Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences
2
Normal Range: Axial length Biggest source of error in IOL power calculations. 96% of axial lengths fall within the range 21.0 to 25.5 mm 60% of AL is between 22.5 and 24.5 mm
3
Ultrasounds are calibrated with average velocities for normal length eyes. These velocities are incorrect for short eyes, causing significant measurement errors. Accurate measurement of axial length in hyperopic eyes is especially important since any error is greatly magnified in proportion to the length of the eye. Immersion or LASER biometry can provide superior results in these cases.
4
In a hyperopic eye with axial length of 20 millimeters each mm error in measurement cause 3.75 D post operative refractive error.
6
Difference in axial length of more than 0.3mm requires confirmation Most individuals have similar axial lengths in each eye except: –unilateral refractive error –coloboma –Staphyloma
10
In the axial myopia, the presence of a staphyloma should be considered, especially if there is difficulty obtaining a distinct retinal spike during A-scan. an immersion A/Bscan approach to axial length measurement in the setting of a posterior staphyloma
11
nearly all eyes with pathologic myopia have some form of posterior staphyloma. For highly myopic eyes (axial > 28mm) a B-scan should be carried out to determine the presence or otherwise of staphylomata
12
Long eye, foveal deep
13
Very long eye, posterior pole staphyloma
14
Buphthalmic globe: very long eye (37·5 mm): deep anterior chamber.
15
Megalocornea: average length eye (23·1 mm): deep anterior chamber (5·2 mm).
16
The Ascan vector is adjusted to pass through the middle of the cornea as well as the middle of the anterior and posterior lens echoes to assure that the vector will intersect the retina in the region of the fovea. IOL Master permits accurate measurements when posterior staphylomata are present. Since the patient fixates along the direction of the measuring beam.
19
Calculation formula: The SRK/T and Holladay formulae worked best overall. SRK I and II were found to be least accurate No statistical difference was found between SRK/T, Hoffer Q and the Holladay formulae
20
Axial LengthFormula <22 mmHoffer Q or SRK/T 22 – 24.5 mmSRK/T or Holloday, Haigis > 24.6SRK/T
22
When using IOL Master do not forget For AL < 22 do immersion A scan For AL > 30 do A and B scan
23
Silicone oil filled vitreous cavity: eye measures 49 mm on B-scan, but actual axial length is 34·3 mm.
24