Circuitry & Function Let’s go back to the starting point
X-RAY CIRCUITY RT 244 – 2012 Circuitry & Function Let’s go back to the starting point Contributions by Mosby, Thompson Publisher, Carlton, Bushberg, and the WWW.
2
The Control Console The control console is device that allows the technologist to set technical factors (mAs & kVp) and to make an exposure. Only a legally licensed individual is authorized to energize the console.
3
Operating Console has meters to measure
kVp, mA, & exposure time Modern units only display mAs Units with ACE’s will have a separate meter for mAs
4
Control Panel All the electric circuits connecting the meters and controls are at low voltage to minimize the possibility of shock.
5
Technique selection The process begins at the control panel such as the one shown. The technologist selects a technique which will specify the kVp, mA, exposure time, and focal spot . When the exposure switch is depressed, manipulation of electricity to produce x-rays begins. It is necessary to follow the sequence of events in two parts. the voltage through to the x-ray tube, then go back and follow the voltage and current through to the filament.
6
Control Panel Circuit Elements
POWER SUPPLY 60 Hz AC to 480 V Panel Power On/Off AUTOTRANSFORMER 1. line compensation a. line meter b. primary side adjustments 2. kVp selection a. secondary side adjustments variable turns ratio 3. filament circuit power 2a. 1a. 2a. 3. 1b.
7
Operating Console Controls:
Line Compensation Quantity = # of x-rays Milliroentges (mR) or (mR/mAs) Quality = the pentrability Kilovolts peak (kVp) or HVL
8
Electric Circuits Modifying electric flow and controlling electricity results in an electric circuit.
9
How the current gets to the TUBE
10
Generator+ Tranformers (where the power comes from)
11
Circuitry: Source: Carlton & Adler (1996)
Circuitry: Source: Carlton & Adler (1996). Principles of radiographic imaging: An art and a science. (96-99). MAIN CIRCUIT Modifies incoming current to produce x-rays Boosts voltage to range necessary produce x-rays. Modifies incoming line power to produce thermionic emission from the filament wire. FILAMENT CIRCUIT Filament circuit adjusts to mA ratings (50, 100, 200, etc.). After mA selection, current sent to step down transformer to modify amps that reach filament on x-ray tube
12
3 Divisions of Circuit Board
PRIMARY (CONTROL PANEL) yellow SECONDARY (HIGH VOLTAGE) blue FILAMENT (LOW CURRENT) purple
14
X-Ray Machine Circuit
15
In section two of this module, the components of the filament circuit were given.
One of the components in that list was the rheostat. Click on the image to see the location of the rheostat. A rheostat is a variable resistor. Just as the autotransformer (discussed in Section 3) can adjust output voltage, the rheostat can vary the amount of resistance in a particular part of a circuit. In the filament circuit section of the x-ray circuit, we are controlling the current that is applied to the filament of the x-ray tube by changing resistance. The x-ray tube will be discussed in detail in Module 7. In Section 6 of this module, we will pull all of the parts of the x-ray circuit together and discuss how they work as one circuit to produce x-rays. But first it is important to understand the role of the rheostat. The technologist has direct control over the rheostat through the mA stations on the control panel. When different mA stations are selected, the amount of resistance in the filament circuit is changed, resulting in a direct control of the current supplied to the filament. The higher the mA station number, the lower the resistance. A 1000 mA station will have very little resistance to the flow of electricity, while a 100 mA station will have considerable resistance.
16
Functional Position Control Console Transformers Tube
18
Complete Single-Phase Rectifier Circuit
high voltage, low current Bushberg, et al., The Essential Physics of Medical Imaging, 2nd ed., p. 126. 18 © UW and Brent K. Stewart PhD, DABMP
19
MAINBREAKER AUTOTRANSFORMER – KVP - Meter EXPOSURE SWITCH TIMER CIRCUIT HIGH VOLTAGE STEP UP TRANSFORMER RECTIFIER - Ma Meter FILAMENT CIRCUIT VARIABLE SELECTPR FILAMENT STEP DOWN TRANSFORMER X-RAY TUBE ROTOR / STATOR
22
© UW and Brent K. Stewart PhD, DABMP
high voltage, low current Bushberg, et al., The Essential Physics of Medical Imaging, 2nd ed., p. 126. 22 © UW and Brent K. Stewart PhD, DABMP
23
A
24
Name the parts of the Circuit board # A
TIMER KVP METER MA OR AMMETER RECTIFIER MA SELECTOR LINE COMPENSATOR INCOMING POWER MAIN BREAKER – VOLTS LINES 9.AUTOTRANSFORMER STEP DOWN TRANS FOCAL SPOT SELECTOR STEP UP TRANSFORMER MINOR KVP SELECTOR MAJOR KVP SELECTOR X-RAY TUBE
25
Important Parts Of The Circuit Board TO ID
1 Incoming Line Voltage 2 Autotransformer 3 KVP Selector 4 Timer No # Ma Selector 5 Primary Side (Low Voltage) 6 Secondary Side (High Voltage) 7 X-ray Tube 8 Rectifier 9 STEP – Up Transformer 10 STEP – Down Transformer
26
5 6 4 7 8 #1 3 9 mA selector 10 2 See Circuitry Review handout and chart for numbers
27
Wavelength is the distance from the peak of one wave to the peak of the next wave.
Frequency refers to the number of waves that go by a specific point in one second. Remember that electromagnetic energy waves all travel at the same speed—the speed of light Measured in Hertz or angstrom
28
Current from the outlet
29
Line Compensation Most imaging systems are designed to operate on 220 V. (some 110 V or 440 V) However power from the wall is not always accurate continuously
30
Line Compensation Wired to the autotransformer is the line compensator
Designed to maintain the accurate voltage required for consistent production of high-quality images Today’s line compensators are automatic and are not displayed on the control panel
31
Line Compensator
32
TRANSFORMERS (Step Up or Step Down)
Increases the VOLTAGE going to the ANODE side of the tube OR 110 volts to 110,000 volts Decreases the CURRENT going to the CATHODE side of the tube 5 Amps to 50 milliamps
33
Transformers Autotransformer Step Up Step Down
The primary circuit is made up of the main power switch (connected to the incoming power supply), circuit breakers , the autotransformer , the timer circuit , and the primary side of the step-up transformer (high tension transformer ). The main power switch is simply the "on/off" switch of the machine and is connected to the power supply of the facility. Circuit breakers are included in the primary circuit to protect against short circuits and electric shock. They are simply switches that open and break the flow of electricity in the event of a surge in power. Because the incoming power supply to the facility is not a consistent 220 volts, a line compensator is also utilized. This device is usually wired to the autotransformer (discussed later) and automatically adjusts the power supplied to the x-ray machine to 220 volts AUTO TRANSFORMER Through self-induction , the autotransformer selects the number of turns on its secondary side necessary to produce a voltage that, when increased about 500 times, will become the kVp selected. This voltage is then sent to the step-up transformer (high tension transformer). The step-up transformer will increase the voltage applied to its primary side by about 500 times to one kilovoltage on the secondary side through mutual induction. This transformer represents the transition from the primary section to the Through self-induction , the autotransformer selects the number of turns on its secondary side necessary to produce a voltage that, when increased about 500 times, will become the kVp selected. This voltage is then sent to the step-up transformer (high tension transformer). The step-up transformer will increase the voltage applied to its primary side by about 500 times to one kilovoltage on the secondary side through mutual induction. This transformer represents the transition from the primary section to the secondary section of the x-ray circuit.
34
AUTOTRANSFORMER
35
AUTOTRANSFORMER RAISES OR LOWERS THE VOLTAGE KVP TAPS LOCATED
220 VOLTS INCOMING CONVERTED FROM 100 T0 300 VOLTS VOLTMETER LOCATED - OFF OF AUTOTRANSFORMER
36
Autotransformer – Self Induction
There is only one wire – but works like when there are 2 wires = The windings are used as the primary and secondary coils The induced voltage varies on where the outside wires are connected (KVP Taps)
37
Autotransformer The power for the x-ray imaging system is delivered first to the autotransformer The autotransformer works on the principle of electromagnetic induction It has one winding and one core There are a number of connections along its length
38
Autotransformer A’s = primary connections & power into the transformer
Other connections allow for variations of voltages
39
Autotransformer Is designed to step up voltage to about twice the input voltage value The increase in voltage is directly related to the number of turns Operates on SELF INDUCTION
40
kVp selection
41
Filaments Operate at currents of 3 to 6 amperes (A)
Current comes from Autotransformer Controls the Ma selection Focal Spot Selector Switch located here
42
Nearly all x-ray equipment operates from an incoming line of ____?
The filament circuit in an x-ray tube operates at about what 10 volts of current
43
X-ray tube current or Filament circuit
A separate circuit crossing from cathode to anode Measured in milliampers (mA) What determines how many x-rays are created?
44
X-ray tube current or Filament circuit
# of e- is determined by the temperature of the filament. The hotter the filament the more e- Are their any limiting factors to thermionic emission?
45
mA selection
46
FILAMENT CURRENT MA METER vs AMMETER ??
MEASURE THE MA GOING TO THE XRAY TUBE
47
RHEOSTAT VARIABLE REISITOR
regulate the amount of resistance in a circuit mA control is found between the AMMETER
48
Generator+ Tranformers (where the power comes from)
49
TRANSFORMER is responsible for supplying a precise voltage to the x-ray machine.
50
TRANSFORMER FORMULAS (STEP UP OR DOWN)
V = voltage N = # turns p = primary s = secondary I = current Vp = N p Vs Ns Vp = I s Vs Ip Np = I s Ns Ip
51
Transformers Autotransformer HV Filament Step-up or step-down
Variable Turns Ratio Controls kV by varying V sent to HVT HV Step-up Fixed TR > 1 (500 to 1000) VS > VP V to kV Filament Step-down Fixed TR < 1 IS > IP (VS<VP) I in filament to cause e- emission
52
Transformer Efficiency
By design Shell – most Closed core Open core Air core Ideal – no loss Reality best = ~95% induction Loss due to Cu resistance wire diameter Eddy currents Laminate core Hysteresis core mag. perm.
53
TRANSFORMERS (Step Up or Step Down)
Increases the VOLTAGE going to the ANODE side of the tube OR 110 volts to 110,000 volts Decreases the CURRENT going to the CATHODE side of the tube 5 Amps to 50 milliamps
54
TRANSFORMERS STEP UP OR DOWN OPEN CORE, CLOSED CORE OR SHELL TYPE
ABOUT 95% EFFICIENT AUTOTRASFORMER = _____ induction Functions to provide ___________ Both types require AC for operation
55
INDUCTION MUTUAL INDUCTION is the? SELF INDUCTION is the
A transformer must have ________to produce an electric or magnetic current ?
56
MUTUAL INDUCTION (STEP UP & DOWN TRANSFORMERS)
57
AC Amplitude and polarity of the current vary with time
AC – sinusoidal wave AC varies amplitude and periodic reversal of polarity
58
STEP UP TRANSFORMER
59
Electricity Is more efficiently tansported over long distances at low currents ahd high voltage in order to avoid large power losses
60
Faraday’s law Regulate the strength of the induced current
Strength of the Magnet Speed of the motion Angle of the magnet Number of turns on the conduction coil
61
ELECTROMAGNETIC INDUCTION
3 WAYS TO CREATE MOTION BETWEEN LINES OF FORCE AND A CONDUCTOR Move the conductor through mag field Move magnetic lines of force Vary the magnetic flux
63
FUSES PREVENT SHOCK FROM A SHORT CIRCUIT –
THE HIGHER CURRENT WILL MELT THE FUSE – STOPPING THE FLOW OF ELECTRICITY CIRCUIT BREAKERS HAVE REPLACED FUSES - POWER TOO HIGH IT WILL CUT OFF – not damage appliance
64
What is directly proportional to the number of x-rays reaching the IR?
Question? What is directly proportional to the number of x-rays reaching the IR?
65
How do you convert msec to sec?
1000ms = 1 sec 100ms = 0.1 sec 10 ms = 0.01 sec
66
mAs Timers Monitors the product of mA and exposure time
Terminates the exposure when the desired mAs value is reached Located on the secondary side of the high-voltage transformer since actual tube current must be monitored
67
mAs Timers Designed to proved the highest mA for the shortest exposure
What is the name of this type of imaging system generator? Hint: most modern and most common
68
TIMER SWITCH Timer switch ends exposure
Timer – length of exposure set at control panel
69
Exposure Timers The timer circuit is separate from the other main circuits of the imaging system It is a mechanical or electronic device whose action is to “make” and “break” the high voltage across the x-ray tube This is done on the primary side of the high voltage transformer.
70
5 types of timing circuits
2 most common: Electronic Timer Computer controlled allow a wide range of time intervals are accurate to intervals as small as 1 ms can be used for rapid serial exposures
71
AEC Control AEC measure the quantity of radiation reaching the IR
Automatically terminates when the IR has received enough radiation for desired OD Two types are common
72
EXPOSURE & TIMER SWITCH
EXPOSURE SWITCH – BEGINS THE EXPOSURE Timer switch ends exposure (unless exposure button is let go prior to end of exposure because of dead man switch) Timer circuit controls number of photons produced. TYPES: MECHANICAL – NO LONGER USED spring wounded, and not very accurate. Synchronous timers: synchronous motor w/60 revolutions/second; min. exposure time is 1/60 sec, and timer in multipulse (i.e., 1/30,1/20). AEC - TIMERS: monitors time and mA and terminates exposure when desired mAs is attained. Designed to provide the highest safe tube current for the shortest time. mA METER OR AMMETER IS located on secondary side of high voltage section.
73
EXPOSURE SWITCH TIMER
74
Back up time for AEC P 116 Bushong (8th ed)
Back-up time should be set (electronic timer to 1.5 the expected exposure) Usually set automatically Exposure timer as short a 1 ms *Reg Rev Q: Manual reset timer = 6 sec (?)
75
APR Anatomically Programmed Radiography (Ch 20)
Radiologic Technologist selects on the console a picture of a written description of the anatomic part to be imaged and the patient body habitus A computer selects the appropriate kVp and mAs.
76
APR
77
Generator+ Tranformers (where the power comes from)
78
generator A generator is a device that converts
mechanical energy into electrical energy
79
X-ray Generator Transformers Rectification Connection to tube
HV (step-up) Filament (step down) Rectification Connection to tube cathode anode F diodes HV
80
generators An AC generator produces a current that is expressed mathematically as a sinusoidal wave ~ Function to change mechanical energy in to electrical energy Electrical current flowing through a conductor in one direction is _______ A battery is a source of direct current
81
GENERATOR THAT CREATE AN ALTERNATING CURRENT ARE CALLED:
AN ALTERNATOR – CONVERT MECHANICAL ENERGY INTO ELECTRICITY
82
WHAT MEASURES ELECTRIC POTIENTAL = VOLT
CURRENT = AMP ELECTRIC CIRCUIT IS THE PATHWAY FOR ELECTRIC CURRENT
83
High-Voltage Generator
Responsible for increasing the output voltage from the autotransformer to the kVp necessary for x-ray production 3 parts: High-voltage transformer, filament transformer and rectifiers
84
High voltage transformer
Or step up transformer Connected to the Major and Minor kVp selector Increases the volts from the autotransformer to kilovolts
85
Step Up Transformer
86
Voltage Rectification
Converts AC to DC current During the negative cycle current can only flow from anode to cathode E- must travel cathode to anode – DC current keeps e- traveling in the correct direction, cathode to anode Attracted to the positive anode
87
Voltage Rectification
88
X-Ray Tube Circuit
89
Rectifier tube failure
2 types A diode electron tube A vacuum tube If one rectifier were malfunctioning the inverted voltage would not be rectified - therefore the resulting mAs would be ½ that expected
90
Filament transformer Or step down transformer
Reduces the current to the filament
91
High-Voltage Generation – converts 110 volts of AC to kilovolts of DC
The generator is a FIXED component of the imaging system, not under the control of the technologist Three basic types: single phase, three phase, and high frequency The generator affects the quality and quantity of photons produced
92
Single Phase Power = Pulsating X-ray beam
What are the 2 types?
93
Half-Wave Rectification – photons are produced & emitted only during positive cycle
100% voltage ripple - quality is the same as full-wave rectification but quantity is half
94
Full-Wave Rectification – same as half-wave except there is no dead time
Half the exposure time is needed for full-wave than half-wave
95
Three-Phase Power : 6 pulse or 12 pulse
Results in higher quality and quantity photons
97
High-Frequency: nearly constant positive voltage
Less than 1% ripple Modern X-ray machines have High-frequency falling-load generator Automatically adjusts to the highest mA at the shortest exposure time possible
98
High Frequency
99
Voltage Ripple Single-phase power has 100% ripple
Voltage varies from zero to the maximum value Three-phase power has 14% ripple Voltage never falls below 86% of the maximum Three-phase, 12 pulse has 4% ripple Voltage never falls below 96% of the maximum
100
Voltage Ripple High-frequency power has 1% ripple
Voltage to the tube never falls below 99% What does this mean for x-ray photon?
101
Voltage generators (power supply)
103
Ripple effect differences
VOLTAGE RIPPLE Ripple effect differences
104
? What is missing?
105
A three-phase generator operates on three single phase currents, each one out of phase by _____ degrees
106
Review Handouts Circuit Board Symbols Function
109
Questions on imaging systems?