Monitoring with the Vision
Alarms BiPAP Vision Product Manual Monitoring
2
Alarm Overview Alarm: Warning Signal of an Event
Event: A Noteworthy Happening An abnormal clinical occurrence Requiring clinical awareness or action BiPAP Vision Product Manual Monitoring
3
Alarm Overview Events Are Related To:
Technical performance of the ventilator system or Patient’s clinical condition BiPAP Vision Product Manual Monitoring
4
Alarm Overview Two Types of Alarms System Adjustable
BiPAP Vision Product Manual Monitoring
5
Alarm Overview Alarm Indications Visual Audible Two Tones Alarm Reset
Manual Self-canceling Nurse Call/Remote BiPAP Vision Product Manual Monitoring
6
System Alarms Check Ventilator Performance Preset and Not Adjustable
Continuous Tone BiPAP Vision Product Manual Monitoring
7
System Alarms Ventilator Inoperative Check Ventilator
Pressure Regulation Proximal Pressure Line O2 Flow Exhalation Port Disconnect BiPAP Vision Product Manual Monitoring
8
Ventilator Inoperative
System Alarm Audible & Visual Why? System Malfunction AC Power Failure What Happens? Ventilator Powers Down Valves Open to Atmosphere allows spontaneous breathing BiPAP Vision Product Manual Monitoring
9
Check Ventilator Abnormal Operating Condition What Happens?
System Alarm Audible & Visual Manual Reset Abnormal Operating Condition What Happens? Ventilator Continues to Operate Unit Requires Service Visual Indicator Remains Until Condition Corrected BiPAP Vision Product Manual Monitoring
10
Check Ventilator Why? What Happens? Internal Battery Voltage is Low
System Alarm Visual & Audible Manual Reset Why? Internal Battery Voltage is Low What Happens? Internal Battery Cannot Sustain Audible Alarm AC Power Will Trickle Charge Error Code 205 BiPAP Vision Product Manual Monitoring
11
Pressure Regulation Why? What Happens? Proximal and Set Pressure Vary
System Alarm Audible & Visual Self-canceling Why? Proximal and Set Pressure Vary +/- 5 cm H2O What Happens? Ventilator Sends Available Flow to Achieve Set Pressure P Regulation BiPAP Vision Product Manual Monitoring
12
Proximal Line Disconnect
System Audible & Visual Self-canceling Proximal Line Disconnect Why? Proximal Pressure < 1 cmH2O What Happens? Ventilator Continues to Deliver Set Pressure ProxLine Disc BiPAP Vision Product Manual Monitoring
13
Oxygen Flow Why? What Happens? Insufficient Oxygen Flow
System Alarm Audible & Visual Manual Reset Why? Insufficient Oxygen Flow What Happens? FIO2 Cannot be Maintained Ventilator Continues to Deliver Set Pressure O2 Flow BiPAP Vision Product Manual Monitoring
14
Exhalation Port Why? What Happens?
System Alarm Audible & Visual Self-Canceling Why? Baseline Flow Decreases Below Intentional Leak Rate 5L/min or 50% What Happens? Ventilator Continues to Deliver Set Pressure Exh Port BiPAP Vision Product Manual Monitoring
15
Learn Function Learns Base Flow Upon system start-up
After mode changes Activated manually With circuit condition changes Modify Alarms screen BiPAP Vision Product Manual Monitoring
16
Disconnect Alarm Why? What Happens? Patient Disconnect
System Alarm Audible & Visual Self-Canceling Why? Patient Disconnect High flow rates for 4 seconds What Happens? Disconnect Ventilator decreases flow to idle BiPAP Vision Product Manual Monitoring
17
Flow Limit Control What Is It ? Recognizes Patient Disconnection
Decreases total flow to idle Flow Returns to Normal Spontaneous patient trigger BiPAP Vision Product Manual Monitoring
18
Benefits of System Alarms
Monitor Operating Specifications Warn of Abnormal Operating Conditions Visual Messages Facilitate Troubleshooting BiPAP Vision Product Manual Monitoring
19
Adjustable Alarms Monitor Specific Patient Parameters
Clinician Adjustable Multiple Tone BiPAP Vision Product Manual Monitoring
20
Adjustable Alarms High Pressure Low Pressure Apnea Low Rate High Rate
Low Pressure Delay Apnea Low Rate High Rate Low Minute Ventilation BiPAP Vision Product Manual Monitoring
21
High Pressure Why? What Happens? Inspiration is Terminated
Adjustable Alarm Audible & Visual Self-canceling Why? Proximal Pressure Exceeds Hi P Alarm Limit What Happens? Inspiration is Terminated Safety Feature Hi P BiPAP Vision Product Manual Monitoring
22
Low Pressure & Low Pressure Delay
Adjustable Alarm Audible & Visual Self-canceling Why? Proximal Pressure Below Lo P Limit for Set Time Delay What Happens? Ventilator Continues to Function Lo P BiPAP Vision Product Manual Monitoring
23
Setting Pressure Alarms
Hi P and Lo P Bracket IPAP Lo P Delay According to patient condition BiPAP Vision Product Manual Monitoring
24
Apnea Adjustable Alarm Audible & Visual Self-canceling Why? Absence of Spontaneous Trigger for Set Time Interval What Happens? Timed Breaths are Delivered at Set Rate Apnea BiPAP Vision Product Manual Monitoring
25
Apnea (con’t) Audible resets only when condition is corrected
Two spontaneous triggers Safety feature Apnea Can be Disabled BiPAP Vision Product Manual Monitoring
26
Low Rate Why? Setting Lo Rate Alarm Limits
Adjustable Alarm Audible & Visual Self-canceling Why? Total Respiratory Rate is Below Lo Rate Limit Setting Lo Rate Alarm Limits Set Below Total Respiratory Rate And Above Set Rate BiPAP Vision Product Manual Monitoring
27
High Rate Why? Total Respiratory Rate is Above Hi Rate Limit
Adjustable Alarm Audible & Visual Self-canceling High Rate Why? Total Respiratory Rate is Above Hi Rate Limit Setting Hi Rate Alarm Limits Set Above Total Respiratory Rate BiPAP Vision Product Manual Monitoring
28
Low Minute Ventilation
Adjustable Alarm Audible & Visual Self-canceling Why? Minute Ventilation Falls Below Set Lo MinVent Limit Setting Lo MinVent alarm limits Set ~ 20% < Total Minute Ventilation Use MinVent estimation BiPAP Vision Product Manual Monitoring
29
Benefits of Adjustable Alarms
Monitor Patient Conditions Warn of Critical Changes in the Patient as determined by the clinician Visual Messages Facilitate Troubleshooting BiPAP Vision Product Manual Monitoring