BINP level meter for cold liquids The method of measurements: change of the capacitance as function of a level of a liquid Originally designed for LN 2 at the CMD-2 detector Used for LXe calorimeter now Sensor: wire 1.2mm at axis of tube 3mm (inner) capacitance 117 pf (0cm) 212 pf (160cm) 0.6 pf/cm Material: stainless steel and teflon clean! Electronics: conversion C f resonance V
2
The lever meter for the LXe calorimeter 1- LXe calorimeter 2- short sensor (145 mm) 3- long sensor (1600 mm) 4- teflon insulator (10 cm step) 5- 1.2 mm stainless steel wire 6- stainless steel tube with inner diameter 3 mm
4
LN 2 level meter at the CMD-2 10 cm
5
LXe level meter at the calorimeter