Download presentation
Presentation is loading. Please wait.
Published byPreston Miller Modified over 9 years ago
1
Hollow Cathode Light Source at MSU Aaron Kingery MSU 2009 REU
2
MOSES Multi-Order Solar EUV Spectrograph Sounding rocket payload Takes images and spectrum simultaneously
3
Multi-Order Grating
4
Extreme Ultraviolet
5
Hollow Cathode Lightsource (HCL)
7
"The Test" 1.Pump the intermediate chamber and HCL chamber down 2.Isolate the HCL chamber 3.Record the pressure in the HCL as a function of time
8
First Tests
10
Ethanol Test 1: No ethanol 2: Ethanol on top 3: Ethanol on bottom 4: No ethanol
11
Virtual Leaks
13
Stabilization Pressure
14
How Do You Find the Leaks? Remove things from the system Check for leaks If the leak is still present, remove more things.
15
How Do You Find the Leaks? Problems: Can't remove the pressure gauge or valve Can't rule out anything if the leak persists.
16
Large Vacuum Chamber Will be used to test MOSES Large enough to put the MOSES payload inside Needs to be clean
17
Residual Gas Analysis Scan
18
Photo diode It was held into possition by the pins The electical connections was also made by the pins. Pin broke off We had make a new way to it to be held and allow it to send a signal Upon testing it was found to be broken
19
To-Do List Get new photo diode Turn on the HCL Test the HCL using Ne Calibrate and Test MOSES
20
Conclusions HCL is almost free of leaks We need a new photo diode to test the HCL Overall a very productive summer
Similar presentations