Radiological Equipment - Basic
2
Objectives For each radiation detection instrument: −Identify capabilities, limitations, and primary uses −Identify the components, controls, and indicators −Perform basic operating procedures −Understand operator maintenance
3
IsoTrak DoseGuard Personal Alarm Dosimeter
4
Introduction Overview Primary uses Capabilities Limitations Equipment safety precautions Components, controls, and indicators
5
Personal Dosimeter Capabilities −Records total gamma/x-ray dose and dose rate −User-adjustable features Limitations −Not intended for survey −Does not detect alpha, beta, neutrons −No display light Best use −Alert wearer of high dose or dose rate −Indoors hospital staff
6
Equipment Safety Precautions Fragile instrument; might survive 1 meter drop Do not submerse; can withstand rain
7
Components, Controls, and Indicators 8.8.8 mR - + Speaker Clip Button Display Chirp icon Battery icon Autoranging 3-digit Unit of measure Battery well
8
Operation Overview Prepare for operation Basic operating procedures Alarms Equipment application After operation
9
Prepare for Operation Before-operation checks −Look for obvious damage Install battery: 1 AAA −Positive terminal inside Power up −Hold button until long beep, then release −Automatic power up when installing battery Check dose reading is 0.0 mR
10
Basic Operating Procedure Default is total dose Check dose rate reading −Press button 1x −Display shows dose rate for 10 sec., returns to default
11
Types of Alarms High radiation −Dose alarm −Dose rate alarm Equipment fault −Overflow alarms −Battery alarms
12
Dose or Rate Alarms Alarm sounds when exceeding alarm set point −Dose alarm: display flash + 1 beep/sec. −Dose rate alarm: display flash + 4 beeps/sec. −Both: display flash + 4 beeps/sec.
13
Dose or Rate Alarms (cont.) To silence audible alarm: press button 1x To reset alarm −Dose rate alarm stops when exposure is reduced −Dose alarm stops when dose is reset to zero
14
Overflow Alarms Occurs when meter exposure exceeds factory- set measurement upper level −Dose overflow (1000 R): display “doo” + 2 beeps/sec. −Rate overflow (300 R/h): display “dro” + 8 beeps/sec. Cannot be reset until removed from exposure
15
Battery Alarms Alarm during startup −Low battery: display shows battery symbol −Dead battery: failure to start Alarm during operation −Low battery: battery symbol flashes, 1 long beep/hour −Dying battery: display “Lob” and 1 beep/sec. Measurement functions stop
16
Respond to Alarms Type of Alarm Dose/rate alarm Overflow alarm Low battery alarm Required Action Take immediate action to reduce exposure Silence and reset alarm Shut down instrument, remove from area Change battery
17
Equipment Application Need for dosimetry −Personal exposure record −Alert user of high dose or dose rate Wear on chest pocket −Instrument inside pocket – prevent dropping −Remove metal objects from pocket – avoid unintentional shielding −Place in plastic bag (optional)
18
After Operating Power down procedure −Press button 2x −Display reads “OFF” −Hold button until long beep, release Perform after-operation maintenance Storage −Remove battery −Store in box with instructions
19
Maintenance Overview Minimal operator care and cleaning No corrective maintenance No troubleshooting procedures
20
Operator Care and Cleaning Wipe exterior surfaces clean Do not submerse; rain resistant but not water tight −If submersed—open battery well, shake out excess water, let it air dry Avoid harsh solvents—may damage plastic case
21
Change Battery During Operation Replace battery if low battery alarm Power Off; change battery; power On −Dose reading and settings not lost
22
IsoTrak DoseGuard Summary Understand capabilities, limitations, uses Recognize components, controls, indicators Perform basic operation Understand operator maintenance
23
Ultra Radiac Personal Radiation Monitor
24
Introduction Overview Primary uses Capabilities Limitations Equipment safety precautions Components, controls, and indicators
25
Radiation Monitor Capabilities −Measures and records gamma/x-ray dose and rate −Multiple user features, data logger, PC download Limitations −Does not detect alpha, beta, neutrons Best uses −Personal dosimeter and rate meter −Used by decon team outdoors
26
Equipment Safety Precautions Do not submerse, but can withstand decon operations
27
Controls and Indicators
28
Digital Display Status indicators Autoranging digital display Units of measure
29
Function Keys Rate and Dose keys: Display dose rate or dose Alarm key: Display “stay time” in minutes Light key: Power On backlight Clear/Test key: Perform operational test On/Off key: Power On and Off Keys have other functions for advanced users
30
Battery Life Indicator Low battery: Flashing “BAT” in upper left corner −Less than 10 hours left −Press Clear/Test key to display minutes of battery left Dying battery: Flashing “b” on blank display −Normal operation stopped
31
Operation Overview Prepare for operation Basic operating procedures Alarms Equipment application After operation
32
Prepare for Operation Before-operation checks −Look for obvious damage Install batteries: 4 AAA −Observe polarity indicator −Starts “sleep” mode Power up −Press On/Off key
33
Prepare for Operation (cont.) Perform operational test −Hold Clear/Test key until “888” appears −Test alarm: Press Rate, Dose, or Alarm keys −After self-test: Flashing “9” = pass or “0” = fail −Press Clear/Test key Check the dose is 0.00
34
Attach Carry Case
35
Basic Operating Procedures Check dose and dose rate Turn on back light Check stay time Data logging Alarms
36
Check Dose and Dose Rate Current dose rate −Normal default reading −Press Rate key from dose display Total dose −Press Dose key −Returns to rate after 10 sec.
37
Backlight Press Light key −Blue light for 10 sec. Why can’t I see the light?
38
Check Stay Time Stay time = number of minutes to safely remain at current dose rate and not sound dose alarm −Based on current dose rate and dose alarm setting Press Alarm button −Normal background = “999” −If rate increases with low dose alarm set point, stay time decreases
39
Data Logging Log data point at: −Any dose alarm −End of normal shift Records up to 300 events −Records date, time, dose, rate, etc. Logging a data point −Power On, rate displayed −Hold Rate + Dose keys until number stops flashing −New number = log sequence
40
Alarms Type AlarmVisual (flashing) Audible (if enabled) Vibration (if enabled) High rate Entire display, Left LED red Beep quicklyVibrate Low rate Rate and Alarm indicators, Left LED green Beep slowlyVibrate High dose Dose and Alarm indicators, Right LED red Beep quicklyVibrate Low dose Dose indicator, Right LED green Beep slowlyVibrate
41
Resetting Alarms High rate alarms −Press Clear/Test key to still vibration alarm −Back away to stop audible and display alarms Low rate alarms −Press Clear/Test key to still audible/vibration alarms −Back away until display alarm stops Dose alarms −Press Clear/Test key to still audible/vibration alarms −Log the data point
42
Respond to Alarms Type of Alarm High dose/rate alarm Low dose/rate alarm Low battery alarm Required Action Take immediate action to reduce exposure Silence and reset alarm Warn others of increased exposure Silence and reset alarm Change batteries
43
Equipment Application Decon line use −Personal exposure record −Alert user of high dose or dose rate −Use “source finder” mode as survey meter Wear on belt −Instrument upside down, face out – flip up to read
44
After Operating Power down −Hold On/Off key until display shows “OFF” then “- - -” Perform after-operation maintenance Storage −Remove batteries −Store in carry case −Store IR reader with RSO
45
Maintenance Overview Minimal operator care and cleaning No corrective maintenance No operator troubleshooting
46
Operator Care and Cleaning Wipe exterior surfaces clean Do not submerse—not water tight Avoid harsh solvents—may damage plastic case
47
Change Batteries Replace batteries if low battery alarm Power Off; change batteries; power On −Ensure correct orientation −Dose reading and settings not lost Sleep mode starts automatically −Power On then Off to prevent battery drain
48
Ultra Radiac Summary Understand capabilities, limitations, uses Recognize components, controls, indicators Perform basic operation Understand operator maintenance
49
Radiation Survey Meter Responder kit and equipment
50
Introduction Overview Primary uses Capabilities Limitations Equipment safety precautions Components, controls, and indicators
51
Survey Meter Kit Capabilities −Measures alpha, beta, gamma, x-ray dose rate −Complete kit in hard case Limitations −Requires separate probes −No dosimeter function Best uses −Area survey −Check radiation from contaminated person/object
52
Equipment Safety Precautions Imbedded radioactive check source—keep covered −1 μCi Cs-137 on side of case Delicate pancake probe face, internal components; do not drop or puncture High internal voltage; power Off before changing probes or opening case
53
Major Components Hard carry case Meter Cable Pancake probe Batteries Gamma probe
54
Additional Components Additional survey meter Cable Extension handle pancake probe
55
Controls and Indicators DET DET 1 2 RESET LIGHT AUDAUD ON OFF F S OFFOFF RATEMETER SCALAR CAL Battery well Calibration switch well Probe connector Display F/S toggle (response time) Audio toggle (chirp) Detector toggle (parameters) Light button Reset button Power/mode switch
56
Digital Display Status messages 6-digit number (1 st & 2 nd digits) Units of measure4-digit number Low battery icon Autoranging decimal and units
57
Operation Overview Prepare for operation Basic operating procedures Alarms Equipment application After operation Practical exercise
58
Prepare for Operation Before-operation checks Install batteries – 2 D cells −Observe polarity indicator Attach probe −Depends on intended use
59
Visually Inspect for Damage Inspection Point – Instrument exterior Radioactive source Instrument display face Instrument switches Electrical connectors (all) Battery well Electrical cord Probe exterior Pancake probe face Do not use (until fixed) if – Cracked, broken open Broken open, loose (missing?) Cracked, broken open Will not move Cannot connect cord Contacts corroded, lid won’t close Frayed, torn open Cracked, broken open Punctured, torn
60
Prepare for Operation (cont.) Set toggle switches for operating mode −Audio on/off, F/S, Detector 1/2 Power On: Set Mode switch to Ratemeter −Display shows background reading Check operation with check source
61
Inspect For Proper Operation Inspection points – Display reads correctly Detector switch set for probe used Reset button clears reading Light button lights backlight AUD toggle ON (chirps) and OFF (mute) F/S toggle – no functional test Do not use (troubleshoot) if – Display blank or does not show correct data Switch cannot be moved or parameters do not change Display does not reset to zero before increasing to normal reading (Can be used without light) (Can be used without chirp) (Can be used in either position)
62
Basic Operating Procedures Use clicker Use gamma probe Change probes Use pancake probe Alarms
63
Use Clicker Clicker sounds as dose rate increases Enable using AUD toggle switch −Disable to mute audible alarm
64
Use Gamma Probe Setup −Attach gamma probe −Set toggle to DET 2 −Set mode switch to Ratemeter Use to search for gamma/x-ray radiation −Display reads radiation level in uR/h to R/h −Very sensitive to low level radiation −Fast clicker rate
65
Change Probes Procedure −Power Off −Change probe −Set detector toggle switch −Power On
66
Use Pancake Probe Setup −Attach pancake probe or extension handle probe −Set toggles to DET 1, Fast mode (F) −Set mode switch to Ratemeter −Remove protective cover Use to search for surface contamination −Display reads in cpm or kcpm (1000 counts/min) −Typical equivalent of ~3300 cpm per mR/h
67
Respond to Alarms Type of Alarm Dose rate alarm -or- Overflow alarm -or- Overload alarm Low battery icon Required Action Take immediate action to reduce exposure Press reset button Change batteries
68
Dose Rate Alarm Alarm sounds when exceeding preset level −Audible continuous sound; “ALARM” is displayed −Alarm is latched; must be acknowledged To acknowledge: Press Reset 1x To reset alarm: Press Reset 2x
69
Dose Rate Alert Alert shown when exceeding preset level −Lower setting than Alarm level −“ALERT” is displayed −Alert is latched; must be acknowledged To acknowledge: Press Reset 1x To reset if alarming: Press Reset 2x
70
Overflow Alarms Overflow alarm: exceeded 100,000 counts −“OFLOW” is displayed −Remove probe from area, press Reset button Overload alarm: radiation level above measurement capability −“OVERLOAD” is displayed −Back out of area −Reset: Press Reset button 2x
71
Equipment Application Area survey −Use gamma probe −Area survey techniques Frisk for contamination −Use pancake probe −Search techniques
72
Area Survey Procedures Use clicker for early warning Enter area slowly −Hold probe waist high in front −Move probe side to side toward suspicious areas Where to look −Walkways – material tracked in or out −Suspect materials spilled −Items touched by contaminated hands/gloves
73
Personal Search Procedures Clicker off to avoid panic Check ambulatory person −Hands −Head and shoulders −Shoes Check nonambulatory person −Hands and head −Both sides, shoulders to feet
74
Materiel Search Procedures Equipment or objects −Handle −Bottom or other touched surface −Parts that extend from main object Patients’ belongings −Bag items first −Hold bag and check both sides (gamma/x-ray) −Save for more thorough search later
75
Quick Vehicle Search Look at YOUR vehicles Driver/passenger area (contaminated person) −Floor and pedals −Items touched (steering wheel, door handles, radio, etc.) −Seats less likely Exterior −Wheel well top inside (dust from below) −Windowsills/wiper wells (dust from above)
76
After Operating Power down After-operation maintenance Short term storage −Disconnect and store in case −Batteries in vs. out Long term storage −Remove batteries, obtain new ones −Track calibration due date
77
Maintenance Overview Operator care and cleaning No operator corrective maintenance procedures No operator troubleshooting
78
Operator Care Before operation checks −Visually inspect for damage −Verify proper operation upon power-up After operation checks −Visually inspect for damage (same criteria) −Clean or decontaminate
79
Cleaning Procedures Wipe exterior surfaces clean −Use care not to puncture pancake probe face Do not submerse—rain resistant but not water tight −If submersed—open battery well and instrument case, shake out excess water, air dry Avoid solvents that could damage polyurethane paint
80
Survey Meter Summary Recognize capabilities, limitations, uses Recognize components, controls, indicators Perform basic operating procedures Understand operator maintenance
81
Area Radiation Monitor
82
Introduction Overview Primary uses Capabilities Limitations Equipment safety precautions Components, controls, and indicators
83
Area Monitor Capabilities −Measures gamma and x-ray dose rate −Wall mounted, 110 V −Internal and external alarm units Limitations −Does not detect beta, alpha, or neutrons −No dosimeter function −One hard wired remote per monitor Best use −Fixed site monitor
84
Equipment Safety Precautions High internal voltage; power Off and unplug before any service Fragile internal components; do not drop Indoors or fair weather use only
85
Major Components Model 272 remote alarmModel 375-10 Digital Area Monitor
86
Controls and Indicators Status LEDs Digital display Adjustment buttons Calibration controls
87
Controls and Indicators (cont.) Alarm RS-232 port 110 V AC Power switch Relay Detector center point Ethernet Left SideRight Side
88
Controls and Indicators (cont.) Alarm Toggle switches Analog monitor Status LEDs Reset button RS 232 port
89
Operation Overview Prepare for operation Basic operating procedures Alarms Equipment application After operation
90
Prepare for Operation Inspect for damage Check connections Adjust alarm horn volume: rotate horn cover Power on
91
Basic Operating Procedures Normal monitoring Alarms
92
Normal Monitoring Fixed site monitor −Power On when used or 24/7 −Check status LEDs and display periodically Network monitoring −May be used as additional remote monitor
93
Respond to Alarms Type of Alarm Dose rate alarm −High: fast beep + High LED −Low: slow beep + Low LED Detector Fail alarm −Det Fail LED + steady tone Low battery Alarm −Low Bat + Det Fail LEDs Overload alarm −Display “OL” + Det Fail LED Required Actions Take immediate action to investigate cause −Reset latched alarm Check power supply; return for service Check power supply Take immediate action to Investigate cause
94
Equipment Application Fixed site monitor −Mount near entry to monitored location −Alerts occupants to source coming in/near door Security monitors the remote or PC
95
Maintenance Overview Minimal operator care and cleaning No corrective maintenance procedures No troubleshooting
96
Operator Care and Cleaning Power Off and unplug before cleaning Wipe exterior surfaces clean Do not submerse Avoid solvents that could damage polyurethane paint
97
Area Monitor Summary Recognize capabilities, limitations, uses Recognize components, controls, indicators Know how to operate
98
Operator Competency Check For each instrument: −Prepare for operation −Check readings −Shut down For dosimeters: explain how it is worn and used For survey meters: explain survey procedures −People, equipment, vehicles
99
Do not print this slide. Chart only used for Note page.
103
New Equipment Summary Familiarization with the equipment Basic operating procedures Basic operator maintenance
104
Do not print this slide. Chart only used for Note page.