1 Methods and Tehniques in Surface Science Prof. Dumitru LUCA “Alexandru Ion Cuza” University, Iasi, Romania

2 Auger Electron Spectroscopy (AES) – short historic and physical background, How AES measurements are performed, Information derived from Auger spectra: Methodology Data analysis Experimental considerations Outline

3 Short historic of the Auger spectroscopy Pierre Auger, in 1925 observed (at first in the cloud chamber, then in photographic plates) the occurrence of electrons with precisely determined energies. These electrons have been later named Auger electrons) may serve to identify their parent atoms. 1953 J. J. Lander – the idea of using the Auger electrons in surface analysis.. The AES has been implemented as an analytic tool in 1967 (Larry Harris), after increasing the method sensitivity by using differential spectra to discriminate the tiny Auger peaks in the electronic spectra. 1968 – Auger spectrometer with CMA in modern configuration. Pierre Auger Beam current as low as 1 nA Probing depth: 0.5 - 10 nm, and < 10 nm lateral resolution(!!!) Covering a wide range of elements which can be detected (except for H and He). High sensitivity: 100 ppm for most of the elements. Nowadays:

4 Auger spectra. Expanation of the Auger effect in free atoms 1. The occurrence of an electron vacancy in a core level (K, L) (core level), by incident electron, X – ray photons, or ions. Little information is available for the energy of the primary and ejected electrons, due to complex cascade of successive collisions with the matrix. Therefore the complex picture in Fig. 1 2. The vacancy is filled by a second electron coming from an upper energy level. 3. The energy of the emitted electron can serve for: - the emission of a X photon (Z > 30) - ejection of a 3-rd (Auger) electron via a non-radiative process. 4. The net result: an atom in a double-ionized state + 2 emitted electrons emisi (the K core level electron and the Auger electron).

5 Auger transition nomenclature KL 1 L 2 Labelling: The 3 letters specify the energy levels implied in the process of emission of the Auger electrons Auger electron Incident electron KE = E K -E L1 -E * L2 - , E K, E L1, and E L2 – the energy levels mentioned in the labeling (generally different from the neutral atom, due to the presence of electron vacancies). AVV

6 Factors influencing the Auger peak area 1. Ionisation cross section 10 keV incident electron beam 3keV incident electron beam KLL LMM MNN

7 2. The Auger yield 3. Backscattering Competition between the Auger process and the X-ray fluorescence. The probability of occurrence of the Auger electrons increases with the decreasing of the differences between the energy levels involved in such transitions. Factors influencing the Auger peak area (cont’d)

8 Auger electron spectra Auger spectrum Kinetic Energy The tiny Auger peaks are difficult to process. They become visible after 10 x magnification. Electrons in the Auger spectra have energy values between 280 eV (KLL, Zn) si 2100 eV (S). Example: The KL 1 L 2 transtion of Si occurs at the energy of 1600 eV (since the difference between the energies of L 1 and K levels for Si is1690 eV, while the difference between L 1 and L 2 levels is 90 eV). After L1 → K de-excitation either an X-ray photon (K  of  Si), or an Auger electron may be emitted having energy of approx. 1690 – 90 = 1600 eV. 1. Direct spectrum

9 Even more features occur in the d[E*N(E)]/dE vs. KE plots. This is the most used kind of Auger electron spectrum. More features of the spectra occur clearly in the dN(E)/dE vs. K.E. Auger electron spectra 2. Differential spectrum

10 Auger spectra of light elements (the y-axis differs for different elements)

11 AES sensitivity Electrons emitted in the solid will “escape” in vacuum if they are not inelastically scattered. The scattered electrons vill have energies less energy than Auger electrons, thus they will occur in the “tail” of the spectrum, towards smaller KE values, along with secondary electrons. Many electrons will fully lose their energy via inelatic collisions in the solid. Therefore, only Auger electrons originating in the surface region (which did not experience inelastic scattering) will be collected by the analyser. Auger electrons Background MNN 95% of the electrons leaving the surface originate in a layer 3 depth.

12 Experimental arrangement

13 Instrumentation Specimen Electron gun Cylindrical mirror analyser (CMA) Ion beam UHV chamber Specimen import unit Electronics Computer and software Electron detector

14 Scanning Auger Microscopy (SAM) AESAuger Electron Spectroscopy SAMScanning Auger Microscopy: Same instrument can provide SEM imaging, Auger spectra and chemical Auger mapping. Specimen Focussing & scanning system of the incident e-beam Ion beam Applications 1keV incident electron beam → penetration depth of about 15 Å. Verification of surface contamination freshly prepared in UHV. Investigation of the thin film growth process + elemental analysis. Depth profiling of concentration of chemical elements.

15 Qualitative analysis 1.First, the main Auger peak positions are identified. 2.These values are correlated with the listed values in the Auger spectra book or standard tables. The main chemical elements are thus identified. 3.The identified element and transition are labelled in the spectrum (close to the negative jump in the differential spectrum). 4. The procedure is repeated for so-far unidentified peaks. The Auger spectrum of a sample under investigation E 0 = 3keV Elemental identification procedure

16 Example: From the differential AES spectrum Ni, Fe and Cr have been identified. Ni Fe Cr Qualitative analysis

17 Information concerning chemical composition Peak shape and the energy values, corresponding to maxima contain information on the nature of the environment, due to addition relaxation effects during the Auger process A full theoretical model is difficult to construct. In practice, Auger spectra of standard samples are used and the results are drawn from spectra comparison.

18 SAM image: red =Al; blue = F; magenta = Al + F Red = Al; green = O red =Al; blue = F; green = O SEM surface image Al+F+O SAM SEM and Auger images of an aluminium oxide surface, in absence and presence of fluorine contramination.

19 1. Measuring the peak-to-peak height dE  N(E)/dE vs. E N(E) vs. E 2. Measuring the peak area (after background subtraction) Quntitative analysis

20 For a homogeneous sample, the Auger peak intensity is given by: I i : Intensity of the detected current, due to the ABC Auger transition of the i element, I P : Incident beam current, N i : Concentration of the element i in the surafce,  i : Ionization cross-section on the A-level of the element i by the electrons from the incident beam,  i : Probability of the Auger ABC transition of the element i, r: ionization cross-section on the A-level of the element i by the electrons scattered in previous processes, : mean free path for inelastic collisions,  : incidence angle of the primary beam, F: Correction factor, dependent on the entrance solid angle in the analyzer, T: Transfer function of the analyzer, D: Detection efficiency, R: Roughness factor of the surface. Remarks 1. Deriving N i from the previous equation is rather difficult, due to large number of implied parameters, 2. In applications, empirical methods are used, which leave from: (a) utilization of standard specimens; (b) utilization of sensitivity factors. Factors affecting the peak intensity

21 Quantitative analysis using standard specimens Advantages: No need to know “obscure” physical quantities: ionization cross-section,  i on the A-level of the element i by the electrons in the primary beam, the Auger yield, backscattering cross-section and electron escape depth values. Drawbacks: Necessity to prepare standard samples, Valid only for homogeneous samples, Quite low accuracy.

22 Quantitative analysis using sensitivity factors Measurements are done under the same conditions on standard samples to cancel the correction factors, associated to the set-up particularities: Sensitivity factors, S i, have been measured for certain energy values of the electrons in the primary beam. They are tabulated for all the chemical elements. The atomic concentration of the elment a in the sample with N elements can be derived using the equation: The method is semi-quantitative, since the back-scattering effects and the escape depth of the electrons are negliged.

23 They do not include the so-called matrix effect of the sample: the inelastic mean free path ( ), the back-scattering factor (r), chemical effects on the peak shape in the Auger spectra, effect of the surface roughness. All these drawbacks result in errors of about 15%. The errors can be diminished to 1% by using standard samples with the same matrix, to derive S i. Fe Cr Fe Ni Peak-to-peak height: S i Cr @ 529eV: 4.70.32 Fe @ 703eV: 10.10.20 Ni @ 848eV: 1.5 0.27 Example Drawbacks of empirical methods