Maselli Beverage Analyzer IB05
2
IB05 Beverage Analyser - Brix - Diet - CO2 Control UC04 Panel
Carbonation analyser UR21 Brix / Diet analyser (refractive index based technology) Electrical Connections Input / Output Standard Parameters: - Brix - Diet - CO2 Optional Parameters: - Sugar inversion - Alcohol - Oxygen Sampling pump
3
Maselli OEM Configuration
End Users OEMs MR03 Control Panel UC04 Carbonation analyser UR21 Brix / Diet analyser (refractive index based technology) Approved by: Krones, KHS, Sidel, Miteco
4
Maselli UC05 Carbometric Theory
5
Gas dissolved in liquid
THE QUANTITY OF THE GAS DISSOLVED IN A LIQUID IS GENERALLY INFLUENCED BY DIFFERENT FACTORS liquid nature gas nature PRESSURE TEMPERATURE The system is made of 2 components in 2 phases so 2 variables (p, T) completely define its status
6
Henry’s Law For gas dissolved in liquid, without any reaction
xGAS: gas molar fraction PGAS: gas partial pressure k: constant At equilibrium status, where free liquid volume is gas saturated, reached after mechanical stressing of liquid, pressure p is PGAS Measuring temperature T allows to determine gas content
7
Gas content tables
8
CO2 in deaerated beverage
Experimentally the quantity of CO2 in the deaerated beverages depends on equilibrium PRESSURE and TEMPERATURE as follows:
9
CO2 measuring method Product sampling
Mechanical stressing to reach the equilibrium condition Temperature and pressure detection Data eleboration and calculation for CO2 content
10
UC05 Construction Main Motor Cap valves By Pass system Measuring Cell
Pt1000 Magnetic Elements Pressure sensor Main Motor Cap valves By Pass system Stirring Element Measuring Cell Vacuum Membrane
11
UC05 Analysis steps Sampling Creation of extravolume / stirring
12
UC05 Analysis steps Washing Pressure, Temperature data acquisition
13
UC05 Mechanical diagram of the cycle
DIAGRAM OF PRESSURE IN THE MEASURING CELL DIAGRAM OF DIAPHRAGM POSITION DIAGRAM OF VALVES POSITION PRESSURE AND TEMPERATURE DATA ACQUISITION STIRRING SYSTEM START STIRRING SYSTEM STOP TOTALLY OPENED VALVES TOTALLY CLOSED VALVES POSITION IN WHICH THE INDUCTIVE SENSOR SIGNAL STARTS THE CYCLE AND THE STIRRING SYSTEM.
14
UC05 Carbonation Analyzer Test of stirring efficiency
Pipe Line pressure Transient time Line in on Line off 7s of Stirring time before measuring pressure 3s of Stirring time before measuring pressure Stirring time No Stirring before measuring pressure
15
Maselli IB05 Diet Theory
16
Focus on Diet Soft drinks
Water Diet Syrup: aspartame phosphoric acid caramel caffeine water etc. DIET COKE Laboratory Analyzer (spectrophotometer): phosphoric acid of the Diet Coke with spectrophotometer (the control established by measuring one of the components) Maselli IB05 Inline: nD of the Diet Coke with refractometer (the control is established by measuring everything that is not water)
17
Diet Analysis Laboratory - Line
Traditional lab methods Maselli IB05 The measurement unit in the laboratory is different from the measure unit in the line A recipe with a specific conversion factor has to be used .
18
Diet Control How good is the correlation???
The IB05 measures the refractometric degree The IB05 has the capatability to control ‘what is not water’ of a diet beverage correlating the results with existing lab methods The capatability has been demostrated performing several experiments with different products How good is the correlation???
19
Maselli IB05 Correlation with the Lab
Diet soft drinks readings: Maselli vs Lab 30 measurements: one sample each 2’ Valve change Data from CCE Wilsonville, USA (QA Manager Will Looney)
20
Maselli Diet from the Line to the Lab
Maselli Diet in the line Our next step is to move this technology from the line to the laboratory! Maselli Diet in the lab DR10 Easy Diet Lab Monitor
21
Refractometric Brix and dilution
22
Refractometric Brix and dilution
23
Refractometric Brix and dilution
24
Refractometric Brix and dilution
25
Refractometric Brix and dilution
26
Conclusion Around 99.7 % of a diet beverage is water
Selective measuraments control only a fraction of what is not water IB05 and DR10 control everything is not water The IB05 measurement is equivalent to the Dr10 measurement IB05 and DR10 have the capatability to follow diluition A linear correlation between refractometric brix and diluition by weight has been proven for all the different beverage In order to match DR10 and IB05 to not a selective measurements a specific coefficient has to be used.
27
IB05/DR10 and traditional methods
28
Maselli IB05 Installation
29
Maselli Diet Flavor Changeover
No, during changeover when the valve to the filler is still closed, the Maselli pump would circulate always the same product. YES, during changeover when the valve to the filler is still closed, the Maselli pump would circulate more product being the sample more representative.