Resolution ä What can be expected ? ä What affects it ? ä How can it be improved ?
2
What is Resolution ? ä How close two points can be and still be seen as two separate points is known as the resolution of an optical system.
3
What is “good” resolution ? ä What are the specifications for Bio-Rad Confocals?
4
Lateral (XY) Resolution ä How closely can two points in the XY plane approach and still be distinguished ? ä In reflection, the specification is 100 nm. ä In fluorescence, the specification is 200 nm.
5
Axial (Z) Resolution ä What is the thinnest sample you can see ? ä In reflection, the specification is 500 nm. ä In fluorescence, the specification is 800 nm.
6
What Affects Resolution ?
7
The Sample Makes a Difference ä Resolution for reflection is different to resolution for fluorescence. ä Something which fluoresces will appear bigger than when it reflects.
8
Other Important Factors ä The light wavelength has a fundamental role to play. ä Vibration destroys resolution ä Optical alignment is crucial ä The condition of the laser ä The cleanliness of the microscope.
9
How can we Improve Resolution ?
10
Here are some Hints
11
Avoid Refractive Index Mismatch ä Ensure you use the correct immersion medium for the sample and lens. ä Do NOT mix manufacturers. ä Keep the optics clean
12
Keep the System Aligned ä Regular alignment of the system is well worth the trouble. ä Regular tuning of the laser is essential
13
Avoid Vibration ä Use an anti-vibration table if possible. ä If not, use a sturdy table, isolated from the room walls. ä Keep vibration sources away from the microscope table. ä Vibration sources include the laser fan, the computer, the epi-fluorescence power supply and printers.
14
Use Good Lenses ä The best investment is in top quality lenses. ä A cheap lens is not good economy as it will never give satisfactory results. ä Plan apo-chromat lenses are the best to use.
15
Pay some Attention to the Room ä Ensuring the room has adequately controlled lighting and is away from drafts and noise is just good sense. ä Extract the laser cooling air from the room. ä Restrict the use of the room to microscopy. General purpose rooms are never properly functional.
16
.........and above All !
17
Pay my bill.