1
Sterilization – In General What’s the process? –Decontamination – removes > 80% of microbes –Sterilization – Kills/inactivates the remaining microbes What are the Sterilants? –Saturated Steam –Ethylene Oxide –Ozone –Hydrogen Peroxide –High Level Disinfectants (Glutaraldehyde, OPA, PAA)
2
Sterilization – In General What’s the process? –Decontamination – removes > 80% of microbes –Sterilization – Kills/inactivates the remaining microbes What are the Sterilants? –Saturated Steam –Ethylene Oxide –Ozone –Hydrogen Peroxide –High Level Disinfectants (Glutaraldehyde, OPA, PAA)
3
Steam Sterilization “Critical” Variables –Time. –Temperature. –Saturated Steam. (Depends on Temperature and Pressure) Other Variables (Incomplete List) –Loading –Sterilant contact with items to be sterilized –Outside weather changes –Boiler Chemicals –Wrapping Materials –Etc, etc, etc…….
4
Saturated Steam Wet Steam/ Liquid Water Superheated Steam
5
Saturated Steam – What is it? It is a colourless gas, containing < 3% liquid water Depends on temperature and pressure It will condense if it cools down a fraction of a degree It releases latent or potential heat energy as it condenses It is the latent or potential heat transferred to microbes that kills/inactivates them
6
Latent Heat = Potential Energy 16 Tons 16 Ton weight is sitting on the floor It has zero Potential energy Researcher knows the 16 ton weight has zero Potential energy. She is not afraid.
7
Latent Heat = Potential Energy 16 Tons 2 meters 16 ton weight is now 2 meters off the ground. It has quite a lot of potential energy with respect to the researcher The researcher is now very afraid of the 16 ton weight’s potential energy
8
Latent Heat = Potential Energy 16 Tons 2 meters Latent or Potential energy from the 16 ton weight is now realized. Especially by the late Researcher.
9
Latent Heat = Potential Energy 16 Tons 16 Tons Steam (Gas) Water (Liquid)
10
The “Experiment” Generating Saturated Steam at 1 Atmosphere Temp Time 100ºC All the energy is now going in To creating “latent” heat in the steam With no increase in temperature It takes 150 KJoules to heat 1 lb of water from 20ºC to 100ºC and 1023 Kjoules to convert 1 lb of water from liquid to gas at 100ºC
11
Typical Steam Sterilizer
12
Sterilization Monitoring We monitor the Process not sterility of items Because the process is validated (by the manufacturer of the sterilizer) we assume that, if the process functions correctly, the goods in the process will be sterile.
13
Process Monitors for Steam Sterilization 3 Major types used: Biological Indicators: –Monitor all variables in the process Chemical Indicators: –Monitor one or more of the critical variables Mechanical Indicators: –Monitor one of the critical variables
14
Process Monitors for Steam Sterilization Individually they provide information, but only together do they tell the whole story! A failure in any one of the monitors implies the process is incomplete in one or more necessary variables
15
Biological Indicators Large # (> 10 5 or 10 6 ) of resistant spores Usually “Self-contained” type Placed in a normally loaded sterilizer At least one per day per cycle (CSA) Final Readout made at 1, 3, 24 or 48 hours Test the process to ensure it is capable of killing microorganisms
16
Vial SporeStrip Ampoule + Growth media & Indicators Filter Cap Self-contained Biological Indicator
17
How do they Work? Colour change indication: –If spore survives, it is incubated –Spore consumes nutrients, excretes acid –Spore creates subsequent generations that also consume nutrients and excrete acids –Purple (or other) dye is acid sensitive –If enough acid is produced, dye will change from purple to yellow (or other colour change) –Process can take up to 48 hours to get reliable prediction of kill
18
How do they Work? Rapid Read BIs –Spore survives the sterilization cycle and is incubated –Nutrients are “tagged” with a U.V. fluorescent dye –As spore consumes nutrients the dye is released and begins to fluoresce –Fluorescence is detected electronically –Process takes up 3 hours to get reliable prediction of kill
19
What bout these “Extended Cycles”? G.stearothermophilus will still be the test spore. Requires a new Process Challenge Device (test pack) not a new BI Extended cycle required because bigger. Heavier more complex sets protect microbes better PCD must provide the same protection to the BI
20
BIs and CIs All current sterilization standards (AAMI, CSA and ISO) require the use of Biological indicators. Chemical Indicators cannot replace Biological indicators because: –Biologics are far more complex –Biologics have the potential to recover after serious damage –Chemistries only approximate the response of Biologics –Chemical kinetics are far simpler than biologic kinetics –Chemical indicators only measure the critical variables of the process. Biologics integrate all the variables of microbial death
21
Understanding Chemical Indicators To fully understand Chemical indicators one must understand: –BIER/CIER Vessels/Test methods used by manufacturers –Stated Values (SVs) –ISO/AAMI Classifications of Chemical Indicators –How it all relates to YOUR sterilization process –What a pass/fail result really means
22
Test Methods for Testing Chemical Indicators Must be reproducible, controlled and consistent Based on standard test “Sterilizers” called BIER or CIER vessels BIER = Biological Indicator Experimental Resistometer CIER = Chemical Indicator Experimental Resistometer BIER/CIER vessels performance specified by ISO 18472 BIER/CIER Vessels behave very differently from a typical health care sterilizer
23
BIER/CIER Vessels Very Small Chamber Very tight controls Highly reproducible results Basis for all Biological and Chemical indicator testing All CI “Stated Values” are based on this vessel
24
Some Requirements of CIER Vessels Come-up time < 10 seconds Come-down time < 10 seconds Vacuum air removal < 2 minutes No steam allowed during air removal Set Temperature within 0.5ºC Set Time within 1 second Pressure within 3.5 Torr One data point recorded per second
25
Contrast CIER Vessel to Health Care Sterilizer Come-up time 5 to 15 minutes Come-down time 5 to 15 minutes Vacuum air removal 5 to 15 minutes Steam used during air removal Set Temperature within 2ºC (?) Set Time within ?? Pressure within ?? One data point recorded per 30 seconds
26
Test Methods for Chemical Indicators All testing done in a CIER Vessel Must meet requirements of ISO 11140-1 ISO 11140 is written for Manufacturers not end users Specifications are broken in to 6 categories Only need to show the ability to reach the stated values.
27
Stated Values: What are they? For any given chemical indicator to be compliant with ISO 11140-1, it must have “Stated Values” for the parameters it measures. Example: For a Class 3 Chemical Indicator, it must have a stated value for temperature. This is the temperature at which we can expect to see the change in the CIER vessel.
28
Stated Values and ISO 11140-1 The standard uses the “Stated Values” to define how the indicator will perform. Example: A Class 4 Indicator Stated Values: –134ºC –4 Minutes ISO allows a range: Colour change (Pass result) must be seen at 134ºC, after 4 minutes exposure in the CIER vessel. A fail (no colour change) result must be seen at 132ºC after only 3 minutes exposure. No additional testing is required at points in between.
29
Stated Values and ISO 11140-1 A Class 6 example: Stated Values: –134ºC, 4 Minutes (Saturated steam) –A pass result must be seen ( in the CIER vessel) after 4 minutes exposure at 134ºC –A Fail result must be seen at 133ºC after 3 min 45 second exposure in a CIER vessel. –No additional testing of points in between is required.
30
Stated Values and ISO 11140-1 The BIG QUESTION So, if a chemical indicator shows a “pass” result in your sterilizer does it mean that the Stated Values given for that particular chemical indicator have been achieved?
31
Chemical Indicators Changes (reactions that cause colour change) start at temperatures well below the stated values for the CI Temperature at which the reaction starts is the initiation temperature Changes (reactions) will continue once initiated and the temperature is above the initiation temperature
32
Rates of Reaction and Initiation Energies In our house my wife allows me the honour of pressing my own shirts. As part of that process I have learned a lot about the chemistry of burning fabrics.
33
Rates of Reaction and Initiation Energies For example, I discovered that if I place an unplugged (cool) iron on a shirt, I can leave it there indefinitely and it will not scorch the shirt. That’s because the energy imparted to the shirt from the cool iron is well below the initiation energy required to start the burning reaction of the fabric. It also doesn’t remove wrinkles
34
Rates of Reaction and Initiation Energies I’ve also learned that I can leave an iron on its lowest setting on the shirt indefinitely and it will not scorch the shirt either. I guess the low setting still doesn’t deliver enough energy to start the burning reaction
35
Rates of Reaction and Initiation Energies Of course, an iron on medium setting will scorch the shirt, but it takes quite a few minutes to do so, On high setting, the wrinkles go, but if I leave the iron on the shirt for even a short time, it will scorch.
36
The Scorched Shirt Example
37
Chemical Indicators Example (Theoretical) CI has stated values of 134ºC, 4 minutes: Initiation temperature may be 100ºC and reaction (change) will start to occur at that temperature and continue while the temperature remains over 100ºC. The rate of change will vary with temperature once initiated. Usually faster as temperature increases
38
BIER/CIER Vessel Testing
39
Same CI in a Health Care Sterilizer
40
Next Big Question(s) If chemical indicators start to react below the stated values: –How fast do they react at lower temperatures? –Will they change if they spend sufficient time at lower temperatures? –What does that mean for monitoring my sterilizer?
41
Extended Cycle Time Temperature Plot: Set Points 132ºC, 18 min)
42
Conclusions so far….. Stated Values cannot be duplicated in Health Care Sterilizers Stated values can be used only for comparative purposes between Chemical Indicators of the same class Stated Values are used by Manufacturers of Chemical Indicators to make CSA/ISO or AAMI Classification claims.
43
CSA/ISO/AAMI Classification of Chemical Indicators All follow the ISO 11140, 6 tier classification structure All state the class has no hierarchical significance. (i.e. a Class 2 is not “better” than a Class 3 or a Class 6 is not “better” than a Class 5.) Classification structure is to provide specifications and targets for manufacturers of chemical indicators It may be used by users for comparing CIs within the same class
44
Definitions Critical Variable –For steam sterilization Time, Temperature and Saturated Steam –For Ethylene Oxide: EtO concentration, humidity, time, temperature Critical Parameter –A set value of a critical variable that must be attained Stated Values –The critical parameters measured by a Chemical Indicator in a CIER vessel
45
CSA/ISO/AAMI Classification Class 1 – Process Indicators –Autoclave tapes, built-in CIs on peel pouches –Simply identify processed from unprocessed packages Class 2 – Special Indicators –Bowie-Dick Type Tests Class 3, 4, 5, and 6 are all Internal Indicators
46
Classes 3 - 6 Class 3 Indicator: –Measures only one of the critical variables of the process Class 4 Indicator: –Measures 2 or more of the critical variables of the process –E.g. Time and temperature, or time saturated steam.
47
Classes 3 - 6 Class 5 Integrating Indicator –Measures all critical variables of the process, and correlates with the test spore for that process across a range of sterilization conditions –For steam sterilization, the Class 5 must correlate with G.stearothermophilus at 121ºC, 128ºC* and 135ºC –Has a known response over a range of sterilization temperatures * May be other temperatures within the range of 121 - 135ºC
48
Class 5 Integrating Indicator Response
49
Class 3 - 6 Class 6 Emulating Indicator –Measures all critical variables of the process –Tightest specifications of all indicators –Stated Values are for a specific time and temperature for steam sterilization
50
Class 6 Emulating Indicator Response
52
Conclusions: Saturated Steam is critical to the sterilization process Non-condensable gases (air, CO 2, N 2 ) are poor heat transfer media and poor sterilants It is the latent heat of saturated steam, not the temperature that kills the microbes
53
Conclusions All process monitors must agree that the cycle was successful A failure in any one of the 3 types of monitor (Biological, chemical or mechanical) means the process was insufficient and processed goods are suspect
54
Conclusions Biological Indicators are the only indicator to integrate all the variables of he sterilization process They are the only direct measure of the lethality of the process The BI response can be approximated by chemical indicators, but CIs cannot replace BIs
55
Conclusions The basis for classification of Chemical Indicators in Canada is CANISO 11140 This standard is written for manufacturers of chemical indicators. The standard has no hierarchical significance Stated values are only realizable in the test vessel (CIER vessel)
56
Conclusions Selection of biological indicators and chemical indicators must be made considering patient safety, the information needed from the indicators, the standard requirements, best practices and economic considerations. No single indicator will likely accommodate every circumstance