NCSLI 2013 Conference 1 How to Achieve a 0.01 µV/V Deviation on Your 10 Vdc Proficiency Test Without using a Josephson Array Author:Gary Bennett – National Instruments Metrology Laboratory Manager Speaker: Jorge Martins – National Instruments Principal Metrology Engineer
2
NCSLI 2013 Conference 2 Learning Objectives Activities required to maintain a 10 Vdc reference standard Tools available to make precision 10 Vdc transfer measurements Overview of techniques to track measured values and project future values of precision standards
3
NCSLI 2013 Conference 3 Opportunity for Internal Support Need to perform artifact calibrations on Fluke 5720A to maintain the 90 day specifications Purchased: – 4x 10 Vdc Reference Standards –Data Proof VoltRef SW and 160B scanner –Nanovoltmeter
4
NCSLI 2013 Conference 4 Requirements 10 Vdc uncertainty for calibrator adjustment is: ±1.5 µV/V 10 Vdc Reference Stability Specification
5
NCSLI 2013 Conference 5 Options Expand the 90 day specification of the 57x0 as per mfg. manual Shorten interval of 10 Vdc Zener references Buy a 10 Vdc with history … that you can trust! Devise a process to expedite the characterization of our Zener References
6
NCSLI 2013 Conference 6
7
7
8
8 Our Chosen Method Calibrate all 4 using MAP program after six months. –Measure the travelling standard Create procedure to measure the travelling standard Review the previous 6 months of VoltRef data Determine uncertainty –Measure all our 10 Vdc references against the travelling standard using the MAP provider’s procedure
9
NCSLI 2013 Conference 9 Measurements Prior to 2 nd Calibration
10
NCSLI 2013 Conference 10 Result of 1 st Proficiency Test
11
NCSLI 2013 Conference 11 What Happened?
12
NCSLI 2013 Conference 12 2 nd Calibration
13
NCSLI 2013 Conference 13 2 nd Calibration Results
14
NCSLI 2013 Conference 14 3 rd Calibration 2 nd Proficiency Test
15
NCSLI 2013 Conference 15 3 rd Calibration, 2 nd Proficiency Test
16
NCSLI 2013 Conference 16 Proficiency Test #2
17
NCSLI 2013 Conference 17 Projected Values and 3 rd Calibration
18
NCSLI 2013 Conference 18 3 rd Calibration
19
NCSLI 2013 Conference 19 Proficiency Test #3
20
NCSLI 2013 Conference 20 4 th Calibration
21
NCSLI 2013 Conference 21 Proficiency Test #4
22
NCSLI 2013 Conference 22 5 th Calibration
23
NCSLI 2013 Conference 23 5 th Calibration Cell #1
24
NCSLI 2013 Conference 24 Uncertainty Components Calibration Uncertainty (Travelling standard) Atmospheric Pressure Temperature Noise Type A Statistical Uncertainty of Projected Value Hysteresis Error
25
NCSLI 2013 Conference 25 Conclusion VoltRef works very well with the proper drift information for each cell Tools are available to make the data gathering less painful Understanding the normal differences in performance of individual references is important Gathering history on reference standards is expensive
26
NCSLI 2013 Conference 26 Questions