IGRT Prostate – the ADHB experience Nicola Gordon Imaging Specialist RT ADHB
2
A little history… 2009 – Orthogonal EPIs #1-3 and weekly – Radiation Oncologists reviewed images 2010 – Introduction of fiducial markers for daily IGRT and RT led image review began Late 2010 – first CBCT capable Linac went clinical – CBCT used for some trial patients 2011 – began using CBCT to monitor bowel and bladder status on a weekly basis Mid 2012 – started using VMAT routinely
3
Why CBCT?
4
Evolution of process CBCT then kV imaging #1-3 & weekly kV then CBCT #1-3 & weekly – Patients asking if it was a scan day – General lack of understanding from patients kV then CBCT daily
5
Bladder and Bowel Advice
6
Bladder filling decision rule 50Gy Isodose curve (maroon) used as minimum bladder filling point.
7
Rectal filling. If fiducial match fit was good on kV imaging then treatment continued, with CBCT only for feedback. If fiducial match fit was not acceptable then the CBCT was used to assess why.
8
But… 2 imaging modalities in each fraction for fiducial patients goes against ALARA principle Time an issue – increased risk of intrafraction motion. Effectively soft tissue matching even with fiducials
9
Soft tissue matching Some inappropriate fiducial positioning Patients who were not suitable for fiducials Prostate Bed patients had soft tissue matching Consultants preferred fiducials for the majority of patients
10
Asking our colleagues… 5 departments use Elekta XVI to match to fiducials on CBCT daily 1 department using Elekta XVI to match soft tissue on CBCT daily 1 department using Siemens to match soft tissue on CBCT daily 2 departments using Varian/ARIA to match to soft tissue on CBCT daily 5 departments using Varian/ARIA to match to fiducials on kV daily 1 department using Varian/ARIA to match to fiducials on CBCT #1-3 and weekly and using kV on the other fractions.
11
Using CBCT to match fiducials Varian OBI gives only couch corrections that can be applied Matches done without rotation so need to be ‘best fit’ If unable to align all 3 fiducials there is probably a reason!
12
How to visualise Fiducial ContourColour Blend
13
Automatch settings. Bounding Box to cover PTV then use Intensity range and structure VOI of the PTV for automatch.
14
The process as it stands now Automatch Manual adjustment as required Fiducial match fit Decision to treat based on bladder and bowel filling Dose escalation to 78 Gy
15
Current Rectum Decision Rules Rectum should not be displaced more than 1cm anteriorly into the PTV contour.
16
Rectum decision rules Before:28 Minutes later:
17
Video of process
18
Other initiatives Updated (simplified) Dietary Advice Weekly MDT ‘On Treatment Chart Round’ Interventions such as ‘tubing’ MOODLEs
19
Time… Post Treatment CBCT study: Prostate VMAT with Fiducials 6 Min Prostate VMAT without Fiducials 7.5 Min Prostate Bed VMAT 6.94 Min Prostate IMRT (Soft Tissue Match) 12.94 Min
20
Intra-Fraction Movement - Range Fiducials: -0.7 to +0.6 (max -0.7 vrt) Non Fiducials: -0.4 to +0.2 (max -0.4 vrt) Prostate Bed: -0.5 to +0.5 (max -0.5 vrt) Prostate IMRT (longer treatment time): -0.3 to +0.4, (max +0.4 - lng)
21
Moving forwards… No more fiducial markers! Finish Accuracy Study Look at toxicity scores
22
Thanks Helen McCabe and the team on MV4 for all their help with Data collection and feedback Sheonagh Nielson Rob Lane All those who responded to the questions we sent out
23
References Moseley DJ. White EA. Wiltshire KL. Rosewall T. Sharpe MB. Siewerdsen JH. Bissonnette JP. Gospodarowicz M. Warde P. Catton CN. Jaffray DA. Comparison of localization performance with implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate. International Journal of Radiation Oncology, Biology, Physics. 67(3):942-53, 2007 Mar 1. Lazos et al, (2012) http://astro2012.abstractsnet.com/handouts/011498_Shifts_Study_ASTR O2012_Poster.pdf http://astro2012.abstractsnet.com/handouts/011498_Shifts_Study_ASTR O2012_Poster.pdf Shi W. Li JG. Zlotecki RA. Yeung A. Newlin H. Palta J. Liu C. Chvetsov AV. Olivier K; Evaluation of kV Cone-Beam CT Performance for Prostate IGRT American Journal of Clinical Oncology. 34(1):16-21, 2011 Feb. NHS (UK) National Cancer Action Team National Radiotherapy Implementation Group Report on IGRT – Guidance for Implementation and Use, August 2012 – Section 18.5