing should be performed to insure adequate safety margins are programmed
49
Loss of Capture Definition:
The emitted pacemaker stimulus does not cause depolarization and resultant cardiac contraction
50
Ventricular Loss of Capture
Programmed Rate Programmed Rate
51
Problem Solving Loss of Capture Causes Dislodged Lead Insulation Break
Program Voltage higher Program Pulse Width higher Reposition pacing electrode
52
Sensing Threshold The sensitivity number is increased until loss of sensing is seen Then the sensitivity number is decreased until sensing is regained This is the Sensing Threshold
53
Ventricular Sensitivity Test
54
Oversensing Definition :
The sensing of events other than P or R-waves by the pacemaker circuitry
55
Ventricular Oversensing
Pacing interval Pacing interval Pacing interval Pacing interval Pacing interval
56
Causes of Oversensing Insulation Break Intermittent Lead Fracture
Myopotentials EMI Problem Solving Program sensitivity to a higher number Program the refractory period longer
57
Undersensing Definition :
Failure of the pacemaker circuitry to sense intrinsic P- or R-waves
58
Ventricular Undersensing
non-sensed R-waves
59
Causes of Undersensing Dislodged Lead Insulation Break Problem Solving Program the sensitivity to a lower number
60
Testing of Chronic Pacing System
Current Voltage Resistance Threshold Threshold Dislodgement of electrode from endocardium H H N Perforation H H N Exit block (fibrosis around electrode tip) H H N Pulse generator malfunction N N N Lead/connector abnormality N N N Battery depletion N N N Partial lead fracture L to N H H F F Complete lead fracture N H H Lead insulation fracture H H L H; high, N; normal, L; low. F; fluctuating
61
Follow-up Check Interrogate Measure Impedance Check Diagnostics
Check Sensing Measure Threshold
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.