EXPOSURE AND TECHNIQUE ERRORS
DA 105
2
Dental Radiography Questions
What film exposure and technique errors may be observed on dental radiographs? What horizontal and vertical angulation errors may be observed on dental radiographs? What is the proper film placement for bite-wing radiographs?
3
Dental Radiography Exposure and Technique Errors Film exposure errors
Technique errors (periapical films) Technique errors (bite-wing films) Miscellaneous technique errors
4
Exposure and Technique Errors
Purpose To describe film exposure problems and periapical, bite-wing, and miscellaneous technique errors
5
Film Exposure Errors Exposure Problems
Unexposed film Film exposed to light Time and Exposure Factor Problems Overexposed film Underexposed film
6
Unexposed Film Appearance The film appears clear Cause
Failure to turn on the x-ray machine Electrical failure Malfunction of the x-ray machine Correction Make certain the x-ray machine is turned on and listen for the audible exposure signal
7
An unexposed film appears clear
8
Film Exposed to Light Appearance Cause Correction
The film appears black Cause The film was exposed to white light Correction Do not unwrap in a room with white light Check the darkroom for light leaks Turn off all lights in the darkroom except the safelight
9
A film exposed to light appears black
10
Overexposed Film Appearance Cause Correction Film appears dark
Excessive exposure time, kilovoltage, milliamperage Correction Check settings and reduce as needed before exposing film
11
An overexposed film appears dark
12
Underexposed Film Appearance Cause Correction The film appears light
The film was underexposed Correction Check the exposure time, kilovoltage, and milliamperage settings on the x-ray machine before exposing the film
13
An underexposed film appears light
14
Technique Errors (Periapical Films)
Film Placement Problems Correct film placement Incorrect film placement Absence of apical structures Dropped film corner Angulation Problems Incorrect horizontal angulation Incorrect vertical angulation PID Alignment Problems Cone-cut with Film Holder Cone-cut without Film Holder
15
Correct Film Placement
Each periapical film must be positioned in a certain way to show specific teeth and related anatomic structures The edge of the periapical film must be placed parallel to the incisal or occlusal surfaces of the teeth and extend 1/8th inch beyond the incisal or occlusal surfaces
16
Correct periapical film placement demonstrates the entire tooth, including the apex and surrounding structures
17
Absence of Apical Structures
Haring (p. 276) (Fig. 20-6) Appearance No apices appear on the film Cause The film was not positioned in the patient’s mouth to cover the apical regions of the teeth Correction Make certain no more than 1/8th inch of the film edge extends beyond the incisal-occlusal surfaces of the teeth
18
Improper film placement; no apices appear on this film
19
Dropped Film Corner Appearance
The occlusal plane appears tipped or tilted Cause The edge of the film was not placed parallel to the incisal-occlusal surfaces of the teeth Correction Make certain the edge of the film is placed parallel to the incisal-occlusal surfaces of the teeth
20
Improper film placement; a dropped film corner is seen when the edge of the film is not placed parallel to the incisal or occlusal surfaces of the teeth
21
Incorrect Horizontal Angulation
Appearance Overlapped contacts Cause The central ray was not directed through interproximal spaces Correction Direct the x-ray beam through interproximal regions
22
Incorrect horizontal angulation results in overlapped contact areas
23
Incorrect Vertical Angulation
Foreshortened images Appearance Short teeth with blunted roots Cause Excessive vertical angulation Correction Do not use excessive vertical angulation with the bisecting technique
24
If the vertical angulation is too steep, the image of the tooth on the film is shorter than the actual tooth; the images are foreshortened
25
Incorrect Vertical Angulation
Elongated images Appearance Long, distorted teeth Cause The vertical angulation was insufficient Correction Use adequate vertical angulation with the bisecting technique
26
If the vertical angulation is too flat, the image of the tooth on the film is longer than the actual tooth; the images are elongated
27
PID Alignment Problems
Cone-cut with film holder Appearance A clear area appears on the film Cause The PID was not properly aligned with the periapical film holder Correction Make certain the x-ray beam is centered over the film
28
A cone-cut is seen when the PID is not properly aligned with the periapical film-holding device
29
PID Alignment Problems
Cone-cut without film holder Appearance A clear area appears on the film Cause The PID was not directed at the center of the film Correction Make certain the x-ray beam is centered over the film
30
A cone-cut appears as a curved unexposed (clear) area on a radiograph
31
Technique Errors (Bite-Wing Films)
Film Placement Problems Correct film placement Incorrect film placement Angulation Problems Incorrect horizontal angulation Incorrect vertical angulation PID Alignment Problems Cone-cut with film holder Cone-cut without film holder
32
Correct Film Placement
Premolar bite-wing The resulting film shows both the maxillary and mandibular premolars and distal contact areas of both canines Molar bite-wing The resulting film shows both the maxillary and mandibular molars
33
Correct film placement for the premolar bite-wing
34
Correct film placement for the molar bite-wing
35
Incorrect Film Placement
Premolar bite-wing Appearance Distal surfaces of the canines are not visible on the film Cause The bite-wing film was positioned too far posteriorly in the mouth Correction Make certain the anterior edge of the bite-wing film is positioned at the midline of the mandibular canine
36
Incorrect film placement for the premolar bite-wing
37
Incorrect Film Placement
Molar bite-wing Appearance Third molar regions are not visible on film Cause The bite-wing film was positioned too far anteriorly in the mouth Correction Make certain the anterior edge of the bite-wing film is positioned at the midline of the mandibular second premolar
38
Incorrect film placement for the molar bite-wing
39
Incorrect Horizontal Angulation
Appearance Overlapped contacts on the film Cause The central ray was not directed through the interproximal spaces Correction Direct the x-ray beam through the interproximal spaces
40
Overlapped interproximal contacts result from incorrect horizontal angulation
41
Incorrect Vertical Angulation
Appearance Images appear distorted on the film Cause The vertical angulation was incorrect Correction Always use a +10-degree vertical angulation with the bite-wing technique
42
Incorrect vertical angulation causes the images to appear distorted
43
Cone-Cut With Film Holder
Appearance A clear area appears on the film Cause The PID was not properly aligned with the film holder Correction Make certain the PID and the aiming ring are aligned
44
A, A cone-cut is seen when the PID is not properly aligned with the bite-wing film-holding device. B, A cone-cut can also be produced with rectangular collimation; the PID is not properly aligned with the bite-wing film-holding device
45
Cone-Cut Without Film Holder
Appearance A clear area appears on the film Cause The PID was not directed at the center of the film Correction Make certain the x-ray beam is centered over the film
46
A cone-cut appears as a curved unexposed (clear) area on a radiograph
47
Miscellaneous Technique Errors
Film Bending Film Creasing Phalangioma Double Exposure Movement Reversed Film
48
Film Bending Appearance Cause Correction
Images appear stretched and distorted Cause The film was bent excessively Correction Check film placement before exposure
49
A bent film appears distorted
50
Film Creasing Appearance Cause Correction
A thin radiolucent line appears on the film Cause The film was creased Correction Do not bend or crease the film excessively
51
A film crease appears as a thin radiolucent line
52
Phalangioma Appearance Cause Correction Haring (p. 282) (Fig. 20-23)
The patient’s finger appears on the film Cause The patient’s finger was positioned in front of the film Correction Make certain the patient’s finger is placed behind the film
53
This film demonstrates a phalangioma; the bones of a patients finger appear on the film
54
Double Exposure Appearance Cause Correction
A double image appears on the film Cause The film was exposed twice in the patient’s mouth Correction Always separate exposed and unexposed films
55
This film demonstrates a double exposure
56
Movement Cause Correction Appearance Blurred images appear on the film
The patient moved during exposure of the film Correction Instruct the patient to remain still while the film is being exposed
57
Movement results in a blurred image
58
Reversed Film Appearance
Light images with a herringbone pattern appear on the film Cause The film was placed backward in the mouth and then exposed Correction Always place the white side of the film adjacent to the teeth
59
A reversed film appears light with a herringbone (or tire-track) pattern. The tire-track pattern is seen on the lead-foil backing within the film packet
60
EXPOSURE AND TECHNIQUE ERRORS
DIAGNOSTIC DENTAL RADIOGRAPH: 1.FILM EXPOSURE ERRORS: EXPOSURE PROBLEMS a) Unexposed film APPEARANCE CAUSE CORRECTION b) Film exposed to light TIME AND EXPOSURE FACTOR PROBLEMS a) Overexposed film b) Underexposed film
61
TECHNIQUE ERRORS(PERIAPICAL FILMS)
FILM PLACEMENT PROBLEMS ANGULATION PROBLEMS PID ALIGNMENT PROBLEMS
62
FILM PLACEMENT PROBLEMS
ERROR APPEARANCE CAUSE CORRECTION CORRECT FILM PLACEMENT INCORRECT FILM PLACEMENT a) ABSENCE OF APICAL STRUCTURES b) DROPPED FILM CORNER
63
ANGULATION PROBLEMS APPEARANCE CAUSE CORRECTION ERROR
INCORRECT HORIZONTAL ANGULATION INCORRECT VERTICAL ANGULATION a) FORESHORTENED IMAGE b) ELONGATED IMAGE
64
PID ALIGNMENT PROBLEMS
ERROR APPEARANCE CAUSE CORRECTION CONE-CUT WITH FILM HOLDER CONE-CUT WITHOUT FILM HOLDER
65
TECHNIQUE ERRORS(BITE-WING FILMS)
FILM PLACEMENT PROBLEMS ANGULATION PROBLEMS PID ALIGNMENT PROBLEMS
66
FILM PLACEMENT PROBLEMS
ERROR CORRECT FILM PLACEMENT 1)PREMOLAR BITE-WING 2)MOLAR BITE-WING INCORRECT FILM PLACEMENT
67
INCORRECT HORIZONTAL ANGULATION INCORRECT VERTICAL ANGULATION
ANGULATION PROBLEMS ERROR APPEARANCE CAUSE CORRECTION INCORRECT HORIZONTAL ANGULATION INCORRECT VERTICAL ANGULATION
68
PID ALIGNMENT PROBLEMS
ERROR APPEARANCE CAUSE CORRECTION CONE-CUT WITH FILM HOLDER CONE-CUT WITHOUT FILM HOLDER
69
MISCELLANEOUS TECHNIQUE ERRORS
APPEARANCE CAUSE CORRECTION 1.FILM BENDING 2.FILM CREASING 3.PHALANGIOMA 4.DOUBLE EXPOSURE 5.MOVEMENT 6.REVERSED FILM