Nevada Test Site Calibration Range and Geometric Calibration of Digital Cameras ISPRS I/2 Bob Kletzli Spectrum Mapping, LLC December 2-4, 2003
2
Outline Engineer Sensor For Geometric Stability
Lock Lenses at Infinity Before Calibration Lab Radiometric Calibration Terrestrial Geometric Calibration In Situ Calibration Calibration Verification NTS
3
Geometric Stability CDD Fixed To Milled Optical Bench
Lens Receptacle Is Opposite Side Each Lens Fixed To Aluminum Socket Lenses Are Interchangeable Alignment Pins Eliminate Rotation TEC Surrounds CCD “Shake and Bake”
4
Lens Focus Lens Focused At Infinity High Frequency Target At 100m
Each Lens Adjusted & Pinned
5
Radiometric Calibration
Lab Tools Include: Integrating Sphere Spectralon Panels Spectral Line Lamps Lab Processes Dark Current Flat Fielding White Balance
6
Terrestrial Range
7
Terrestrial Camera Stations
8
Bundle Result 1
9
Bundle Result 2
10
Bundle Result 3
11
Bundle Summary
12
Insitu Range (NM) 400+ points by RTK 2-3 cm accuracy
Photo Identifiable Variable terrain Used for both laser & photography Building corners surveyed
13
Insitu Range Closeup
14
NTS Calibration Range
15
NTS LandSat
16
NTS Site
17
Flight Plan
18
GPS Trajectory NLV to NTS
19
GPS Fwd/Rev RMS (NLV)
20
GPS Fwd/Rev RMS (NTS)
21
1:12,000 Images/Control
22
1:12,000 Flight
23
1:18,000 Images/Control
24
1:18,000 Flight
25
1:24,000 Images/Control
26
1:24,000 Flight
27
Summary Digital Cameras Can be Calibrated to Conventional Specifications Calibrations Are Maintained Over Time Regional Test Ranges Are Extremely Useful For Calibration and Verification