Presentation is loading. Please wait.

Presentation is loading. Please wait.

Alex Bode Emmanuel Cordova Mikel Dualos Christa Hamilton Medical Instrumentation Spring 2011.

Similar presentations


Presentation on theme: "Alex Bode Emmanuel Cordova Mikel Dualos Christa Hamilton Medical Instrumentation Spring 2011."— Presentation transcript:

1 Alex Bode Emmanuel Cordova Mikel Dualos Christa Hamilton Medical Instrumentation Spring 2011

2 Overview Clinical aspects Interface design Electronic module Video Presentation Design Issues and Observations Lessons Learned

3 Clinical Aspects Morphology of Pulse Waveform The wave originates as left ventricle contracts. Aortic valve opens. Rapid rise in pressure as the aortic valve opens and the blood is pushed out of the left ventricle into the aorta. Fall in pressure in left ventricle until the aortic valve closes when aortic pressure exceeds that of the ventricle.

4 Clinical Aspects Morphology of Pulse Waveform

5 Clinical Aspects Finger Tip Blood Flow

6 Clinical Aspects Finger Tip Blood Flow

7 Clinical Aspects Optical properties of Hemoglobin

8 Clinical Aspects Optical Plethysmography

9 Mechanical Interface Design Phototransistor Semiconductor device used to capture light. Converts light energy into electrical energy. When light from a source object falls on the unit, it is amplified and passed to the base collector diode. Works best around 800 nm.

10 Mechanical Interface Design IR vs. Red LED Both signals required much time and manipulation to produce Red LED took less time depending on the individual Red LED chosen since it passed trough skin easier/ wavelength not as high as Infrared Infrared Red LED

11 Mechanical Interface Design Transducer Mount Configuration 1 Configuration 2

12 Electronic Module Design Schematic *tape to keep offset potentiometer in place

13 Band-Pass Filters Normal resting Heart rate is 70 bpm=1.17 Hz Range of 20 bpm to 200 bpm 0.33 Hz to 3.33 Hz High pass filter=0.33 Hz Low pass filter=3.33 Hz

14 Frequency Response Curve Fc=1/(2*pi*R*C) R=resistor value (ohms) C=capacitance (Farads) HP=0.33 Hz when C=10 uF R=47k ohms LP=3.33 Hz when C=0.47 uF R=100k ohms

15 Band-Pass Filter The capacitive and resistive input components make up the high-pass filter. The capacitive and resistive feedback components make up the low-pass filter. Rf/Rx+1=Gain

16 Operational Amplifier Gain Output range for the LED bar graph display is.2 to 1.2 Volts. Signal peak to peak is 20 mV. (1.2-.2)/20m=50

17 Schematic

18 LED Bar Graph Display

19

20 Video Presentation

21 Design Issues and Observations Hand positing was very particular. Very time consuming to get a signal Since it was difficult to get a signal, using the EKG waveform allowed us to continue working on the other segments of the circuit during development Some noise was still visible in final output signal Use of infrared LED may have reduced noise

22 Lessons Learned O No pressure should be applied onto skin from either transducer or LED, reduces capillary volume O 10 bar LED display type depends on pin 9 connection. The collector voltage must be around 2.5 volts (half input voltage) in order for the photo transistor to properly transmit signal. (R1 = 10M  )

23 Lessons Learned When choosing resistor and capacitor values, it is important to take into account the time constant RC. this affects how long it takes for 10 bar LED to move past the transient phase where all bars are lit and slowly decline.

24 Group 6


Download ppt "Alex Bode Emmanuel Cordova Mikel Dualos Christa Hamilton Medical Instrumentation Spring 2011."

Similar presentations


Ads by Google