First DQE measurment of 500 µm thick Si hybrid pixel sensor with Photon counting readout for X-ray medical imaging M.Caria, J.Chaput, L.Sarry, B.Surre IWORID 2004 - Glasgow
2
IWORID 2004 - B.SURRE2 Introduction Aim: to measure system DQE following a standart protocole Requirements: –IEC 62220-1 –RQA 5 (~ 70 kVp) –Flat field corrected data
3
IWORID 2004 - B.SURRE3 Establishment of Standard X-ray spectrum Adjust kVp of generator to get RQA 5 spectrum such as: Spectrum Added filtration (mm Al) HVL (mm Al) RQA 3107 RQA 5217.1 RQA 7309.1 RQA 94011.5
4
IWORID 2004 - B.SURRE4 Measurement geometry
5
IWORID 2004 - B.SURRE5 Tube calibration @ RQA 5
6
IWORID 2004 - B.SURRE6 Set up ? Well known Photon counting chip = MEDIPIX 2 (chip set just above the noise) Bump bonded with 500 µm thick Si detector MUROS 2 interface Medisoft Thanks to the MEDIPIX collaboration
7
IWORID 2004 - B.SURRE7 Depletion voltage 20 % overdepletion 120 V
8
IWORID 2004 - B.SURRE8 Conversion function Pixel size 55x55 µm² Fluence: 91 #/(pixel.µGy) Slope: 5.9 count/(pixel.µGy) G=6.5%
9
IWORID 2004 - B.SURRE9 MTF (ESF) N = 1/tan( α ) = 12 Oversampling = 0.055 mm/12 Angle α
10
IWORID 2004 - B.SURRE10 SNR ? 20 flood images to compute the flat field coefficients
11
IWORID 2004 - B.SURRE11 ESF, RQA 5, 10 mAs
12
IWORID 2004 - B.SURRE12 7. Step (300 pts), RQA 5, 10 mAs Step(x)=ESF(x-x_shift)-ESF(x)
13
IWORID 2004 - B.SURRE13 DFT(Step) (300 pts), RQA 5, 10 mAs MTF(f)=|DFT(real step,f)|/|DFT(perfect step,f)|
14
IWORID 2004 - B.SURRE14 DFT(Step) (300 pts), RQA 5, 10 mAs MTF(f)=|DFT(real step,f)|/|DFT(perfect step,f)|
15
IWORID 2004 - B.SURRE15 DFT(Step) (300 pts), RQA 5, 10 mAs MTF(f)=|DFT(real step,f)|/|DFT(perfect step,f)|
16
IWORID 2004 - B.SURRE16 MTF, RQA 5, 10 mAs
17
IWORID 2004 - B.SURRE17 MTF, RQA 5, 10 mAs
18
IWORID 2004 - B.SURRE18 MTF, RQA 5, 10 mAs Presampling resolution: 7.7 lp/mm Visible detail limit = 0.3
19
IWORID 2004 - B.SURRE19 NPS @ RQA 5 dose
20
IWORID 2004 - B.SURRE20 NTF² @ RQA 5
21
IWORID 2004 - B.SURRE21 DQE @ RQA 5 DQE(f) = GxMTF²/NTF²
22
IWORID 2004 - B.SURRE22 NPS considerations & calculation Can we consider our results correct? >>> Calculations and simulations…
23
IWORID 2004 - B.SURRE23 Experimental NPS dose
24
IWORID 2004 - B.SURRE24 Simulated NPS dose
25
IWORID 2004 - B.SURRE25 NPS comparison: sim & exp
26
IWORID 2004 - B.SURRE26 1-D NPS comparison, 2 directions
27
IWORID 2004 - B.SURRE27 1-D NPS comparison
28
IWORID 2004 - B.SURRE28 1-D NPS comparison Interpolation increases the low frequencies
29
IWORID 2004 - B.SURRE29 Conlusions DQE has been calculated on 500 µm thick Si, bonded to a MEDIPIX2 chip A standard measure protocol have been used DQE is not so evident to find… How to improve this DQE ? (remember medicine) Next step: comparison with detectors of different thickness, study in parallel with detectors from the market, further calculations…
30
IWORID 2004 - B.SURRE30 Acknowledgements The MEDIPIX collaboration J.CHAPUT, M.CARIA L.SARRY You, for your attention