Pulsed and square wave voltammetry
2
Digital voltammetry waveforms – staircase used to approximate a ramp for LSV;
All modern potentiostats use this approach, also easy to use other input waveforms All sorts of pulsed voltammetry methods were developed in s by Sir Geoffrey Barker in UK, and later s modernized by Janet and Bob Osteryoung in the US
3
Basis of all pulsed methods: Response of reversible system to a potential pulse;
Measuemenst at end of pulse discriminates against charging current E 60 ms time measurement Faradaic I Charging (decays faster)
4
Normal Pulsed Voltammetry (simplest) DL about 15-fold lower than cyclic voltammetry (CV)
Input waveform output
5
Input waveforms output Normal pulse voltammetry Differential Pulse voltammetry Ep nM detection limits
6
Square Wave Voltammetry – complex waveform, derivative output most sensitive instrumental electrochemical method Input waveform Ep nM detection limits; Slightly better than Differential pulse output
7
SWV outputs Net or difference current Forward Current Reverse current
8
NPV SWV difference current I x 1000 O1 + e == R1 R1 + e == R2 Better resolution, Best sensitivity
9
SWV Output Net or difference current forward reverse
10
SWV parameters - increasing frequency (effect of DE is similar)
11
Approx DL NPV 10-6 M/n DPV 2x10-9 M/n SCV or LSV (CV) 5x10-5 M/n SWV 10-9 M/n