ACap™ Confirm
3
ACap™ Confirm Benefits
Automatic algorithm to adjust the atrial pacing amplitude to address changing atrial thresholds Clinic time saved by daily automatic atrial amplitude testing and adjustment (8 – 23 hrs) Threshold trend with follow-up EGMs to quickly evaluate patients status Evaluate early lead integrity issues
4
ACap™ Confirm What to do
At activation Run Setup Test Run Capture Test At follow-up Read ACap™ Confirm results and trend
5
ACap™ Confirm Polarity
Bipolar pacing Uni- or bipolar sensing ER detection always unipolar Uni- or bipolar back-up pulse configuration
6
ACap™ Confirm Setup
7
ACap™ Confirm Setup The ER will be determined during the ACap™ Confirm Setup using Paced Depolarization Integral (PDI)
8
ACap™ Confirm Determine ER sensitivity
Ensures ER is large enough to separate from polarization Captured pulses = ER + Polarization Non-captured pulses = Polarization Only ER sensitivity is automatically set to a safe level between ER & polarization measurements
9
ACap™ Confirm How the Setup Test is Run
test rate 90 bpm default (max 120 bpm) 4 paced beats at each amplitude 1-2 beats between for program changes 4.5 4.25 4.0 3.75 3.5 3.0 2.5 2.0 1.5 1.0 0.75 0.5 0.25 0.0 Back up pulse, 5.0 V, is delivered after primary pulse on non captured 0.75V)
10
ACap™ Confirm Setup Test, Back-up pulse
40 ms after primary pulse 5.0 V, >0.5 ms unipolar or bipolar
11
ACap™ Confirm Setup Test, AV Delay
180 120 AV delay as programmed as long as capture in Setup Test Test dictates AV delay 120 ms when backup pulse is delivered (timed from primary atrial pulse)
12
ACap™ Confirm Evoked Response detection
Blanking Sensing (Alert) 4.5 ms 31ms PDI-area of ER in A. unipolar tip EGM channel PDI Paced Depolarization Integral ER Evoked Response
13
ACap™ Confirm Programmable Options
On A voltage is added to the threshold to define a new automatic pulse amplitude to be used until the next search (requires a 1.7:1 safety margin) Monitor Capture threshold is automatically measured and plotted on the trend but no automatic amplitude adjustment is made Off No measurements performed no weekly threshold trend data available
14
ACap™ Confirm Capture - Loss of Capture
Capture Confirmed Loss of Capture ER Sensitivity = 500 PDI = 750 ER Sensitivity = 500 PDI = 150
15
ACap™ Confirm Capture Test
16
ACap™ Confirm How the Capture Test is Run (Capture @ 0.825V)
Ensure atrial pacing: in-clinic test rate 90 bpm default (max 120 bpm) 4 paced beats at each amplitude Decrease amplitude in 0.25 V steps. 3 Consecutive pulses at the same amplitude confirm LOC (backup pulse delivered) Increase amplitude in V steps until 2 consecutive pulses capture at the same amplitude (capture threshold) Back up pulse, 5.0 V, is delivered after every primary pulse on non captured events
17
ACap™ Confirm How the Capture Test is Run (Capture @ 0.825V)
2.0 1.75 1.825 1.5 1.25 1.0 0.825 0.75
18
ACap™ Confirm Capture Test
Starts at programmed atrial pulse amplitude If programmed atrial pulse amplitude > V start the threshold search at V Decrease amplitude in 0.25 V steps until 3 consecutive pulses at the same amplitude confirm LOC (backup pulse delivered). Below 0.25 V, the step size is V Increase amplitude in V steps until 2 consecutive pulses capture at the same amplitude (capture threshold)
19
ACap™ Confirm Safety Margin
To 5.0V To 5.0V To 5.0V To 5.0V To 5.0V +2.0V +2.0V +1.5V +1.0V +1.0V +1.0V +1.0V +1.0V N/A 4.5 4.25 4.0 3.75 3.5 3.0 2.5 2.0 1.5 1.0 0.75 0.5 0.25 0.0 Adding safety margin related to capture threshold
20
ACap™ Confirm Safety Margin
21
ACap™ Confirm Out of Clinic Test
22
ACap™ Confirm Periodic Out of Clinic Threshold Search
Programmable: 8 or 24 (23) hrs Device overdrives the atrial rate Minimize fusion Maximize patient comfort by interspersing atrial test pulses throughout normal pacing
23
Threshold Search Algorithm
Monitoring Period Search initiation: open window of 16 atrial intervals Determine the rate overdrive interval for atrial test pulses 16 pulses should cover two respiratory cycles Continuously update rate overdrive interval Fusion avoidance
24
Threshold Search Algorithm Out of Clinic Periodic Threshold Search
Test Mode Conditions under which test occurs Fast 100% Atrial paced during monitoring period and one of AF Suppression Off AF Suppression On, but not actively overdriving Slow At least one p-wave during monitoring period and one of AF Suppression AF Suppression On and AF Suppression actively overdriving rate
25
Threshold Search Algorithm
Fast Mode 100% atrial pacing during monitoring period (AF Suppression is OFF or not actively overdriving)
26
Threshold Search (Fast) AF Suppression OFF or Not Actively Overdriving
Search initiation: 16 beat monitoring period 100% atrial pacing in monitoring period: AF supression off / no AF related atrial pacing Test pulse overdrive interval: Current pacing interval – 40 ms If another algorithm calculates a shorter interval than the test pulse overdrive interval, i.e.- Rate Response, the shorter interval is used Overdrive rate limit is 120 bpm
27
ACap™ Confirm How the Fast Mode is Run (Capture @ 0.825V)
Only Paced Events Monitoring, 16 beats 2.0 Monitoring, 16 beats Monitoring, 16 beats Monitoring, 16 beats 1.75 1.875 1.5 1.25 1.0 0.75 0.75 0.75 0.875 5.0 5.0 5.0 0.875
28
Threshold Search (Fast) cont’d
Question What if a p-wave were sensed during the threshold test? Answer A new 16 beat monitoring period would begin
29
ACap™ Confirm Sensing interrupts the test
FAST MODE Intrinsic event Restart of monitoring 0.75 2.0 1.75 1.25 1.5 1.0 0.875 1.875 Monitoring, 16 beats 5.0 I N T R I N S I C
30
Threshold Search Algorithm
Slow Mode at least one atrial event is sensed during monitoring period (AF Suppression is OFF or not actively overdriving)
31
Threshold Search (Slow) AF suppression OFF - Not Actively Overdriving
Search initiation: 16 beat monitoring period At least one sensed P- wave occurs during monitoring period Test pulse interval = MATRI - the max of either 3 x ARIAD 40 ms
32
Threshold Search (Slow) AF suppression OFF - Not Actively Overdriving
Test pulse interval = MATRI - the max of either a) 3 x ARIAD b) 40 ms MATRI Mean ATrial Rate Interval (from monitoring period, 16 atrial intervals) ARIAD Atrial Rate Interval Average Deviation (during monitoring period) Disregard the two shortest as well as the 6 longest intervals, then calculate the average deviation on the remaining 8 intervals
33
Threshold Search (Slow) AF Suppression OFF - Not Actively Overdriving
Rate Observation Period (Slow Mode) Started after 16 beat monitoring period 10 cycles in duration Look for a long atrial interval, ≥ MATRI (mean atrial rate interval) Following long interval, deliver test pulse at test pulse interval If a long atrial interval is not seen within 10 cycles: Test pulse is not delivered New 16 beat monitoring period is started
34
ACap™ Confirm (Slow) Monitoring and Rate Observation Period
LONG INTERVAL 956ms 972ms 986ms Rate Observation <10 beats Test Pulse following a long interval > MATRI 902ms Test Pulse Rate interval MATRI – (3xARIAD (or 40ms)) 977 – (3x25)ms = 902 (67 min-1) Monitoring, 16 beats I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C Interval Variation from mean value 1000ms 23ms 892ms 85ms 923ms 54ms 952ms 25ms 968ms 9ms 898ms 79ms Mean ATrial Rate Interval (MATRI) = 977 ms Atrial Rate Interval Average Deviation (ARIAD) = 25 ms 1000ms 1000ms 1000ms 1000ms 1000ms 1000ms
35
Threshold Search (Slow) AF suppression OFF - Not Actively Overdriving
New monitoring period is started after test pulse delivery If no atrial events are sensed in alert during monitoring period, search switches to Fast Mode Test proceeds as described before Attempt test pulse delivery in the first cycle after long atrial interval Test pulse delivery is inhibited by a sensed atrial event New monitoring period would be started
36
ACap™ Confirm How the Slow Mode is Run
Calculate mean atrial rate interval (MATRI) and atrial rate interval deviation (ARIAD) No long atrial interval is seen within 10 cycles Rate Observation Period New Monitoring Period p-waves Calculate MATRI and ARIAD Rate Observation Period test pulse after long interval New Monitoring Period No p-waves Fast-mode p-wave during threshold test Initiates new monitoring period New Monitoring Period No p-waves Fast-mode 2.0 2.0 1.75 I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C LONG INTERVAL I N T R I N S I C 1.5 1.25 1.0 0.75 Monitoring, 16 beats 5.0
37
ACap™ Confirm How the Slow Mode is Run
Calculate mean atrial rate interval (MATRI) and atrial rate interval deviation (ARIAD) Test pulse interval MATRI minus the max of either: 3 x ARIAD or 40 ms 0.75 2.0 1.75 1.25 1.5 1.0 0.875 1.875 Monitoring, 16 beats 5.0 I N T R I N S I C LONG INTERVAL
38
ACap™ Confirm How the Slow Mode is Run
Calculate mean atrial rate interval (MATRI) and atrial rate interval deviation (ARIAD) Test pulse interval MATRI minus the max of either: 3 x ARIAD or 40 ms Test pulse delivery is inhibited by a sensed atrial event. New monitoring period would be started 0.75 1.75 1.25 1.5 1.0 0.875 1.875 Monitoring, 16 beats 5.0 I N T R I N S I C 2.0 LONG INTERVAL
39
ACap™ Confirm How the Slow Mode is Run. (Capture @ 0.825V)
Calculate mean atrial rate interval (MATRI) and atrial rate interval deviation (ARIAD) Test pulse interval MATRI minus the max of either: 3 x ARIAD or 40 ms 2.0 2.0 2.0 Monitoring, 16 beats 1.75 I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C I N T R I N S I C LONG INTERVAL I N T R I N S I C LONG INTERVAL I N T R I N S I C 1.5 1.25 1.0 0.75 Monitoring, 16 beats 5.0
40
Threshold Search Algorithm
AF suppression Mode AF Suppression ON and Actively Overdriving
41
Threshold Search AF Suppression Mode Search (ON and Actively Overdriving)
During AF Suppression overdrive, No monitoring periods are started after: Test pulses Test pulse inhibition Back-up pulses If AF Suppression stops overdriving, a new monitoring period is started. If no atrial events are sensed → Fast Mode. If at least one atrial event is sensed → Slow Mode
42
Threshold Search AF Suppression Mode Search (ON and Actively Overdriving)
Second overdrive stimulus is a test pulse @ AF Suppression rate interval Max. AF Suppression Rate Second overdrive stimulus is a test pulse @ AF Suppression rate interval Second overdrive stimulus is a test pulse @ AF Suppression rate interval AFx Intrinsic Rate Basic Rate
43
ACap™ Confirm Threshold Search Interactions
44
ACap™ Confirm Threshold Search Interactions
Ongoing ACap Confirm threshold search is terminated upon wand placement Search is reinitiated upon wand removal If an automatic ACap Confirm threshold search is scheduled to start while wand is in place: Search delayed until wand is removed If weekly trend log is scheduled to be logged, it will be logged upon wand removal
45
ACap™ Confirm Threshold Search Interactions
If AMS entry occurs during a threshold search the algorithm is suspended until one hour after AMS exit Atrial threshold search is suspended/delayed until the following processes are complete Automatic P and R-wave measurements Lead surveillance Actively storing a SEGMs Ventricular AutoCapture threshold
46
ACap™ Confirm Threshold Search Exceptions
Amplitude is set to 5.0 V (high output) if: Capture verification occurred at 0.0 V Amplitude increases to V and capture fails If ACap Confirm goes to High Output Mode, a new threshold search will be performed after one hour If capture threshold fails or is inhibited a new threshold search may be re-initiated in one hour, 8 or 24 hour clock is not reset If the atrial rate >120 bpm the threshold search will be canceled and attempted in 1 hour
47
ACap™ Confirm High Output Mode
High Output Mode popup message when: ACap Confirm is at maximum output ACap Confirm has been auto-programmed to OFF and the atrial pulse amplitude has been set to 5.0 V High Output Mode popup message when ACC is Monitor: ACap Confirm could not determine the threshold Are these for in-clinic measurements only? Does it really program it off after 1 HOM determination? Because next slide says that it will try again in 1 hour.
48
ACap™ Confirm Interactions
If an Atrial lead configuration switch occurs: ACap Confirm in Monitor Mode: ACap Confirm is turned OFF (Atrial pulse amplitude remains at programmed value) If ACap ConfirmON: ACap Confirm is turned OFF (Atrial pulse amplitude is set to 5.0 V. Pulse width unchanged
49
ACap™ Confirm Interactions
At ERI, if ACC mode is ON ACC is programmed OFF Pulse amplitude is set to 2 x (average of 4 most recent atrial capture thresholds), minimum 2.0 – maximum 5.0 V At ERI, if ACC mode is Monitor Pulse amplitude remains at programmed value
50
ACap™ Confirm Summary ACap™ Confirm Setup Test prior to enabling the algorithm Periodic threshold searches and amplitude adjustments Programmable Modes: Off, On and Monitor Follow-up EGMs and trend data available
51
ACC and VAC Differences
V. AutoCapture ACap Confirm Threshold search, no beat to beat capture verification or capture recovery Threshold search, beat to beat capture verification Capture detection: PDI method Capture detection: PDI and DMAX method Fusion Avoidance -Rate overdrive Fusion Avoidance - Short AV delay 40 ms backup (5.0V) 100 ms backup (5.0V) 3 consecutive LOCs to confirm LOC 2 consecutive LOCs to confirm LOC Programmable “ON” option allows amplitude adjustment after search capture Amplitude adjusted after search Output set to defined safety margin (1.7:1) Output set to threshold V margin