Ahura First Defender: Acquisition and Application in the Academic Environment 2013 College and University Hazardous Waste Conference Kyle Dagestad Sean Whalen Iowa State University
2
Introduction The Money - Sean The Selection - Sean The Equipment - Kyle The Goods and The Bads- Kyle Safety and Maintenance – Kyle
3
The Money Homeland Security Buffer Zone Program – Grant through ISU-Police FTIR and RAMAN for unknowns $82,500 – Equipment only Purchased the Ahura First Defender ($35,000) and Tru Defender ($47,500) EH&S owns equipment Tru Defender on loan to Ames Fire HAZMAT
4
The Selection Two primary suppliers – Ahura Scientific – Smith’s Detection Disclaimer Representative Presentations Field Tests Matrix
5
Field Tests Two Unknown “Spills” – Citric Acid – Soda Ash Lab Unknowns – Corrosives – Solvents – Metals – Mixtures
6
Field Test Ahura Either Yes or No Results – 50% spill – 73% unknowns – High fluorescence is biggest problem Smith’s Provides options Results – 50% spill – 47% unknowns – False Positives Ketamine as TNT Trimethylamine HCl as molybdenum oxide or propanol
7
Field Test Results
8
Matrix Pointed out best unit Ahura scored highest Ahura more expensive Had to write justification
10
The Equipment How does it work? Raman Spectroscopy: Molecular “fingerprint” of the molecule Class 3B laser excites the molecular bonds and scatters measurable light to identify unknowns ISU Ahura library- 10,152
11
Why the Ahura First Defender? Currently used by law enforcement, firefighters, and the military Very durable (4 Feet & 1 Meter) No calibration needed-confidence check User friendly Ahura Customer Support- 24/7- 365
12
Identification Capabilities Drugs (illegal & prescription) White powders Inorganic compounds- (sulfuric acid, nitric) Inorganic oxides- (rust, titanium dioxide) Ionic compounds-(sulfates, phosphates) Substances in water Identifies through container
13
Identification Limitations No covalent bonds Dark colored materials Highly fluorescent materials- (dyes, ketchup) The sun can disrupt scan Most pure metals and elemental substances Biologicals- confused by mix of proteins, and fats Radiation, gasses, and vapors Highly diluted substances
14
Safety Follow your organization’s Laser Safety Protocols - Eyewear Maintain 14” distance Laser activated- don’t look into lens NOT intrinsically safe
15
Maintenance Cleaning Water, alcohol, household bleach Keep lens & nose cone clean Updating Chemical library updates “LibBuild Mode” Good spectrum, but no match? Export- “Ahura Reachback”
16
Demonstration Performance Point and shoot through a bottle? No problem! Why you shouldn’t scan dark colored materials video.