Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry
2
The traditional educational spectrophotometer Single beam Single channel
3
Multichannel Spectrometer
4
Spectrophotometry – the mundane Output from Spec-20: Absorbance = 0.390 Entire spectrum shows λ max and choice of analysis wavelength 2 spectrometers for lab of 24 students is sufficient
5
The multichannel spectrometer is easily configured for emission spectrometry using a fiber optic probe for light input. This has been used as a portable fluorescence emission spectrometer.
6
Fluorescence A typical luminescence spectrometerOur luminescence spectrometer Handheld UV Source Multichannel Spectrometer Emission MonochromatorMulitchannel Detector
7
Fluorescence Emission Spectrum
8
Absorbance and Fluorescence Emission Spectra of Quinine in Water
9
Fluorescence Emission Spectra
10
The multichannel spectrometer has also been used as a detector for qualitative analysis in a traditional flame test …
11
Fiber Optic Collection of Na Flame Emission Spectrum
13
Qualitative Analysis “Group IV” Cations: Na +, K +, Ca 2+, Mg 2+, NH 4 + Visual Flame Test: Na: persistent yellow K : transient red or lavender Ca: unreliable, transient red Na masks K when both are present (Co blue filter) False + and false – for Ca common with visual flame test
14
Qualitative Analysis Spectrometer Flame Test: Na and K easily detected and distinguished Ca not detected, chemical test only
15
Qualitative Analysis Spectrometer Flame Test Student Results: Na > 95% correct on 1 st try K > 95% correct on 1 st try Ca > 98% correct on 1 st try Fewer false +, fewer false – No blue filter for visual observation of K
16
Quantitative Analysis Emission intensity is transient and variable Use another alkali as an internal standard Sample introduction with a nichrome wire does not provide adequate quantitative results Aspirate a sample using a continuous sprayer into the air intake of a Bunsen burner for improved sample introduction
17
Compressed Air Spectrometer Fiber optic
20
K unknown = 200.0 ppmMeasured = 202.5 ppm Student results not that good, but calibrations are OK Detection limits using spray: 30-40 ppm for all alkalis Detection limits much lower using nichrome wire
21
Conclusions A multichannel spectrometer is useful for teaching spectrophotometry Using a fiber optic probe the spectrometer was configured for the following tasks: As a fluorescence emission spectrometer For qualitative analysis of alkalis For quantitative analysis of alkalis using an internal standard calibration